Keithley Series 2200 Reference Manual

Other manuals for Series 2200: User Manual, Operation Manual
Manual is about: Programmable DC Power Supplies

Summary of Series 2200

  • Page 1

    Reference manual 2200s-901-01 rev. C / march 2016 *p2200s90101c* 2200s-901-01c www.Keithley.Com series 2200 programmable dc power supplies a greater measure of confidence.

  • Page 2: Reference Manual

    Programmable dc power supplies reference manual © 2016, keithley instruments cleveland, ohio, u.S.A. All rights reserved. Any unauthorized reproduction, photocopy, or use the information herein, in whole or in part, without the prior written approval of keithley instruments is strictly prohibited. A...

  • Page 3: Safety Precautions

    Safety precautions the following safety precautions should be observed before using this product and any associated instrumentation. Although some instruments and accessories would normally be used with nonhazardous voltages, there are situations where hazardous conditions may be present. This produ...

  • Page 4

    For safety, instruments and accessories must be used in accordance with the operating instructions. If the instruments or accessories are used in a manner not specified in the operating instructions, the protection provided by the equipment may be impaired. Do not exceed the maximum signal levels of...

  • Page 5: Table Of Contents

    Introduction ............................................................................................................... 1-1 welcome .............................................................................................................................. 1-1 extended warranty .................

  • Page 6

    Table of contents series 2200 programmable dc power supplies reference manual command and query structure .................................................................................................. 3-4 command entry ................................................................................

  • Page 7

    Series 2200 programmable dc power supplies reference manual table of contents [source:]output[:state] ............................................................................................... 4-20 [source:]output:timer:delay ........................................................................

  • Page 8

    Table of contents series 2200 programmable dc power supplies reference manual next steps ................................................................................................................. 6-1 additional series 2200 information .............................................................

  • Page 9: Welcome

    In this section: welcome .................................................................................. 1-1 extended warranty ................................................................... 1-1 contact information .................................................................. 1-1 cd-rom ...

  • Page 10: Cd-Rom Contents

    Section 1: introduction series 2200 programmable dc power supplies reference manual 1-2 2200s-901-01 rev. C / march 2016 cd-rom contents a cd-rom is shipped with each series 2200 order. The series 2200 product information cd-rom contains: • reference manual: includes advanced operation topics and ma...

  • Page 11: Optional Accessories

    Series 2200 programmable dc power supplies reference manual section 1: introduction 2200s-901-01 rev. C / march 2016 1-3 optional accessories optional accessory keithley part number ieee-488.2 interface board for the pci bus kpci-488lpa usb cable type a to b 1m (3.2 ft) usb-b-1 usb cable type a to b...

  • Page 12

    Section 1: introduction series 2200 programmable dc power supplies reference manual 1-4 2200s-901-01 rev. C / march 2016 electrical ratings for the power connection model line selector switch setting frequency fuse rating max power 2200-20-5 110 v / 220 v 50 / 60 hz for 110 v: 5 a th, 250 v for 220 ...

  • Page 13: Operating Requirements

    In this section: operating requirements ........................................................... 2-1 installing the system ................................................................. 2-3 operating basics ...................................................................... 2-5 operation ex...

  • Page 14

    Section 2: general operation series 2200 programmable dc power supplies reference manual 2-2 2200s-901-01 rev. C / march 2016 figure 1: place bottom of instrument on flat stable surface 2. Before operating, ensure that the ambient temperature is between +0 °c and +40 °c (+32 °f to +104 °f). To ensur...

  • Page 15: Installing The System

    Series 2200 programmable dc power supplies reference manual section 2: general operation 2200s-901-01 rev. C / march 2016 2-3 installing the system this section contains information on how to install your model 2200 power supply. • unpack the instrument and check that you received all the items list...

  • Page 16

    Section 2: general operation series 2200 programmable dc power supplies reference manual 2-4 2200s-901-01 rev. C / march 2016 voltage output check to verify basic voltage functions without a load, follow these steps. 1. Remove all leads from the output connectors. 2. Turn on the power supply. 3. Pre...

  • Page 17: Operating Basics

    Series 2200 programmable dc power supplies reference manual section 2: general operation 2200s-901-01 rev. C / march 2016 2-5 operating basics front panel at a glance the front panel of the keithley instruments series 2200 contains the following items: • power switch • display • arrow keys • the nav...

  • Page 18

    Section 2: general operation series 2200 programmable dc power supplies reference manual 2-6 2200s-901-01 rev. C / march 2016 (3) arrow keys up, down, right, and left arrow keys (▲ and ▼) and enter key. (4) navigation wheel rotate the navigation wheel to increase or decrease digits or to select menu...

  • Page 19

    Series 2200 programmable dc power supplies reference manual section 2: general operation 2200s-901-01 rev. C / march 2016 2-7 key descriptions the keys provide front panel control and configuration. See front panel at a glance (on page 2-5) for an illustration of each key location. The following tab...

  • Page 20

    Section 2: general operation series 2200 programmable dc power supplies reference manual 2-8 2200s-901-01 rev. C / march 2016 keypad special operations see the following table for the series 2200 keypad special operations. Operation key press display the instrument firmware version. Shift + 8 displa...

