Falco Systems WMA-02 User Manual - page 2
About this manual
This user manual is an integral part
of the Falco Systems WMA-02
amplifier product. Please read it
carefully and pay attention to the
recommendations
and
the
instructions for safe use.
General description
The Falco Systems WMA-02 is a
high voltage amplifier that runs from
an 24 – 30V DC power supply and
can be battery powered if required,
e.g. for use in a Faraday cage. Its
internal high voltage generator is
inductor-free and hence does not
spread magnetic interference in
sensitive experimental setups. Its
large voltage range makes it an
excellent choice for use with MEMS
devices, EO-modulators, PZT (piezo)
positioning systems, beam steering
components, and many more.
The amplifiers’ full-power frequency
response
has
a
-3dB
corner
frequency at 13kHz. The input
impedance is 100kΩ and the output
50kΩ.
The output noise is at this price level
impressively low at 5mVrms. For
extremely low noise applications
where bandwidth is not an issue, a
dedicated WMA-02LF version is
available that contains an output low
pass filter capacitor which reduces
the bandwidth to 50Hz, but makes
the noise drop to 0.2mVrms. This
bandwidth-limited amplifier is ideal
for MEMS or piezo positioning with
manual voltage adjustment requiring
extremely low noise.
The WMA-02 contains an over-
voltage protection for the daily use
and abuse in the laboratory and is
long-term short-circuit proof. The
input (indicated “Input” on the WMA-
02 label), output (indicated “Output”)
and DC power supply (Indicated “24
– 30V DC 120mA”) connections are
all standard BNC.
The amplification is 20x (fixed). With
a range of –175V to +175V, a –3dB
bandwidth of 13kHz, possibility for
battery powered operation, and
below 5mVrms output noise, this
amplifier
will
enable
many
measurements
in
different
applications.
Safety
- This product is able to produce over
175V,
but
the
50kΩ
output
impedance limits the current to a
rather safe maximum of 3.5mA.
Although this voltage is painful it will
under normal circumstances not be
lethal. Nevertheless take care. This
is a high voltage unit and safety
measures
should
be
taken
accordingly.
- The internal circuitry of the amplifier
operates at high voltage. Only
qualified
personnel
from
Falco
Systems
should
service
this
amplifier.
- When the amplifier is turned on or
off, a short voltage spike may appear
at the output which may damage
circuitry already connected to it.
- Replace the fuse with 250V
315mAT 5x20mm rated fuses only.
- The Falco Systems WMA-02 is only
suitable for indoor use in a class II
!