Edaphic AZ8205 Operation Manual - Fig. I
Fig. I
psi
4. The meter has defaulted as 80 psi
calibration point , the adjustable
pressure range is from 78.0 to 82.0
. if calibration pressure source is
not 80psi , to increase the figure by
pressing key , or decrease the
figure by pressing key to set
calibration point as required.
5. Save the calibration point by
pressing key , "SA" and small
"CA"appears on the display (See
Fig. J ) in 2 seconds , the meter
auto-skip to the negative pressure(
-) point for next calibration mode.
REC
REC
REC
REC
9
1. First , please manually set the
display to zero ( no pressure
applied to the connector ), refer to
the Manual zero procedure.
2. Turn the meter off.
3. Press & simultaneously
,"CA" appears on the display, (See
Fig.I )the meter enters to the
calibration mode ,make sure the
pressure unit to be pointed under
the arrow mark " " is "psi" to
start positive (+) pressure
calibration
DIF
DIF