Game Ready GRPro 2.1 User Manual

Summary of GRPro 2.1

  • Page 1

    This user’s manual is also available online in english, french, german, italian, japanese, and spanish at www.Gameready.Com under products. Model numbers 550550-03, 550550-03-rn, 550550-53 g r pro ™ 2.1 c o n t r o l u n i t user’s manual.

  • Page 2

    T h e g a m e r e a d y d i f f e r e n c e t h e g a m e r e a d y d i f f e r e n c e game ready ® was built upon the concept that the principles behind spacesuit technology could be used to develop an injury treatment system. As a designer for the national aeronautics and space administration (na...

  • Page 3

    Slee p tim e tim e run ice bath tem p. Pres sure off low med high avg . Wra p tem p. 55/1 3 50/1 0 45/7 40/5 ºf/º c d e s c r i p t i o n o f t h e g r p r o 2 . 1 s y s t e m the following items are included in your grpro ™ 2.1 system: a wrap (comprised of an inner atx series heat exchanger and an ...

  • Page 4

    Detailed instructions for use ________________________________ 2 modes of operation _____________________________________ 2 buttons ________________________________________________ 2 adjusting temperature __________________________________ 3 display _________________________________________________ ...

  • Page 5

    Do not use this device without your physician’s specific recommendations for the frequency and duration of your treatments. The most common recommendation is to use cold therapy at least 4 times a day, for approximately 30 minutes each time, with at least a 30-minute break between treatments. While ...

  • Page 6

    Adjusting temperature to adjust the temperature being applied during treatment, simply turn the temperature knob towards 3 snowflakes for the maximum amount of cold, or towards 1 snowflake for the least amount of cold. Notice that as you adjust the knob, the “target temperature” on the display will ...

  • Page 7

    Pressure: off low med high off low med high off low med high off low med high when selecting a program: sleep time time run pressure off low med high 55/13 50/10 45/7 40/5 ºf/ºc target temperature actual temperature displays pressure. Displays amount of time unit will run. Displays amount of time un...

  • Page 8

    Operating the system to operate your grpro 2.1 system, you need: • control unit filled with ice and water as indicated by the fill line labels within the reservoir. Optimal performance is achieved by first adding 1.5 liters of water, and then filling the reservoir to the top with ice. • game ready s...

  • Page 9

    4 place the control unit in the location where you plan to use it. The control unit should only be used in the upright orientation as shown. The control unit will leak if placed on its side. 2 add water to fill-line indicated on label within the reservoir. Do not overfill. Add ice to top of reservoi...

  • Page 10

    6 connect the larger end of the connector hose (with the red button) to the control unit. Make sure you hear it click. To disconnect, simply press the red button and remove the connector from the control unit. 7 apply the selected wrap (consult the use guide accompanying each wrap for application in...

  • Page 11

    Manual mode: set the time in five minute increments by pushing the +/- buttons. Set the pressure by pushing the pressure button. You can select from 4 pressure settings: no pressure, low pressure (5-15mmhg), (medium pressure (5-50mmhg), high pressure (5-75mmhg). Program mode: push the program button...

  • Page 12

    Control unit the exterior of the control unit and the visible interior surfaces of the reservoir can be cleaned with a soft cloth and one of the following cleaning agents: • mild detergent • 70% isopropyl alcohol • 100% simple green ® • quaternary ammonium (such as virex ® – typically only found in ...

  • Page 13

    M a i n t e n a n c e the reservoir filter should be inspected, cleaned, and/or replaced as necessary. 1 identify the filter within the ice reservoir. 2 using two fingers, grasp and squeeze the two protruding prongs. 3 slide the filter out. 4 rinse debris from the filter and be sure there are no obv...

  • Page 14

    The grpro 2.1 control unit can be used with any of the following accessories: • any game ready wrap using atx series heat exchangers (wraps made by any other manufacturer cannot be used with this system) • game ready supplied fsp group, inc. Power supply model fsp 030-rcam and hospital grade power c...

  • Page 15

    Compression therapy (vasopneumatic compression) using the game ready system or any compression therapy device should not be used in patients: • who are in the acute stages of inflammatory phlebitis in the affected region. • who have any history or current clinical signs suggestive of deep vein throm...

  • Page 16

    • who have compromised local circulation or neurologic impairment (including paralysis or localized compromise due to multiple surgical procedures) in the affected region. • who have rheumatoid arthritis in the affected region. • who have a known and uncontrolled peptic ulcer since wraps confining a...

  • Page 17

    U l c l a s s i f i c a t i o n en-14 protection against electric shock: the grpro 2.1 system is considered to be class i (protective earth) when connected to the fsp group, inc. Model fsp 030-rcam power supply. Protection against harmful ingress of water: this product provides ordinary protection a...

  • Page 18

    E l e c t r o m a g n e t i c c o m p a t i b i l i t y electromagnetic compatibility information according to iec/en 60601-1-2 table 1 for emissions guidance and manufacturer’s declaration – electromagnetic emissions the grpro 2.1 is intended for use in the electromagnetic environment specified bel...

  • Page 19

    Table 2 for transient electromagnetic immunity guidance and manufacturer’s declaration – electromagnetic immunity the grpro 2.1 is intended for use in the electromagnetic environment specified below. The customer or the user of the grpro 2.1 should assure that it is used in such an environment. Immu...

  • Page 20

    Table 3 for rf electromagnetic immunity guidance and manufacturer’s declaration – electromagnetic immunity the grpro 2.1 is intended for use in the electromagnetic environment specified below. The customer or the user of the grpro 2.1 should assure that it is used in such an environment. Immunity te...

  • Page 21

    Table 4 for rf immunity recommended separation distances between portable and mobile rf communications equipment and the gr pro 2.1 the grpro 2.1 is intended for use in an electromagnetic environment in which radiated rf disturbances are controlled. The customer or the user of the grpro 2.1 can help...

  • Page 22

    Error what does it mean? What can i do? Air pressure sensor: the control unit has detected a problem calibrating the air pressure circuit on start-up. • this is most likely to occur if you have restarted the system with an inflated wrap attached. • detach the wrap, pressing it flat to expel the air ...

  • Page 23

    Error what does it mean? What can i do? Under pressure: the control unit can’t reach its target maximum compression. This often indicates that there is a leak in the pneumatic compression circuit, either in the connector hose, wrap or control unit. Or it may occur because the hook and loop fastener ...

  • Page 24

    En-21 error what does it mean? What can i do? Self-test error – fluid circuit: the control unit has detected an electronic problem in the fluid circuit on start-up. • turn off the control unit. • turn the control unit back on and resume treatment. • if the problem persists, contact customer service....

  • Page 25

    En-22 manufacturer’s warranty coolsystems warrants that the grpro 2.1 control unit, if properly used, will be free from defects in material and workmanship for a period of two (2) years after the date the grpro 2.1 control unit was purchased. If the grpro 2.1 control unit, which is the subject of th...

  • Page 26

    Period. No warranties apply after that period. Some jurisdictions do not allow limitations on how long an implied warranty lasts, so this limitation may not apply to you. Limitations of liability coolsystems’ responsibility under this, or any other warranty, implied or express, is limited to repair ...

  • Page 27

    Game ready control unit user’s manual pn 703758 rev e ©2012 coolsystems, inc. All rights reserved. Brought to you by coolsystems, inc. 1800 sutter street, suite 500 concord, ca 94520 1.888.Gameready (1.888.426.3732) +1.510.868.2100 www.Gameready.Com.