  • Page 21

    Series 2200 programmable dc power supplies reference manual section 2: general operation 2200s-901-01 rev. C / march 2016 2-9 configuration menus description use the configuration menus to configure (setup) the instrument. Sometimes the configuration menus are called the settings menus. See the foll...

  • Page 22

    Section 2: general operation series 2200 programmable dc power supplies reference manual 2-10 2200s-901-01 rev. C / march 2016 primary menu secondary menu first option additional option(s) allows you to ... The instrument will advance to each step only after it receives a trigger signal. Repeat allo...

  • Page 23

    Series 2200 programmable dc power supplies reference manual section 2: general operation 2200s-901-01 rev. C / march 2016 2-11 primary menu secondary menu first option additional option(s) allows you to ... Immediate set rear panel port mode to immediate trigger. >ri mode off turn remote inhibit (ri...

  • Page 24

    Section 2: general operation series 2200 programmable dc power supplies reference manual 2-12 2200s-901-01 rev. C / march 2016 operating conditions messages description off power supply output is off. Cv instrument is operating in constant voltage (cv) mode. Cc instrument is operation in constant cu...

  • Page 25

    Series 2200 programmable dc power supplies reference manual section 2: general operation 2200s-901-01 rev. C / march 2016 2-13 rear panel at a glance the rear panel of the series 2200 contains the following items: • cooling vents • factory test port • usb device port • gpib ieee-488 port • 110 v/220...

  • Page 26

    Section 2: general operation series 2200 programmable dc power supplies reference manual 2-14 2200s-901-01 rev. C / march 2016 (2) factory test port do not use the factory test port. Unauthorized use of the factory test port could damage this product. (3) usb device port the usb 2.0 high-speed devic...

  • Page 27

    Series 2200 programmable dc power supplies reference manual section 2: general operation 2200s-901-01 rev. C / march 2016 2-15 110 v/220 v selector switch the 110 v/220 v selector switch is located on the bottom of the instrument as shown in the following figure. Figure 5: selector switch front pane...

  • Page 28

    Section 2: general operation series 2200 programmable dc power supplies reference manual 2-16 2200s-901-01 rev. C / march 2016 configuring the instrument for your application the configuration menus (see configuration menus description (on page 2-9)) include settings, like ovp and max volt, that det...

  • Page 29

    Series 2200 programmable dc power supplies reference manual section 2: general operation 2200s-901-01 rev. C / march 2016 2-17 initializing to the default setup use the default setup to get the power supply into a default initial state: 1. Remove all leads from the output connectors. 2. Turn on the ...

  • Page 30

    Section 2: general operation series 2200 programmable dc power supplies reference manual 2-18 2200s-901-01 rev. C / march 2016 save and recall setups you can store up to 40 different setups in setup memory locations (1 to 40). Each setup includes the set voltage limit, the set current limit, and the...

  • Page 31

    Series 2200 programmable dc power supplies reference manual section 2: general operation 2200s-901-01 rev. C / march 2016 2-19 setting the overvoltage protection the ovp set control turns overvoltage protection (ovp) on or off. If overvoltage protection is turned on and the instrument senses a volta...

  • Page 32

    Section 2: general operation series 2200 programmable dc power supplies reference manual 2-20 2200s-901-01 rev. C / march 2016 recall the power supply on/off output state at power on the output recall control determines the on or off output state after the power supply is powered on. If you select o...

  • Page 33

    Series 2200 programmable dc power supplies reference manual section 2: general operation 2200s-901-01 rev. C / march 2016 2-21 recall the power supply operating parameters at power on the save last control determines whether the power supply saves its most recent settings, such as voltage and curren...

  • Page 34

    Section 2: general operation series 2200 programmable dc power supplies reference manual 2-22 2200s-901-01 rev. C / march 2016 lock the navigation wheel the knob lock control can be used to lock the navigation wheel so it cannot be used to change settings or to select menu items. To lock the navigat...

  • Page 35

    Series 2200 programmable dc power supplies reference manual section 2: general operation 2200s-901-01 rev. C / march 2016 2-23 using remote sense use remote sensing to regulate the output voltage at the device under test. This feature lets you compensate for the voltage drop in the leads between the...

  • Page 36

    Section 2: general operation series 2200 programmable dc power supplies reference manual 2-24 2200s-901-01 rev. C / march 2016 define a list of voltage and current steps the edit list feature lets you create up to seven sequences of steps, each with a voltage level, current level, and time duration....

  • Page 37

    Series 2200 programmable dc power supplies reference manual section 2: general operation 2200s-901-01 rev. C / march 2016 2-25 7. Press enter. The instrument displays: repeat 1 8. Use the keypad or navigation wheel to select the number of times to repeat the list. In this example, 2 is used. This me...

  • Page 38

    Section 2: general operation series 2200 programmable dc power supplies reference manual 2-26 2200s-901-01 rev. C / march 2016 run a list of voltage and current steps to run a defined list of voltage and current steps, do the following: 1. Set the output voltage to what you want it to be set at befo...

  • Page 39

    Series 2200 programmable dc power supplies reference manual section 2: general operation 2200s-901-01 rev. C / march 2016 2-27 figure 9: gpib connector to allow many gpib connections to other instruments, stack the cable connectors. Two screws are located on each connector to ensure that connections...

