EZ Smart 606 User Manual - page 2
P. 12
3.
As soon as enough blood has
filled the Confirmation Window
(see picture)
of the Test Strip,
the
EZ Smart-
606
Meter will
beep and begin counting down
from
8
to
1
.
Confirmation Window
4.
Your blood glucose result
will appear on the LCD
panel and will be stored into
t h e M e t e r m e m o r y
automatically.
Note:
To ensure accurate results, make sure the Confir-
mation Window of the Test Strip is completely filled with
your blood sample.
5.
The
EZ Smart-
606
results range between 20 to 600 mg/dL (1.1 to 33.3 mmol/l). If a
HI
message appears on your display, your
EZ Smart-
606
Meter has detected
that your blood glucose level is higher than 600 mg/dL(33.3 mmol/l). It is suggested
that you review your testing procedure and test again with a new
EZ Smart-
606
Test Strip to confirm the result.
If the same result occurs, consult your
healthcare professional immediately.
6.
If a
Lo
message appears on your display, your
EZ Smart-
606
Meter has de-
tected that your blood glucose level is lower than 20 mg/dL (1.1 mmol/l). It is
suggested that you review your testing procedure and test again with a new
EZ
Smart-
606
Test Strip to confirm the result.
If the same result occurs, consult
your healthcare professional immediately.
If no further tests are performed, the
EZ Smart-
606
Meter will automatically shut
off after 4 minutes.
P. 13
After finishing the test, we recommend using a tissue paper to remove
EZ
Smart-
606
Test Strip from the
EZ Smart-
606
Meter for proper disposal.
Caution:
Any change or administer of medication based on the
EZ Smart-606
blood
glucose test results without the consent advice of a physician or
healthcare professional is not recommended.
Memory Features
The
EZ Smart-606
Meter automatically stores the 28 most recent blood glucose results. It also provides you
with an average of your blood glucose test results.
1.
Press the M button to
turn on the
EZ Smart-
606
Meter
.
2.
All segments of the LCD display
will appear.
3.
The 28 set data aver-
age will automati-
cally appear, and
flash for few seconds.
4.
The most recent test result(1 st set) will
automatically appear.
Press the M button, the Meter displays
the data order release the M botton
shows its reading, accordingly. When the
28 memories are full, the oldest result is
removed and replaced with the newest
result.
P. 14
Exiting the Memory Mode
1. Press and release the M button to the end of the memory(000 will show). Press the S button
again to turn the meter off.
2.Meter will shut down automatically after 15 seconds with no further action.
Results marked as
ctl
(Control Solution) will not be averaged into the nor be stored into the
memory.
Care and Storage
1.
Handle the
EZ Smart-
606
Meter with care. Dropping or throwing the Meter may cause dam-
age to the device.
2.
Do not expose the Meter, Test Strips, and Control Solution to extreme conditions, such as high
humidity, heat, freezing cold or dust.
3.
Always wash hands with soap and water and rinse and dry completely before handling the
Meter and Test Strips.
4.
When cleaning the Meter, gently wipe the exterior surface using a damp soft cloth.
DO NOT
USE ANY ORGANIC SOLVENT
for cleaning.
5.
The
Meter should be stored at room temperature in a dry and clean space.
DO NOT STORE
IN DIRECT SUNLIGHT OR AREAS WITH HIGH HUMIDITY AND/OR DUST.
It is advised
that you store the Meter and its accessories into the provided carrying case.
We have provided a log book for your recording convenience. To learn more about how to record
your results, please refer to the log book for further instructions.
P. 15
For healthcare professionals using this system on multiple patients, please be aware that all
items that come in contact with human blood should be handled as potential biohazards.
Users should follow the guidelines for prevention of blood-borne transmittable diseases in a
healthcare setting for potentially infectious human blood specimens as recommended in the
National Committee for Clinical Laboratory Standards,
Protection of Laboratory Workers from
Instrument Biohazards and Infectious Disease Transmitted by Blood, Body Fluids and Tissue:
Approved Guideline.
NCCLS document M29-A [ISBN 1-56238-339-6]
.
Display Messages
DISPLAY
WHAT IT MEANS
ACTION
System check verifying that all seg-
ments of the Meter are functioning.
This message appears immedi-
ately if your Meter has not been
encoded.
No Action Necessary
Code Meter
(see page 7)
This is the current code number
in use.
This code number should
match the code number
printed on your
EZ Smart-
606
Test Strip vial.
P. 16
DISPLAY
WHAT IT MEANS
ACTION
The system is ready to accept a blood
sample.
The system is ready to accept a Con-
trol
Solution sample.
EZ Smart-
606
Meter detects blood
glucose level is higher than 600 mg/
dL(33.3 mmol/l).
The
EZ Smart-
606
Meter detects
blood glucose level is lower than 20
mg/dL (1.1 mmol/l).
You may now apply a blood sample.
You may now apply a drop of Con-
trol Solution sample.
Indicates high blood glucose level.
Repeat test. If result still occurs,
consult your physician immediately.
Indicates low blood glucose level.
Repeat test. If result still occurs,
consult your physician immediately.
P. 17
DISPLAY
WHAT IT MEANS
ACTION
Temperature is too low to
perform the test.
