Jelrus Infinity L-30 Operator's Manual

Manual is about: Burnout Furnaces

Summary of Infinity L-30

  • Page 1

    I n f i n i t y b u r n o u t f u r n a c e s tm operator’s manual a6732 iso 13485:1996 iso 9001:2000 air techniques inc. Hicksville, ny.

  • Page 2: Table Of Contents

    1 introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 . Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 safety ....

  • Page 3: Warranty

    Thank .You .For .Purchasing .An .Infinity burnout furnace . . . We .Have .Designed .And .Manufactured .This .Furnace .Using .The .Latest .In .Microcomputer .Technology .To .Give .You .Many . Years .Of .Dependable .Service . .The .Controls .On .Your .New .Infinity .Are .Different .From .Those .You .M...

  • Page 4: Specifications

    Specifications infinity m-30 infinity l-30 electrical 100v . 50/60hz . 1150w 115v . 50/60hz . 1150w 220/230v . 50/60hz . 1150w 240v . 50/60hz . 1150w 100v . 50/60hz . 1610w 115v . 50/60hz . 1610w 220/230v . 50/60hz . 1610w 240v . 50/60hz . 1610w capacity .8-1 .3/4” .Rings .Or .2-3” .Rings 15-1 .3/4”...

  • Page 5

    Display description 1 . . Night .Time . . . Press .To .Start .A .Delayed .Program . .The .Number .Entered .Is .The .Time . . . (delay .Start) . . . Required .For .The .Program .To .Be .Completed .And .Ready .To .Cast . . . . . . . . . . . When .A .Program .Is .Running, .Press .To .Display .The .Time...

  • Page 6

    5 figure 1 infinity m-30 and l-30 identification 13 10 11 9 8 7 6 5 12 4 3 2 1 front panel heating chamber vent tube key parts identification and explanation.

  • Page 7

    Key parts identification and explanation lower back panel (figure 2) display description 1 . . Power .Cord . A . .On .100v .And .115v .Furnaces, .The .Ac .Power .Cord .Is .Permanently . . . . Connected .To .The .Unit . . . B . . On .220/230v .And .240v .Furnaces, .The .Ac .Power .Cord .Is .Detachabl...

  • Page 8: Installation

    Unpack and set-up installation 7 1 . . Unpack .The .Contents .Of .The .Box . .Remove .The .Following .Materials: . A . .Vent .Tube .- .Remove .From .Bubble .Wrap .And .Insert .In .Hole .On .Top .Of .Furnace . . B . .Calibration .Table .Kit .- .Remove .Plastic .Bag .Containing .Two .Tablets .And .Sav...

  • Page 9: Installation

    Installation 8 when .A .Program .Is .Completed, .20 ."beeps" .Sound .Every .15 .Minutes .To .Remind .The .Operator .That .The . Material .Is .Ready .To .Cast . 1 . . .Be .Sure .The .Infinity .Is .In .The .Idle .Mode .- .No .Program .Us .Running . 2 . . .Press . . . .And .(while .Holding) .Press .Pro...

  • Page 10: Operation

    Program and operate - one stage program .(figure .1) 9 operation 1 . . Turn .The .Power .Switch .On . 2 . . Press .Program .Number . . .Use . . . . . . . .To .Display .The .Desired .Program .Number .(p1 .- .P30) . 3 . . Press .Enter ./ .Review .To .Select .The .Displayed .Program .Number . .Stage .1...

  • Page 11: Operation

    Operation program and operate - two stage program (figure .1) 10 1 . .Follow .One .Stage .Program, .Steps .1 .- .6 . 2 . .Press .Enter ./ .Review . . .Stage .1 .Light .Turns .Off, .Stage .2 .And .Heat .Rate .Lights .Turn .On . . .Main . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ....