  • Page 40

    Section 2: general operation series 2200 programmable dc power supplies reference manual 2-28 2200s-901-01 rev. C / march 2016 to connect the series 2200 to the ieee-488 bus, line up the cable connector with the connector located on the rear panel. Install and tighten the screws securely, making sur...

  • Page 41: Operation Examples

    Series 2200 programmable dc power supplies reference manual section 2: general operation 2200s-901-01 rev. C / march 2016 2-29 operation examples you can save tests in 40 memory locations to save up to 40 setups or use the series 2200 list mode to define custom test sequences of up to 80 steps. This...

  • Page 42

    Section 2: general operation series 2200 programmable dc power supplies reference manual 2-30 2200s-901-01 rev. C / march 2016 timer example this example sets up the power supply to output 6 v, 1 a for 7 seconds before the output automatically turns off. Keystrokes perform the following procedure us...

  • Page 43

    Series 2200 programmable dc power supplies reference manual section 2: general operation 2200s-901-01 rev. C / march 2016 2-31 step number voltage level (v) current level (a) duration (sec) 1 2 5 0.001 2 5 5 0.001 3 1 1 1.000 4 7 3 0.001 5 3 1 0.001 keystrokes use the following procedures to configu...

  • Page 44

    Section 2: general operation series 2200 programmable dc power supplies reference manual 2-32 2200s-901-01 rev. C / march 2016 to configure the third step in the list: 1. Input s 003= 1.000 v. Press enter. The display shows s 003= 0.0000a. 2. Input s 003= 1.000 a as the current limit. Press enter. T...

  • Page 45: Default Setup

    Series 2200 programmable dc power supplies reference manual section 2: general operation 2200s-901-01 rev. C / march 2016 2-33 programming commands to reset the power supply: *rst to set the trigger source to be internal: trigger:source:bus to configure the list: source:list:step 5 source:list:mode ...

  • Page 46

    Section 2: general operation series 2200 programmable dc power supplies reference manual 2-34 2200s-901-01 rev. C / march 2016 clear and sanitize procedures this section helps customers with data security concerns to sanitize or remove memory devices from the series 2200 instrument. These products h...

  • Page 47

    Series 2200 programmable dc power supplies reference manual section 2: general operation 2200s-901-01 rev. C / march 2016 2-35 memory devices the following tables list nonvolatile and volatile memory devices. Detailed procedures to sanitize or clear these devices are shown after the tables. Nonvolat...

  • Page 48

    Section 2: general operation series 2200 programmable dc power supplies reference manual 2-36 2200s-901-01 rev. C / march 2016 data export devices the following table lists the data export device. There are no procedures to disable these devices. Data export device type and minimum size function use...

  • Page 49: Section Overview

    In this section: section overview ...................................................................... 3-1 communication interfaces ........................................................ 3-1 command syntax ..................................................................... 3-3 command groups ....

  • Page 50

    Section 3: remote operation series 2200 programmable dc power supplies reference manual 3-2 2200s-901-01 rev. C / march 2016 using the gpib connector for ieee-488 (gpib) operation. Use a shielded cable, such as the keithley instruments model 7007-1 or model 7007-2. Connect a gpib cable from the seri...

  • Page 51: Command Syntax

    Series 2200 programmable dc power supplies reference manual section 3: remote operation 2200s-901-01 rev. C / march 2016 3-3 command syntax you can control the power supply through the gpib or usb interface using commands and queries. This section describes the syntax these commands and queries use ...

  • Page 52

    Section 3: remote operation series 2200 programmable dc power supplies reference manual 3-4 2200s-901-01 rev. C / march 2016 command and query structure commands consist of set commands and query commands (usually simply called commands and queries). Commands change power supply settings or perform ...

  • Page 53

    Series 2200 programmable dc power supplies reference manual section 3: remote operation 2200s-901-01 rev. C / march 2016 3-5 commands commands cause the power supply to perform a specific function or change one of its settings. Commands have the structure: [:][[]...] a command header is made up of o...

  • Page 54

    Section 3: remote operation series 2200 programmable dc power supplies reference manual 3-6 2200s-901-01 rev. C / march 2016 scpi commands and queries the power supply uses a command language based on the standard commands for programmable instruments (scpi) standard. The scpi standard was created b...

  • Page 55

    Series 2200 programmable dc power supplies reference manual section 3: remote operation 2200s-901-01 rev. C / march 2016 3-7 parameter types many power supply commands require parameters. Parameters are indicated by angle brackets, such as . There are several different types of parameters, as listed...

  • Page 56

    Section 3: remote operation series 2200 programmable dc power supplies reference manual 3-8 2200s-901-01 rev. C / march 2016 chaining commands and queries you can chain several commands or queries together into a single message. To create a chained message, first create a command or query, then add ...

  • Page 57: Command Groups

    Series 2200 programmable dc power supplies reference manual section 3: remote operation 2200s-901-01 rev. C / march 2016 3-9 • you can use upper case, lower case, or a mixture of both. No embedded spaces are allowed between or within nodes. :source:frequency 10mhz is the same as :source:frequency 10...

  • Page 58: Command Reference

    In this section: commands listed in alphabetical order .................................... 4-2 *cls (no query form) ............................................................... 4-2 configure:sound[:state] .................................................... 4-3 *ese ................................

  • Page 59: *cls (No Query Form)

    Section 4: command reference series 2200 programmable dc power supplies reference manual 4-2 2200s-901-01 rev. C / march 2016 system:key ......................................................................... 4-35 system:local (no query form) ............................................ 4-36 syste...