Temperature is too high to
perform the test.
An error message indicating a
problem with the
EZ Smart-
606
Test Strip or Glucode Chip.
An error message indicating a
problem with the
EZ Smart-
606
Meter.
Repeat the test in a warmer setting
(10 - 40
o
C or 50 - 104
o
F). Wait 15
minutes before re-testing.
Repeat the test in a cooler setting
(10 - 40
o
C or 50 - 104
o
F). Wait 15
minutes before re-testing.
Review the instructions and try
again with a new
EZ Smart-
606
Test
Strip or Glucode Chip.
Check the meter again with the
Glucode Chip. If the problem
persists, contact Tyson Bioresearch,
Inc. Customer Service for help.
P. 18
DISPLAY
WHAT IT MEANS
ACTION
Battery power is too low for fur-
ther usage.
Battery power is low. Meter will
provide approximately 50 more
measurements.
All testing values recorded in the
memory have been shown.
Replace with a new 3-Volt lithium
battery (CR-2032).
Replace with a new 3-Volt lithium
battery (CR-2032).
Press S button or wait for 15 sec-
onds to turn off the meter.
An error message indicating a prob-
lem with the coding of
EZ Smart-
606
Meter.
Check the meter again with the
Glucode Chip. If the problem
p e r s i s t s , c o n t a c t T y s o n
Bioresearch, Inc. Customer Ser-
vice for help.
P. 19
Other problems that may occur
ACTION
The
EZ Smart-
606
Test Strip has not been
inserted into the Meter properly.
Review the instructions and re-insert the
Test Strip correctly(white side face up).
Defective
EZ
Smart-
606
Test Strip.
Replace with a new
EZ Smart-
606
Test
Strip.
Insufficient blood sample.
Repeat the test with new
EZ Smart-
606
Test Strip.
Test Strip remains in the test port for more
than 4 minutes prior to testing.
Meter will automatically turn-off.
Re-insert the Test Strip to the test port.
LCD Screen on the Meter is blank when trying
to preform a test.
Contact Tyson Bioresearch, Inc.
Customer Service for help.
P. 21
v
The Lancets are in conformity with the
MDD 93/42/EEC.
Wuxi Xinda
Medical Device Co., Ltd.
No.42 Xixin Road, Zhangjing
Wuxi, Jiangsu,214194,PRC
+86-510-379-3149
v
The Lancing Device is being clasified as
MDD Class I.
Limitations of the Procedure
Caution: The
EZ Smart-
606
System is designed for
in vitro
use only and is not intended to test
on newborns.
The
EZ Smart-
606
Test Strips are designed for use with fresh capillary whole blood samples.
False results may occur when performing the test while severely dehydrated, severely
hypotensive, in shock or in a hyperglycemic-hyperosmolar state. If you believe you are
suffering from any of the above symptoms, consult a healthcare professional immediately.
EMDAR BV
Ijsselburecht 3, Postbus 5486
6802 EL Arnhem, Netherlands
+31-263-653-375
Wuxi Xinda
Medical Device Co., Ltd.
No.42 Xixin Road, Zhangjing
Wuxi, Jiangsu,214194,PRC
+86-510-379-3149
EMDAR BV
Ijsselburecht 3, Postbus 5486
6802 EL Arnhem, Netherlands
+31-263-653-375
P. 22
Service and Warranty
IMPORTANT:
The
EZ Smart-
606
Blood Glucose Monitoring System is designed for
in vitro
diagnostic use only. Tyson Bioresearch Inc. cannot endorse the performance of the
EZ
Smart-
606
System when used with Test Strips other than those designed for the
EZ
Smart-
606
Meter. The
EZ Smart-
606
System manufacturer warranty is valid only when used properly
within the guidelines of the provided Users Manual and is invalid when the
EZ Smart-
606
System and
EZ
Smart-
606
Test Strip are used improperly.
Manufacturer Warranty:
Tyson Bioresearch, Inc. guarantees that this device will be free of
defects in materials and workmanship for a period of three years from the date of original purchase.
During the stated three-year period, our company shall repair or replace any
EZ Smart-
606
Meter
found defective with a new
EZ Smart
-606
Meter.
This warranty does not apply to the performance of a
EZ Smart-
606
Meter that has been acciden-
tally damaged, altered, misused, tampered with or abused in any way. In no event shall our
company be liable to the purchaser or any other person for any incidental, consequential, or
punitive damages arising from or in any way connected with the purchase or operation of the
EZ
Smart-
606
Meter or its parts.
For manufacturer warranty services, purchaser must contact Tyson Bioresearch, Inc. for
help.
P. 23
Customer Service
If you have any questions or concerns regarding this product and or its operation, or any attempt
to correct a problem fails, please call our Customer Service Unit. Our trained specialists will be
happy to assist you, train you, or even reaffirm your results.
Customer Service Hotline:
886-37-585998
our office hours are from
8:30am (GMT0030) - 5:30pm (GMT0930)
or E-Mail us at
service@tysonbio.com
When you call our Customer Service Unit, please have your
EZ
Smart-
606
Meter,
EZ Smart
-
606
Test Strip and all other system supplies available. This will allow us to answer any of your ques-
tions with speed and efficiency.