Icare tonometer one User And Maintenance Manual

Summary of tonometer one

  • Page 1

    U ser ’s an d m ai n ten an ce m an ual english.

  • Page 2

    Icare tonometer one manual  english 2 www.Icaretonometer.Com icare® one (model: ta02) instruction manual v1.12 01/14 en 0044 this device complies with: medical device directive 93/42/eec canadian medical device regulations copyright © 2012 icare finland oy made in finland icare finland oy Äyritie 2...

  • Page 3

    Icare tonometer one manual  english 3 www.Icaretonometer.Com warning! Health care professionals must inform patients not to modify or discontinue their treatment plan without receiving instructions from the health care professional. Warning! Do not change your medication or dosage without consultin...

  • Page 4

    Icare tonometer one manual  english 4 www.Icaretonometer.Com usage contraindications the ta02 tonometer is contraindicated in the presence of the following conditions: corneal scarring, microphtalmos, buphthalmos, nystagmus, keratoconus, abnormal central corneal thickness, and topical anesthetics. ...

  • Page 5

    Icare tonometer one manual  english 5 www.Icaretonometer.Com turning the device on press the measurement button to turn the tonometer on. All leds are displayed briefly. One of the lower iop indicator leds remains lit for 2 seconds while other leds switch off. The led indicates the battery voltage....

  • Page 6

    Icare tonometer one manual  english 6 www.Icaretonometer.Com accuracy of display: display is divided into 11 ranges: 5-7 mmhg 7-10 mmhg 10-14 mmhg 14-18 mmhg 18-21 mmhg 21-24 mmhg 24-27 mmhg 27-30 mmhg 30-35 mmhg 35-40 mmhg 40-50 mmhg setting up the tonometer before taking measurements before takin...

  • Page 7

    Icare tonometer one manual  english 7 www.Icaretonometer.Com tonometer forehead and cheek supports with adjustment wheels an optional eye support can be used for easy alignment and additional stability during measurement. When the eye support is used, the cheek support is not used and needs to be a...

  • Page 8

    Icare tonometer one manual  english 8 www.Icaretonometer.Com inserting probe into probe base 3. After the probe has been inserted, be careful not to point the tonometer downward to prevent the probe from falling out. 4. Press the measurement button once to activate the inserted probe. During activa...

  • Page 9

    Icare tonometer one manual  english 9 www.Icaretonometer.Com in normal mode a value is shown briefly after each successful measurement. When the six measurements in sequence are completed and the measurement result is valid, the final value range is displayed. If there is too much variation between...

  • Page 10

    Icare tonometer one manual  english 10 www.Icaretonometer.Com a new measurement needs to be taken if: repeat error led is displayed, for example, if the probe made contact with the eyelid. Encountering unusual values, for example, over 22 mmhg or below 8 mmhg. Error messages and indicators icare to...

  • Page 11

    Icare tonometer one manual  english 11 www.Icaretonometer.Com cleaning and disinfection the forehead and cheek supports and the eye support must be cleaned for each new patient with disinfectant for example 70% alcohol solution using wipe. Do not immerse the tonometer in water or other liquid. If t...

  • Page 12

    Icare tonometer one manual  english 12 www.Icaretonometer.Com inserting new batteries service procedures replace the batteries when the device indicates low battery voltage. Clean or change the probe base if the probe does not move smoothly (clean led is lit). Note no other service procedures can b...

  • Page 13

    Icare tonometer one manual  english 13 www.Icaretonometer.Com operation environment temperature: +10 °c to +35 °c relative humidity: 30 % to 90 % atmospheric pressure: 800 hpa – 1060 hpa storage environment temperature: -10 °c to +55 °c relative humidity: 10 % to 95 % atmospheric pressure: 700 hpa ...

  • Page 14

    Icare tonometer one manual  english 14 www.Icaretonometer.Com symbols caution bf-type device single-use disposable serial number manufacturing date lot number sterilized using irradiation stand by use by do not discard this product with other household-type waste. Send to appropriate facility for r...