Replacing the battery
If, after normal recharging, the battery cannot sustain a full treatment
session, recharge again, as it is probably approaching the end of its useful
life. If even after a second recharge the battery still cannot complete a
treatment session, follow the steps here below:
•
Contact the manufacture or authorized dealer for original spare
part;
•
Open the battery compartment on the back side of MAG1000,
disconnect the red/black wire from connector and remove the
exhausted battery;
•
Insert new battery and connect the red/black wire to the plug;
•
Close battery compartment
WARNING: for disposal exhausted battery follow the instructions on
chapter “Disposal”. Do not open or burn battery. Do not short circuit poles.
Keep battery away from sparks or naked flames. In the event of internal
electrolyte coming into contact with skin or garments, wash immediately
with water. In the event of electrolyte coming into contact with eyes, rinse
thoroughly and seek medical attention.
WARNING: It is suggested to recharge the device once a month in case
of prolonged inactivity. This is for a correct battery maintenance.
Maintenance
Cleaning
Clean the equipment from the dust using a dry soft cloth.
Resistant stains can be removed using a sponge soaked in solution of
water and alcohol (20%).