Eagle Eye SG-1000 User Manual

Manual is about: Specific Gravity Tester

Summary of SG-1000

  • Page 1

    Sg-1000 specific gravity tester user manual the best battery testing equipment. The best customer support. Period. Www.Eepowersolutions.Com copyright © 2012 eagle eye power solutions, llc v1.0

  • Page 3

    Eagle eye sg-1000c user manual 2 trouble? If abnormal measurements are experienced, it may be due to low battery power. Replace with new alkaline battery. Caution care should be taken to avoid damage to the prism. Use soft tissue paper when wiping the prism surface to avoid scratching. Never make co...

  • Page 4: Table Of Contents

    Eagle eye sg-1000c user manual 3 table of contents caution ........................................................................................................ 2 1. Names and functions of parts ................................................................ 4 2. Inserting the nozzle ..............

  • Page 5

    Eagle eye sg-1000c user manual 4 1. Names and functions of parts.

  • Page 6

    Eagle eye sg-1000c user manual 5 1. Display the lcd provides a digital readout of the measured values. 2. Select key used to select a measurement item. The item to be displayed switches between temperature and measure each time this key is pressed. (specific gravity of electrolyte) 3. Zero key used ...

  • Page 7

    Eagle eye sg-1000c user manual 6 2. Inserting the nozzle the nozzle is packed separately from the main unit. Insert the nozzle to main unit following the procedure described below. (1) push the nozzle into the hole located (2) then, fit the nozzle into the slit by at the bottom of the sample chamber...

  • Page 8

    Eagle eye sg-1000c user manual 7 3. Inserting the battery (1) using a screwdriver, turn the locking pin to the “open” position and open the battery cover. (2) insert the battery as indicated, take note of the polarity. Note: the battery (006p) provided with the unit as an accessory is to be used to ...

  • Page 9

    Eagle eye sg-1000c user manual 8 4. Attaching the wrist strap attach the wrist strap to the hook located on the lower side of the main unit. When measuring samples, loop the strap around your wrist to avoid any accidents..

  • Page 10

    Eagle eye sg-1000c user manual 9 5. Zero setting (1) remove the nozzle cap (3) confirm that the sample chamber has been filled with water. If bubbles are present in the sample chamber, operate the tube pump again to refill water into the sample chamber. (2) dip the tip of the nozzle in water, and sq...

  • Page 11

    Eagle eye sg-1000c user manual 10 (4) press the zero key. “000” will blink three times and “000” will be displayed. (5) zero-setting is now complete. Remove the nozzle from the water and squeeze the tube pump to drain the water from the sample chamber. Allow the unit to equilibrate to the ambient te...

  • Page 12

    Eagle eye sg-1000c user manual 11 6. Measurement of sample (1) press the select key to move the triangle mark to the desired measurement item on the display. (2) dip the tip of the nozzle into the electrolyte sample and squeeze and release the tube pump a few times to draw the sample into the sample...

  • Page 13

    Eagle eye sg-1000c user manual 12 (4) the displayed value will remain on the lcd for 5 minutes. (5) to reset the display, press and hold down the start switch for 2 seconds. (6) before storing the sg1000f, squeeze and release the tube pump several times to drain the sample from the chamber. Then, re...

  • Page 14

    Eagle eye sg-1000c user manual 13 7. Explanation of error messages the sg-1000 will indicate an error message when there is a problem with the operation on the instrument. Error messages that may be displayed are explained below. “aaa” zero-setting error zero-setting was performed without water in t...

  • Page 15

    Eagle eye sg-1000c user manual 14 (1) press the select key to move the triangle mark to temperature, and the temperature will be displayed. (2) the temperature indicated is the temperature of the sample chamber and is not a direct measurement of the temperature of the sample. 9. Maintenance (1) to r...

  • Page 16

    Eagle eye sg-1000c user manual 15 symptom remedy the unit will not turn on. The contact plate in the battery compartment may have been bent creating a gap between the plate and the battery. Restore contact between the contact plate and battery by using a screwdriver to return the plate to the normal...

  • Page 17

    Eagle eye sg-1000c user manual 16.

  • Page 18

    Eagle eye sg-1000c user manual 17 10. Trouble? If any problem in operation is suspected, check the following reference list for repair before contacting the manufacturer. Symptom what to check 1.”hhh” is displayed at zero-setting. Does the battery need replacement? 2. “aaa” is displayed at zero-sett...

  • Page 19

    Eagle eye sg-1000c user manual 18 *is the power level of the battery low? The unit does not measure if the battery is low. The battery level can be checked by using a multimeter. If the mattery is low, replace it with a new battery. (refer to page 8 on how to replace the battery and other remarks.) ...

  • Page 20

    Eagle eye sg-1000c user manual 19 *is the tube pump damaged? *is the nozzle damaged? *is the cover of the sample chamber damaged? *is the nozzle loose at the insertion point of the sample chamber? Squeeze the tube pump to draw water into the sample chamber. At the same time check the tube pump, nozz...

  • Page 21

    Eagle eye sg-1000c user manual 20 *is there a difference between the room temperature and sample temperature? The unit automatically compensates for temperature by detecting the temperature of the sample in the sample chamber. If the temperature of the sample is different from that of the unit, wait...