  • Page 61: *esr? (Query Only)

    Section 4: command reference series 2200 programmable dc power supplies reference manual 4-4 2200s-901-01 rev. C / march 2016 *esr? (query only) returns the contents of the standard event status register (sesr). *esr? Also clears the sesr (since reading the sesr clears it) see appendix c, registers ...

  • Page 62

    Series 2200 programmable dc power supplies reference manual section 4: command reference 2200s-901-01 rev. C / march 2016 4-5 fetch:voltage[:dc]? (query only) this command returns the last measured output voltage stored in the communications buffer of the power supply. A new measurement is not initi...

  • Page 63: *idn? (Query Only)

    Section 4: command reference series 2200 programmable dc power supplies reference manual 4-6 2200s-901-01 rev. C / march 2016 *idn? (query only) returns the power supply identification code in ieee 488.2 notation. Group status syntax *idn? Returns a string that includes , , , and as defined in the f...

  • Page 64: *opc

    Series 2200 programmable dc power supplies reference manual section 4: command reference 2200s-901-01 rev. C / march 2016 4-7 measure:voltage[:dc]? (query only) this command initiates and executes a new voltage measurement, and returns the measured output voltage of the power supply. Group measureme...

  • Page 65: *psc

    Section 4: command reference series 2200 programmable dc power supplies reference manual 4-8 2200s-901-01 rev. C / march 2016 *psc sets and queries the power-on status flag that controls the automatic power-on execution of srer and eser. When *psc is true, the srer and eser are set to 0 at power-on....

  • Page 66: *rcl (No Query Form)

    Series 2200 programmable dc power supplies reference manual section 4: command reference 2200s-901-01 rev. C / march 2016 4-9 *rcl (no query form) restores the state of the power supply from a copy of its settings stored in the setup memory. The settings are stored using the *sav command. If the spe...

  • Page 67: [Source:]Current[:level]

    Section 4: command reference series 2200 programmable dc power supplies reference manual 4-10 2200s-901-01 rev. C / march 2016 saves the state of the power supply into a specified nonvolatile memory location. Any settings that had been stored previously at the location are overwritten. You can later...

  • Page 68: [Source:]Digital:data

    Series 2200 programmable dc power supplies reference manual section 4: command reference 2200s-901-01 rev. C / march 2016 4-11 [source:]digital:data this command sets the output state of the rear-panel ttl control output and queries the state of the rear-panel ttl control input. When the port mode i...

  • Page 69: [Source:]Digital:function

    Section 4: command reference series 2200 programmable dc power supplies reference manual 4-12 2200s-901-01 rev. C / march 2016 [source:]digital:function this command sets or queries the function of the ttl control lines on the rear panel of the power supply. *rst value is trigger. Group source synta...

  • Page 70: [Source:]Function:mode

    Series 2200 programmable dc power supplies reference manual section 4: command reference 2200s-901-01 rev. C / march 2016 4-13 [source:]function:mode this command can be in either fixed mode or list mode. When this command is in fixed mode, the power supply responds to discrete commands. When this c...

  • Page 71

    Section 4: command reference series 2200 programmable dc power supplies reference manual 4-14 2200s-901-01 rev. C / march 2016 returns is the number of times the active list is set to execute. Examples list:count 6 would set the active list to execute 6 times before stopping. [source:]list:current[:...

  • Page 73: [Source:]List:step

    Section 4: command reference series 2200 programmable dc power supplies reference manual 4-16 2200s-901-01 rev. C / march 2016 [source:]list:save (no query form) this command saves the active list file to a storage location in non- volatile memory. Group source syntax [source:]list:save related comm...

  • Page 74: [Source:]List:width

    Series 2200 programmable dc power supplies reference manual section 4: command reference 2200s-901-01 rev. C / march 2016 4-17 [source:]list:voltage[:level] this command sets the voltage level of a specified step in a list in units of v or mv. Group source: list syntax [source:]list:voltage[:level] ...

  • Page 75: [Source:]Output:dfi:source

    Section 4: command reference series 2200 programmable dc power supplies reference manual 4-18 2200s-901-01 rev. C / march 2016 [source:]output:dfi:source this command associates the dfi ttl output on the rear panel with a specified bit in the status byte register (sbr). Once the bit is associated wi...

  • Page 76: [Source:]Output:pon[:state]

    Series 2200 programmable dc power supplies reference manual section 4: command reference 2200s-901-01 rev. C / march 2016 4-19 [source:]output:pon[:state] this command configures the power supply to power up with its output turned off, or to return the output to the state it was in when it powered d...

  • Page 77: [Source:]Output:ri:mode

    Section 4: command reference series 2200 programmable dc power supplies reference manual 4-20 2200s-901-01 rev. C / march 2016 [source:]output:ri:mode this command sets the input mode of the ri (remote inhibit) input pin. In order for this command to be effective, the rear panel ttl input and output...

  • Page 78: [Source:]Output:timer:delay

    Series 2200 programmable dc power supplies reference manual section 4: command reference 2200s-901-01 rev. C / march 2016 4-21 examples output on output? Might return 0, which would indicate that the output is off. [source:]output:timer:delay this command sets the time duration of the output timer. ...