  • Page 12: Operation

    Operation program and operate - three stage program (figure .1) 11 1 . .Follow .Two .Stage .Program, .Steps .1 .- .4 . 2 . .Press .Enter ./ .Review . . .Stage .2 .Light .Turns .Off, .Stage .3 .And .Heat .Rate .Lights .Turn .On . . .Main . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ....

  • Page 13: Operation

    Operation program and operate - four stage program (figure .1) 12 1 . .Follow .Three .Stage .Program, .Steps .1 .- .4 . 2 . .Press .Enter/review . . .Stage .3 .Light .Turns .Off, .Stage .4 .And .Heat .Rate .Lights .Turn .On . . .Main . . . .Display .Shows .Four .Heat .Rate .Choices . . .Choose .One:...

  • Page 14: Operation

    Operation review a program (figure .1) 13 1 . . .Turn .The .Power .Switch .On . 2 . . .Press .Program .Number . 3 . . .Press . . . . . . . .To .Select .The .Program .Number .To .Be .Reviewed . . .The .Program .Number .(p1 .- .P30) . . . . .Will .Appear .On .The .Main .Display . 4 . . The . Number . ...

  • Page 15: Operation

    Operation edit while a program is running (figure .1) 14 1 . . .To .Identify .Which .Program .Is .Running, .Press .Program .Number . Note: the program number cannot be changed using while the program is running. 2 . . .To .Determine .The .Time .Remaining .For .The .Completion .Of .The .Program, .Pre...

  • Page 16: Operation

    Night time delay start operation sample one stage program (to .Be .Completed .And .Ready .To .Cast .In .24 .Hours) . Heat rate temp hold time stage 1 stage 2 stage 3 stage 4 24:00 10 °f .(6°c) 1600 °f .(871°c) 1:00 no no no * .* .* * .* .* * .Though .Any .Number .May .Appear, .The .Unit .Is .Deactiv...

  • Page 17: Operation

    To .Start .Immediately, .Press .Start ./ .Stop . Follow .The .Same .Programming .Procedures .If .A .Three .Or .Four .Stage .Program .Is .Desired . Press .Night .Time .(delay .Start) .To .Start .An .Overnight .Burnout . . .(see .Sample .On .Stage .Program .To .Pro- gram .Night .Time .(delay .Start) ....

  • Page 18: Error Codes

    Error codes note: . .“beeps” .Occur .When .The .Error .Code .Appears .On .The .Main .Display error code description probable cause invalid .Entry . Error: . .Stage, . Heat .Rate .And . Temp .Lights .Flash occurs . When . The . Heat .Rate . Is . Set . To . Cool . But . The . Temp . Of . That . Stage ...

  • Page 19: Error Codes

    Error codes error code description probable cause shorted . Thermocouple . Or .Defective . . Heater .Plates occurs .If .The .Thermocouple .Wires .Are .Shorted .Or .The .Heater .Plates .Are . Defective .Causing .Infinity .To .Achieve .A .Heat .Rate .Of .Less .Than .6% .Of .The . Maximum .Attainable ....

  • Page 20: Service

    19 caution: the infinity should be serviced only by qualified service technicians. Be sure to unplug the power cord and wait for the furnace to cool before performing any service oper- ation. For help with operating or servicing your jelrus equipment, please call jelrus any time between 9:00am and 5...

  • Page 21: Service

    20 1 . . If .A .Power .Failure .Occurs, .The .Infinity .Memorizes .The .Conditions .Prior .To .The .Loss .Of .Power . . .When .The .Power . . . . . Returns, .The .Infinity .Returns .To .The .Proper .Point .In .The .Program . . 2 . . When .Power .Is .Returned, .The .Hr:min .Light .Flashes .Indicating...

  • Page 22: Field Service

    1 . . Remove .The .Back .Panel .Of .The .Furnace . 2 . . Remove .The .Nuts .Which .Hold .The .Heating .Plate .Wires .To .The .Power .Terminals .And .Straighten .The .Heating . . Plate .Wires .(figure .4) . 3 . . Open .The .Furnace .Door .And .Find .The .Two .Ceramic .Sections .At .The .Front .Of .Th...

