Game Ready GRPro 2.1 User Manual - page 11
Manual Mode:
Set the time in five minute increments by pushing the +/- buttons.
Set the pressure by pushing the pressure button. You can select
from 4 pressure settings: No pressure, Low Pressure (5-15mmHg),
(Medium Pressure (5-50mmHg), High Pressure (5-75mmHg).
Program Mode:
Push the program button to enter Program Mode. In Program Mode,
the unit will operate continuously according to the selected program.
You will need to replenish ice and water as previously indicated in
step 2.
You can select from the following programs: (Push the Program
Button to scroll through the available programs.)
Program 1:
30 minutes on, 30 minutes sleep. No pressure.
Program 2:
30 minutes on, 30 minutes sleep. Low pressure.
Program 3:
30 minutes on, 30 minutes sleep. Medium pressure.
Program 4:
30 minutes on, 60 minutes sleep. No pressure.
Program 5:
30 minutes on, 60 minutes sleep. Low pressure.
Program 6:
30 minutes on, 60 minutes sleep. Medium pressure.
Program d:
Drain Mode. Please refer to page 2 in this manual for
details.
10
You can choose to operate in either Manual Mode or Program Mode.
Manual Mode allows you to adjust time and pressure settings as you
choose. Program Mode allows you to choose one of six programs that
provide treatment for a set time then sleep (providing no treatment) for a set time,
continuously, at a specific pressure setting (refer to list of available programs below).
The unit automatically starts in Manual Mode.
S T O R A G E
When you are done using the System for the day:
•
Unplug the AC Adapter and the Connector Hose from the Control Unit.
•
Push the door release button to open the door.
•
Carefully pour out the ice and water.
•
Wipe off any excess water.
•
Store the Control Unit with the lid fully open to allow the interior to dry and preserve
the reservoir seal. Store the Control Unit in the Carry Bag or in another safe location.
Remember that your GRPro 2.1 Control Unit is a valuable piece of equipment and should
be treated with great care, like a laptop computer.
Storage Temperature Conditions: 33° - 120°F (1° - 50°C)
Relative Humidity: 10% - 95% non-condensing
Caution
: Do not keep in extreme cold or hot temperatures (below 33°F or above 120°F
or below 1° or above 50°C). Do not leave in a hot or freezing car. Do not leave the Control
Unit in direct sunlight. The UV light may damage or discolor the Control Unit.
EN-8