  • Page 79

    Section 4: command reference series 2200 programmable dc power supplies reference manual 4-22 2200s-901-01 rev. C / march 2016 [source:]output:timer[:state] this command turns the output timer function on and off. When the timer is activated and a duration is set, the output of the power supply will...

  • Page 82

    Series 2200 programmable dc power supplies reference manual section 4: command reference 2200s-901-01 rev. C / march 2016 4-25 [source:]voltage:protection:state this command activates, deactivates, or checks the status of overvoltage protection (ovp). Group source syntax [source:]voltage:protection:...

  • Page 83: [Source:]Voltage:range

    Section 4: command reference series 2200 programmable dc power supplies reference manual 4-26 2200s-901-01 rev. C / march 2016 [source:]voltage:range this command limits the maximum voltage that can be programmed on the power supply. This command corresponds to the front-panel max voltage setting th...

  • Page 84: *sre

    Series 2200 programmable dc power supplies reference manual section 4: command reference 2200s-901-01 rev. C / march 2016 4-27 *sre service request enable (sre) sets and queries the bits in the service request enable register (srer). See appendix c, registers (on page c-3) for more information about...

  • Page 85: Status:operation:enable

    Section 4: command reference series 2200 programmable dc power supplies reference manual 4-28 2200s-901-01 rev. C / march 2016 srer have the binary value 00100000. Status:operation:enable this command sets and queries the contents of the operation enable register (oenr). The oenr is an eight-bit mas...

  • Page 86

    Series 2200 programmable dc power supplies reference manual section 4: command reference 2200s-901-01 rev. C / march 2016 4-29 status:operation[:event]? (query only) this command returns the contents of the operation event register (oevr). After executing this command the operation event register is...

  • Page 87: Status:questionable:enable

    Section 4: command reference series 2200 programmable dc power supplies reference manual 4-30 2200s-901-01 rev. C / march 2016 status:questionable:enable this command sets and queries the contents of the questionable enable register (qenr). The qenr is an eight-bit mask register that determines whic...

  • Page 88

    Series 2200 programmable dc power supplies reference manual section 4: command reference 2200s-901-01 rev. C / march 2016 4-31 status:questionable[:event]? (query only) this command returns the contents of the questionable event register (qevr). After executing this command, the quest event register...

  • Page 89

    Section 4: command reference series 2200 programmable dc power supplies reference manual 4-32 2200s-901-01 rev. C / march 2016 status:questionable:ntransition this command sets the negative transition filter of the questionable event register. The filter contents cause the corresponding bit in the q...

  • Page 90

    Series 2200 programmable dc power supplies reference manual section 4: command reference 2200s-901-01 rev. C / march 2016 4-33 status:questionable:ptransition this command sets the positive transition filter of the questionable event register. The filter contents cause the corresponding bit in the q...

  • Page 91: *stb? (Query Only)

    Section 4: command reference series 2200 programmable dc power supplies reference manual 4-34 2200s-901-01 rev. C / march 2016 *stb? (query only) the byte query returns the contents of the status byte register (sbr) using the master summary status (mss) bit. Refer to appendix c, registers (on page c...

  • Page 92: System:key

    Series 2200 programmable dc power supplies reference manual section 4: command reference 2200s-901-01 rev. C / march 2016 4-35 system:key this command can produce the same effect as pressing one of the front-panel buttons. The instrument must be in local mode in order for this command to simulate a ...

  • Page 93: System:posetup

    Section 4: command reference series 2200 programmable dc power supplies reference manual 4-36 2200s-901-01 rev. C / march 2016 system:local (no query form) this command sets the power supply for control from the front- panel. Group system syntax system:local related commands system:local (no query f...

  • Page 94

    Series 2200 programmable dc power supplies reference manual section 4: command reference 2200s-901-01 rev. C / march 2016 4-37 system:remote (no query form) this command sets the power supply to remote control mode. Group system syntax system:remote related commands system:local (no query form) (on ...

  • Page 95: *trg (No Query Form)

    Section 4: command reference series 2200 programmable dc power supplies reference manual 4-38 2200s-901-01 rev. C / march 2016 *trg (no query form) this command generates a trigger event. Group trigger syntax *trg related commands trigger[:immediate] (no query form) (on page 4-38) examples *trg trig...

  • Page 97: *wai (No Query Form)

    Section 4: command reference series 2200 programmable dc power supplies reference manual 4-40 2200s-901-01 rev. C / march 2016 *wai (no query form) this command prevents the instrument from executing further commands or queries until all pending commands are complete. Group synchronization syntax *w...

  • Page 98

    Series 2200 programmable dc power supplies reference manual section 4: command reference 2200s-901-01 rev. C / march 2016 4-41 example 1 this example is written in the c programming language, the virtual instrument software architecture (visa) standard can be used. It demonstrates basic communicatio...

  • Page 99

    Section 4: command reference series 2200 programmable dc power supplies reference manual 4-42 2200s-901-01 rev. C / march 2016 reading checkerror("unable to read voltage"); errorstatus = viprintf(pws4000,"measure:current?"); //measure the output current checkerror("unable to write the device"); er...

  • Page 100

    Series 2200 programmable dc power supplies reference manual section 4: command reference 2200s-901-01 rev. C / march 2016 4-43 { clock_t goal; goal = wait + clock(); while(goal > clock()); } example 2 this example shows a command sequence that configures a three-step list to execute once in continuo...