  • Page 23

    Infinity m-30 (all voltages) infinity l-30 100, 115 volt infinity l-30 220/230, 240 volt 22 figure 4 infinity with upper rear panel removed heating plate wire power terminal heater (power) leads rear insulating panel thermocouple thermocouple clamp ceramic insulating bushing insulating filler rear i...

  • Page 24: Field Service

    1 . . Remove .The .Front .Panel .(figure .5) . 2 . . Note .And .Record .The .Color .And .Location .Of .Each .Wire .On .The .Thermocouple .Terminals .Located .On .The .Main . Pc .Board . .Remove .Both .Wires . 3 . . Remove .The .Calibration .Jack .Connector .(j1) .From .The .Main .Pc .Board .By .Pull...

  • Page 25

    Figure 5a other components 24 figure 5 infinity with front panel / main pc board removed.

  • Page 26: Field Service

    1 . . Remove .The .Bottom .Front .Panel .And .The .Upper .And .Lower .Rear .Panels .Of .The .Furnace . .(note .The .Position .Of . . . The .Vent .Louvers .On .The .Rear .Panels .) 2 . . Note .And .Record .The .Color .And .Location .Of .The .Wires .On .The .Thermocouple .Terminals .Located .On .The ....

  • Page 27: Field Service

    Field service replacement of the triac (pn 15475) 26 1 . . Remove .The .Lower .Rear .Panel .(figure .2) . 2 . . Note .And .Record .The .Color .And .Location .Of .Each .Of .The .Three .Wires .On .The .Triac .Terminals . (figure .5a) . .Remove .Each .Of .The .Three .Wires .By .Pulling .Straight .Up .O...

  • Page 28: Spare Parts List

    27 spare parts list description 100v 115v 220/230v 240v power .Cord .Kit-japan .& .U .S . . 15288 . 15288 . N/a . N/a power .Cord .Kit-australia . N/a . N/a . 23206 . 23206 power .Cord .Kit-england . N/a . N/a . 23202 . 23202 power .Cord .Kit-europe . N/a . N/a . 23203 . 23203 power .Cord .Kit-italy...

  • Page 29: Spare Parts List

    28 spare parts list description 100v 115v 220/230v 240v power .Cord .Kit-japan .& .U .S . . 15289 . 15289 . N/a . N/a power .Cord .Kit-australia . N/a . N/a . 23206 . 23206 power .Cord .Kit-england . N/a . N/a . 23202 . 23202 power .Cord .Kit-europe . N/a . N/a . 23203 . 23203 power .Cord .Kit-italy...

  • Page 30: Program Card Forms

    Program card forms stage 1 stage 2 stage 3 stage 4 heat temp hold rate time program # metal(s) stage 1 stage 2 stage 3 stage 4 heat temp hold rate time program # metal(s) stage 1 stage 2 stage 3 stage 4 heat temp hold rate time program # metal(s) stage 1 stage 2 stage 3 stage 4 heat temp hold rate t...

  • Page 31: Program Card Forms

    Program card forms stage 1 stage 2 stage 3 stage 4 heat temp hold rate time program # metal(s) stage 1 stage 2 stage 3 stage 4 heat temp hold rate time program # metal(s) stage 1 stage 2 stage 3 stage 4 heat temp hold rate time program # metal(s) stage 1 stage 2 stage 3 stage 4 heat temp hold rate t...

  • Page 32

    Pn .15813 .Rev . .E infinity .Is .A .Trademark .Of .Air .Techniques .Inc . ©2004 .Air .Techniques .Inc . Jelrus international, .A .Division .Of .Air .Techniques .Inc ., .Has .Been .A .Leading .Manufacturer .Of dental .Laboratory .Equipment .Essential .To .The .Fabrication .Of .Porcelain .Caps, .Crow...