Jbc TI-A Instruction Manual - page 3
of
4
3
1. Pull out the sensor
2. Adjust at an ambient temperature
with slotted screw driver.
3. Wiring like the drawing below.
4. Apply a voltage of 23.65mV and verify the temperature shown in the thermometer is 350°C (662°F)
5. Take note of the deviation to correct future measurements.
6. Insert the sensor ST-A.
NOTES: - Thermometer can be adjusted through the potentiometer you will find inside the unit.
- The calibration date is on the label on the bottom
Digital
mVolt Meter
TI-A
Variable
power supply
ST-A
CALIBRATION OF TI-A
2
+
-
-
-
+
+
1