Icare tonometer one User And Maintenance Manual - page 12
Icare tonometer ONE manual English
12
www.icaretonometer.com
Inserting new batteries
SERVICE PROCEDURES
Replace the batteries when the device indicates low battery voltage.
Clean or change the probe base if the probe does not move smoothly (CLEAN LED is lit).
NOTE
No other service procedures can be carried out by the user. All other service and repair must be carried out by the manufacturer or a certified
service center.
SPARE PARTS
Single-use probes (10 included with the tonometer when purchased)
Probe base kit
Eye supports
Mode switch key
USB cable
Batteries
carry on case
These can be purchased from a local distributor of Icare products.
TECHNICAL AND PERFORMANCE DATA
Type TA02
The device conforms to CE regulations
Dimensions: approximately 11 cm x 8 cm x 3 cm
Weight: approximately 150 g
Power supply: 2•CR123 batteries (make sure you use batteries with built-in PTC protection, For Example Energizer Lithium Photo
123 3V CR123A)
Measurement range: 5-50 mmHg
Display range: 5-50 mmHg
Accuracy (95% tolerance interval relatively to manometry): ± 1.8 mmHg
Repeatability (coefficient of variation):
< 9% @ IOP ≤20 mmHg
< 3% @ IOP >20 mmHg
Accuracy of display: display is divided into 11 ranges:
5-7 mmHg
7-10 mmHg
10-14 mmHg
14-18 mmHg
18-21 mmHg
21-24 mmHg
24-27 mmHg
27-30 mmHg
30-35 mmHg
35-40 mmHg
40-50 mmHg
Display unit: mmHg
The serial number is located on the inside of the battery compartment cover
There are no electrical connections from the tonometer to the patient
The device has BF-type electric shock protection