Haier DW-86L288 User Manual

Other manuals for DW-86L288: Service Manual, User Manual
Manual is about: Ultra-Low Temperature Refrigerator

Summary of DW-86L288

  • Page 1

    0070505464 read this user's manual carefully before using the ult freezer. Place this manual in a convenient location for future reference. Haier medical and laboratory co., ltd. This product is suitable for ultra low temperature storage of products in applications such as clinical, pharmaceutical, ...

  • Page 2

    Table of contents g technical data eneral information pre-start up instructions maintenance features of the haier ult freezer .........................................................1 description of key parts and functions..................................................2 safety precautions..........

  • Page 3

    24 your haier ult freezer has a 3 year warranty on the compressors and 5 years parts and labor on the rest of the unit. Equipment and instrument service center. The warranty card should be kept for future service work. Should you need service or to inquire about your freezer, please use the warranty...

  • Page 4

    2 door handle side panel for refrigeration compartment front protection panel for refrigeration compartment reserved window for temperature recorder diagram of parts outer door inner door description of key parts and functions control panel power switch for battery control panel door handle lock pro...

  • Page 5

    Optional accessories this recorder has been calibrated at the factory. It retains the temperature accuracy and calibration even when there is a power supply interruption. Should there be a need to recalibrate the recorder, please follow the procedures below. A)let the unit run continuously until it ...

  • Page 6

    The appliance requires 220v~/50hz for 628/388/288/386/310 or 220v~/50hz and 220v~/60hz for 628/626.A regulator with a capacity of 4000w or higher must be installed if the voltage if beyond the rated voltag 8/626(115v~/60hz ),and the extended length must not exceed 3m. Haier haier e of ±10%.If the po...

  • Page 7

    20 1.Temperature recorder the ult freezer is equipped with a 6-inch temperature recorder for the chamber temperature. When the unit is powered up, the recorder starts working. For proper operation of the temperature recorder, please follow the procedures below. 1.Open the recorder door to access the...

  • Page 8

    6 do not touch any electrical components or power cord if your hands are damp or wet. Such action might lead to an electric shock. After restarting your unit after a power outage or shutdown, ensure that all settings are correct. Accidental change in setting may damage the stored products. Do not pl...

  • Page 9

    18 specifications description model description exterior wall material interior wall material inner doors shelves porthole for testing insulation compressors evaporator condenser refrigerant temperature controller temperature display temperature sensor alarm device remote alarm terminals battery acc...

  • Page 10

    8 installation environment 1.Ambient temperature: the ult is designed to operate between 5 to 28 .The maximum ambient temperature is 32 . For best results, the operating ambient should be in between 18 to 25 . If necessary, use an air-conditioning system to achieve the required ambient condition. 2....

  • Page 11

    16 should there be any malfunctions in the system, please attempt to answer the following questions before notifying maintenance or calling equipment & instrument service center haier freezer does not start up. Freezer's refrigeration ability is reduced. The unit is noisy. Is the power supply normal...

  • Page 12

    10 warning your ult unit should be cared for by dedicated personnel. Every 2 to 4 hours, the unit's working status should be inspected and recorded on a daily basis. Should there be a malfunction in the unit, the freezer temperature rises. If the problems cannot be corrected in a short time, please ...

  • Page 13

    14 notes: a flashing alarm light cannot be cancelled unless the root cause for the alarm is eliminated. The buzzing alarm can be temporarily silenced for 30 minutes by pressing the alarm silence button. However, if the problem is not fixed, the buzzer alarm will resume after 30 minutes. The freezer ...

  • Page 14

    12 temperature alarm set point can only be set to a value of at least 5 colder than the set point temperature of the unit. It cannot be set to the lowest temperature limit. 5.Setting the security code: when the unit is first setup to operate, the security code to unlock the control is 06. After unlo...