Hach TU5200 User Manual

Other manuals for TU5200: User Manual
Manual is about: Laboratory Laser Turbidimeter

Summary of TU5200

  • Page 1

    Doc022.5 2.80489 tu5200 09/2014, edition 1 user manual beta -beta draft-.

  • Page 2: -Beta Draft-

    Beta -beta draft-.

  • Page 3: Table of Contents

    Table of contents specifications ..............................................................................................................3 general information ..................................................................................................4 safety information....................

  • Page 4: Troubleshooting

    Clean the vial compartment .......................................................................................20 troubleshooting .......................................................................................................20 read the diagnostics information ...............................

  • Page 5: Specifications

    Specifications specifications are subject to change without notice. Specification details dimensions (w x d x h) 41 x 28 x 12.5 cm (16 x 11 x 7.7 in.) weight 2.37 kg (5.23 lb) enclosure rating ip20 protection class iii for the instrument ii for the external power supply pollution degree/installation...

  • Page 6: General Information

    Specification details certifications ce compliant; us fda accession number: 1420492-xxx. This product complies with iec/en 60825-1 and to 21 cfr 1040.10 in accordance with laser notice no. 50. Warranty eu: 2 years 1 this is based on formazin primary standard. General information in no event will the...

  • Page 7: -Beta Draft-

    Precautionary labels read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol on the instrument is referenced in the manual with a precautionary statement. Electrical equipment marked with this symbol may not be disposed o...

  • Page 8: -Beta Draft-

    This instrument is a class 1 laser product. There is invisible laser radiation when the instrument is defective and when the instrument lid is open. This product complies with en 61010-1, "safety requirements for electrical equipment for measurement, control and laboratory use" and with iec/en 60825...

  • Page 9: -Beta Draft-

    Certification canadian radio interference-causing equipment regulation, iecs-003, class a: supporting test records reside with the manufacturer. This class a digital apparatus meets all requirements of the canadian interference-causing equipment regulations. Cet appareil numérique de classe a répond...

  • Page 10: -Beta Draft-

    Figure 1 product overview 1 lid 6 usb port type a 2 vial compartment 7 usb port type b 3 display screen 8 ethernet port 4 power button 9 rfid scanner (optional) 5 power connection 10 usb port type a product components make sure that all components have been received. Refer to figure 2 . If any items...

  • Page 11: Installation

    Figure 2 product components 1 turbidimeter 4 power supply with adapters 2 stablcal ® calibration set 5 dust cover 3 sample vials 6 vial stand installation c a u t i o n multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document. This instrument is rat...

  • Page 12: -Beta Draft-

    Connect to power c a u t i o n personal injury hazard. Never remove covers from the instrument. This is a laser-based instrument and the user risks injury if exposed to the laser. C a u t i o n personal injury hazard. Do not look into the vial compartment when the instrument is connected to power. R...

  • Page 13: -Beta Draft-

    Figure 3 main reading screen 1 time and date 6 up/down navigation arrows 2 comments 1 7 menu icons 3 message display 8 options (contextual: send data, compare log, setup, import, edit) icon 4 value and unit 9 read icon (contextual: read, options, select, delete) 5 heading bar (contextual: shows the ...

  • Page 14: Startup

    Table 2 icon description (continued) icon description configures the instrument settings. Shows instrument-specific data about the instrument: firmware information, instrument backup, instrument updates, signaling information and factory service data. Shows help and more information. If there is mor...

  • Page 15: -Beta Draft-

    Add an operator id 1. Push login>options. 2. Select an option. Option description new adds a unique operator id. Edit changes an existing operator id and assigns a security level to an operator id. Delete removes an existing operator id. Initialize rfid tag enables an operator id tag (optional) to b...

  • Page 16: -Beta Draft-

    Option description compare log shows the data log from the online turbidimeter to compare results with the laboratory turbidimeter. Shows the data log management options: send data to a job list, view data as a graph, remove data and send data to a connected usb flash drive, pc or printer. Send data...

  • Page 17: -Beta Draft-

    Put the vial into the instrument c a u t i o n personal injury hazard. Never remove covers from the instrument. This is a laser-based instrument and the user risks injury if exposed to the laser. C a u t i o n personal injury hazard. Do not look into the vial compartment when the instrument is conne...

  • Page 18: -Beta Draft-

    Do a measurement only do the steps that follow if there is a sample vial in the vial compartment and the instrument lid closed. Refer to put the vial into the instrument on page 15. 1. The measurement starts automatically when the lid is closed. If not, push read to start the measurement or select a...

  • Page 19: -Beta Draft-

    Import a sample id list make a sample id list to import to the instrument via usb flash drive. 1. On an external device, open a spreadsheet program. 2. Make three columns that identify the sequential sample id number, the sample id name and the sampling date (optional). Make sure that all headings a...

  • Page 20: -Beta Draft-

    Prepare for a verification or calibration n o t i c e do not to touch or scratch the glass of the sample vial. Contamination or scratches on the glass can cause measurement errors. The measurement surface must stay clean and have no scratches. Use a no-lint cloth to remove dirt, fingerprints or part...

  • Page 21: Maintenance

    The instrument is factory calibrated. Do calibration procedures as necessary. Do the steps that follow to do a full range calibration procedure. 1. Clean the vial, then carefully invert the vial. Do not cause air bubbles. Refer to prepare for a verification or calibration on page 18. 2. For vials wi...

  • Page 22: Troubleshooting

    N o t i c e do not disassemble the instrument for maintenance. If the internal components must be cleaned or repaired, contact the manufacturer. Clean spills c a u t i o n chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and national regulations. 1. Obey a...

  • Page 23: -Beta Draft-

    Problem possible cause solution the standard value is too high. The wrong standard is in the vial compartment. Check the calibration standard and do a calibration on page 18. There are bubbles in the standard. ???What is the solution??? There is dirt in the vial compartment. Clean the vial compartme...

  • Page 24: -Beta Draft-

    Problem possible cause solution selected documents do not show on the display. The document is not a pdf file. Make sure that the document is a pdf file. The network connection does not show. There is no connection to the dhcp server. Contact the network administrator. If necessary, contact technica...

  • Page 25: -Beta Draft-

    Accessories description item no. Micro fiber cloth (for vial cleaning) lzy945 secondary standards, lzy901 secondary standards, 1 to 4 ntu lzy902 vial wiper for manual cleaning lzy903 vial compartment wiper for manual cleaning lzy910 english 23 beta -beta draft-.

  • Page 26: -Beta Draft-

    24 english beta -beta draft-.

  • Page 27: -Beta Draft-

    Beta -beta draft-.

  • Page 28: -Beta Draft-

    Hach company world headquarters p.O. Box 389, loveland, co 80539-0389 u.S.A. Tel. (970) 669-3050 (800) 227-4224 (u.S.A. Only) fax (970) 669-2932 orders@hach.Com www.Hach.Com hach lange gmbh willstätterstraße 11 d-40549 düsseldorf, germany tel. +49 (0) 2 11 52 88-320 fax +49 (0) 2 11 52 88-210 info@h...