Eagle Eye SG-1000 User Manual - page 20
Eagle Eye SG-1000C User Manual
19
*Is the Tube Pump damaged?
*Is the Nozzle damaged?
*Is the cover of the Sample
Chamber damaged?
*Is the Nozzle loose at the
insertion point of the Sample
Chamber?
Squeeze the Tube Pump to draw water
into the Sample Chamber. At the same
time check the Tube Pump, Nozzle, and
cover of the Sample Chamber for
damage. Make sure the Nozzle is
securely attached. (Leakage of liquid or
air can be observed if the Tube Pump,
Nozzle or cover has been damaged, or
the Nozzle is loose.)
If damage is found, replace the
damaged part. (Two spare Nozzles have
been provided as accessories.) Parts can
be purchased from your dealer or
manufacturer. See page 4 and 5.
*Is the prism surface stained or
blurred?
Remove the cover of the Sample
Chamber from the unit and check the
prism surface for stains or blurs.
If stained or blurred, using a soft tissue,
wipe with water or ethyl or isopropyl
alcohol.