  • Page 101

    Section 4: command reference series 2200 programmable dc power supplies reference manual 4-44 2200s-901-01 rev. C / march 2016 example 4 this example is written in the c programming language; ni visa can be used. The program demonstrates setting up and executing a list. It shows how to use the opera...

  • Page 102

    Series 2200 programmable dc power supplies reference manual section 4: command reference 2200s-901-01 rev. C / march 2016 4-45 sendscpi("voltage 0v"); //set the voltage to 0v sendscpi("output 1"); // turn output on sendscpi("trigger:source bus"); //set trigger source to bus /* configure the list*/ s...

  • Page 103

    Section 4: command reference series 2200 programmable dc power supplies reference manual 4-46 2200s-901-01 rev. C / march 2016 strcat(commandstring, ""); errorstatus = viprintf(pws4000, commandstring); checkerror("can't write to power supply"); pdest = strchr(commandstring, '?'); //search for quer...

  • Page 104: Troubleshooting Guide

    In this section: what to do if the power supply does not turn on ...................... 5-1 what to do if the power supply does not turn on to solve problems you might encounter when turning on the instrument, follow these steps: 1. Verify that there is ac power to the power supply. First, check tha...

  • Page 105: Next Steps

    In this section: additional series 2200 information ........................................... 6-1 additional series 2200 information this manual has prepared you to start using your new series 2200 instrument. For additional information on the series 2200, refer to: • the product information cd-ro...

  • Page 106: Cleaning

    In this appendix: cleaning ................................................................................... A-1 cleaning inspect the power supply as often as operating conditions require. To clean the exterior surface, perform the following steps: 1. Remove loose dust on the outside of the power ...

  • Page 107: Performance Verification

    In this appendix: performance verification ........................................................... B-1 performance verification procedures ...................................... B-11 performance verification this topic contains performance verification procedures. Additional test equipment is re...

  • Page 108

    Appendix b: calibration series 2200 programmable dc power supplies reference manual b-2 2200s-901-01 rev. C / march 2016 test record model serial procedure performed by date test passed failed self test dc voltage accuracy with remote sense dc voltage accuracy without remote sense dc voltage readbac...

  • Page 109

    Series 2200 programmable dc power supplies reference manual appendix b: calibration 2200s-901-01 rev. C / march 2016 b-3 dc voltage accuracy with remote sense instrument dut voltage test current min measured max 0% test voltage 2200-20-5 0 v 2.5 a -0.0030 v __________ 0.0030 v 2200-30-5 0 v 2.5 a -0...

  • Page 110

    Appendix b: calibration series 2200 programmable dc power supplies reference manual b-4 2200s-901-01 rev. C / march 2016 dc voltage accuracy without remote sense instrument dut voltage test current min measured max 0% test voltage 2200-20-5 0 v 2.5 a -0.0030 v __________ 0.0030 v 2200-30-5 0 v 2.5 a...

  • Page 111

    Series 2200 programmable dc power supplies reference manual appendix b: calibration 2200s-901-01 rev. C / march 2016 b-5 dc voltage readback accuracy instrument dut voltage test current measured voltage voltage readout absolute difference maximum difference 0% test voltage 2200-20-5 0 v 2.5 a ______...

  • Page 112

    Appendix b: calibration series 2200 programmable dc power supplies reference manual b-6 2200s-901-01 rev. C / march 2016 dc voltage line regulation instrument min line max line nom - min max - nom largest max value 2200-20-5 __________ __________ __________ __________ __________ 0.0080 v 2200-30-5 _...

  • Page 113

    Series 2200 programmable dc power supplies reference manual appendix b: calibration 2200s-901-01 rev. C / march 2016 b-7 dc current accuracy instrument test voltage dut current min measured current max 0% test current 2200-20-5 10 v 0.0 a -0.002 a __________ 0.002 a 2200-30-5 15 v 0.0 a -0.0026 a __...

  • Page 114

    Appendix b: calibration series 2200 programmable dc power supplies reference manual b-8 2200s-901-01 rev. C / march 2016 dc current readback accuracy instrument test current measured current current readout absolute difference maximum difference 0% test current 2200-20-5 0 a ________ ________ ______...

  • Page 115

    Series 2200 programmable dc power supplies reference manual appendix b: calibration 2200s-901-01 rev. C / march 2016 b-9 dc current line regulation instrument test voltage min line max line nom line nom - min max - nom largest max value 2200-20-5 10 v ______ ________ ________ ________ ________ _____...

  • Page 116

    Appendix b: calibration series 2200 programmable dc power supplies reference manual b-10 2200s-901-01 rev. C / march 2016 voltage noise at 7 mhz instrument voltage test load r 1 measured rms maximum rms measured pk - pk maximum pk - pk 2200-20-5 5 Ω ________ 1 mv ________ 3 mv 2200-30-5 10 Ω _______...

  • Page 117

    Series 2200 programmable dc power supplies reference manual appendix b: calibration 2200s-901-01 rev. C / march 2016 b-11 performance verification procedures use the procedures in this topic to verify the performance of the series 2200 instruments. Performance verification conditions the following c...

  • Page 118

    Appendix b: calibration series 2200 programmable dc power supplies reference manual b-12 2200s-901-01 rev. C / march 2016 equipment required these procedures use external equipment to directly check warranted characteristics. The following table lists the required equipment..

  • Page 119

    Series 2200 programmable dc power supplies reference manual appendix b: calibration 2200s-901-01 rev. C / march 2016 b-13 test equipment item connectors minimum requirements example 1. Connectors connectors for hook up wire assemblies and test resistors, item numbers 5, 6, 9, 10, and 11. • (qty. 6) ...

  • Page 120

    Appendix b: calibration series 2200 programmable dc power supplies reference manual b-14 2200s-901-01 rev. C / march 2016 item connectors minimum requirements example 2200-20-5 5 Ω hl10006e5r000jj 2200-30-5 10 Ω hl10006e10r00jj 2200-32-3 12 Ω hl10006e12r00jj 2200-60-2 40 Ω hl10006e40r00jj 2200-72-1 ...

  • Page 121

    Series 2200 programmable dc power supplies reference manual appendix b: calibration 2200s-901-01 rev. C / march 2016 b-15 4. If any of the messages listed in the following table appear in the display, it indicates that the eeprom is experiencing the condition associated with the message. 5. If there...

  • Page 122

    Appendix b: calibration series 2200 programmable dc power supplies reference manual b-16 2200s-901-01 rev. C / march 2016 figure 18: dc voltage accuracy with remote sense test setup.

  • Page 123

    Series 2200 programmable dc power supplies reference manual appendix b: calibration 2200s-901-01 rev. C / march 2016 b-17 4. Power on the dut. Ensure the warm-up criteria has been met as described in the performance verification conditions. 5. Set the voltmeter as follows: a. Set to measure dc volts...

  • Page 124

    Appendix b: calibration series 2200 programmable dc power supplies reference manual b-18 2200s-901-01 rev. C / march 2016 check dc voltage accuracy without remote sense and voltage readback accuracy equipment required (item 2) ac power source (item 5) high current hook up wire (bold line connections...

  • Page 125

    Series 2200 programmable dc power supplies reference manual appendix b: calibration 2200s-901-01 rev. C / march 2016 b-19 figure 20: voltage accuracy regulation and protection test setup 2. Set the voltmeter as follows: a. Set to measure dc volts. B. Set to auto range. C. Verify that the math mx+b f...

  • Page 126

    Appendix b: calibration series 2200 programmable dc power supplies reference manual b-20 2200s-901-01 rev. C / march 2016 check dc voltage line regulation this check procedure uses the same test setup as the previous procedure. (see check dc voltage accuracy without remote sense and voltage readback...

  • Page 127

    Series 2200 programmable dc power supplies reference manual appendix b: calibration 2200s-901-01 rev. C / march 2016 b-21 10. Change the ac power source output to match the voltage selector switch setting of the dut 65as shown in the following table. Dut voltage selector switch ac power source volta...

  • Page 128

    Appendix b: calibration series 2200 programmable dc power supplies reference manual b-22 2200s-901-01 rev. C / march 2016 check dc voltage load regulation this check procedure uses the same test setup as the previous procedure. (see check dc voltage accuracy without remote sense and voltage readback...

  • Page 129

    Series 2200 programmable dc power supplies reference manual appendix b: calibration 2200s-901-01 rev. C / march 2016 b-23 12. Enter the largest of the three values calculated in step 11 into the largest column in the dc voltage load regulation (on page b-6) table..

  • Page 130

    Appendix b: calibration series 2200 programmable dc power supplies reference manual b-24 2200s-901-01 rev. C / march 2016 check dc overvoltage detection this check procedure uses the same test setup as the previous procedure. (see check dc voltage accuracy without remote sense and voltage readback a...

  • Page 131

    Series 2200 programmable dc power supplies reference manual appendix b: calibration 2200s-901-01 rev. C / march 2016 b-25 8. Enter the voltmeter reading (where ovp triggering occurred) into the dc overvoltage detection (on page b-6) table for checking dc overvoltage detection. 9. Reset ovp. A. Press...

  • Page 132

    Appendix b: calibration series 2200 programmable dc power supplies reference manual b-26 2200s-901-01 rev. C / march 2016 check dc current and dc current readback accuracy equipment required (item 2) ac power source (item 5) high current hook up wire (bold line connections) (item 3) electronic load ...

  • Page 133

    Series 2200 programmable dc power supplies reference manual appendix b: calibration 2200s-901-01 rev. C / march 2016 b-27 2. Set the voltmeter as follows: a. Set to measure dc volts. B. Set to auto range. C. Set to show amps (instead of volts) by multiplying the voltmeter result by 20. • press shift...

  • Page 134

    Appendix b: calibration series 2200 programmable dc power supplies reference manual b-28 2200s-901-01 rev. C / march 2016 check dc current line regulation this check procedure uses the same test setup as the previous procedure (see check dc current and dc current readback accuracy). 1. Change the ac...

  • Page 135

    Series 2200 programmable dc power supplies reference manual appendix b: calibration 2200s-901-01 rev. C / march 2016 b-29 8. Change the ac power source output to the voltage specified in the following table. Dut voltage selector switch ac power source voltage 110 115 v 220 230 v 9. Enter the voltmet...

  • Page 136

    Appendix b: calibration series 2200 programmable dc power supplies reference manual b-30 2200s-901-01 rev. C / march 2016 check dc current load regulation this check procedure uses the same test setup as the previous procedure (see check dc current line regulation (on page b-28)). 1. Set the voltmet...

  • Page 137

    Series 2200 programmable dc power supplies reference manual appendix b: calibration 2200s-901-01 rev. C / march 2016 b-31 instrument minimum test voltage 2200 - 20 - 5 18.0 v 2200 - 30 - 5 27.0 v 2200 - 32 - 3 28.8 v 2200 - 60 - 2 54.0 v 2200 - 72 - 1 64.8 v 9. Enter the voltmeter reading into the d...

  • Page 138

    Appendix b: calibration series 2200 programmable dc power supplies reference manual b-32 2200s-901-01 rev. C / march 2016 check voltage noise (7 mhz) equipment required (item 5) high current hook up wire (bold line connections) (item 12) oscilloscope (item 10) load resistor (item 13) oscilloscope 1x...

  • Page 139

    Series 2200 programmable dc power supplies reference manual appendix b: calibration 2200s-901-01 rev. C / march 2016 b-33 5. Set the oscilloscope as follows: a. 1 mv/division b. 1 m Ω input resistance c. 20 mhz bandwidth (bw) limit d. Ac coupled e. Auto-trigger f. 1 ms/div g. Set to measure vp-p and...

  • Page 140

    Appendix b: calibration series 2200 programmable dc power supplies reference manual b-34 2200s-901-01 rev. C / march 2016 check voltage noise (20 mhz) equipment required (item 5) high current hook up wire (bold line connections) (item 15) coaxial cable (bnc m-m) (item 10) load resistor (item 16) bnc...

  • Page 141

    Series 2200 programmable dc power supplies reference manual appendix b: calibration 2200s-901-01 rev. C / march 2016 b-35 5. Set the oscilloscope as follows: a. 1 mv/division b. 1 m Ω input resistance c. 20 mhz bandwidth (bw) limit d. Ac coupled e. Auto-trigger f. 1 ms/div g. Set to measure vp-p and...

  • Page 142

    Appendix b: calibration series 2200 programmable dc power supplies reference manual b-36 2200s-901-01 rev. C / march 2016 check current noise equipment required (item 5) high current hook up wire (bold line connections) (item 12) oscilloscope (item 11) 1 Ω load resistor (item 14) oscilloscope 10x pr...

  • Page 143

    Series 2200 programmable dc power supplies reference manual appendix b: calibration 2200s-901-01 rev. C / march 2016 b-37 5. Set the oscilloscope as follows: a. 1 mv/division b. 1 m Ω input resistance c. 20 mhz bandwidth (bw) limit d. Ac coupled e. Auto-trigger f. 1 ms/div g. Set to measure vp-p and...

  • Page 144: Status And Events

    In this appendix: status and events .................................................................... C-1 status reporting structure ........................................................ C-1 registers ................................................................................. C-3 status ...

  • Page 145

    Appendix c: status model series 2200 programmable dc power supplies reference manual c-2 2200s-901-01 rev. C / march 2016 figure 25: error and event handling process.

  • Page 146: Registers

    Series 2200 programmable dc power supplies reference manual appendix c: status model 2200s-901-01 rev. C / march 2016 c-3 registers the registers in the event reporting system fall into two functional groups: • status registers contain information about the status of the power supply. They include t...

  • Page 147

    Appendix c: status model series 2200 programmable dc power supplies reference manual c-4 2200s-901-01 rev. C / march 2016 sbr bit functions bit function 7 (msb) osb operation status bit. Indicates that an operation event has occurred. 6 rqs request service. Obtained from a serial poll. Shows that th...

  • Page 148

    Series 2200 programmable dc power supplies reference manual appendix c: status model 2200s-901-01 rev. C / march 2016 c-5 operation condition register (ocr) the operation condition register is made up of eight bits, which note the occurrence of events as shown here. Figure 28: operation condition re...

  • Page 149

    Appendix c: status model series 2200 programmable dc power supplies reference manual c-6 2200s-901-01 rev. C / march 2016 sesr bit functions bit function 7 (msb) ———— this bit is not used. 6 ———— this bit is not used. 5 ———— this bit is not used. 4 ps protection shutdown. 3 ri remote inhibit. This b...

  • Page 150

    Series 2200 programmable dc power supplies reference manual appendix c: status model 2200s-901-01 rev. C / march 2016 c-7 enable registers this topic describes the following four types of enable registers: • event status enable register (eser) (on page c-7) • service request enable register (srer) (...

  • Page 151

    Appendix c: status model series 2200 programmable dc power supplies reference manual c-8 2200s-901-01 rev. C / march 2016 operation enable register (oenr) the oenr consists of bits defined exactly the same as bits 0 through 7 in the oevr. You can use this register to control whether or not the opera...

  • Page 152

    Series 2200 programmable dc power supplies reference manual appendix c: status model 2200s-901-01 rev. C / march 2016 c-9 error/event queue the event queue stores detailed information on up to 32 events. When 32 events stack up in the event queue, the 32nd event is replaced by event code 350, "queue...

  • Page 153

    Appendix c: status model series 2200 programmable dc power supplies reference manual c-10 2200s-901-01 rev. C / march 2016 command errors the following table shows the command error messages generated by improper syntax. Check that the command is properly formed and that it follows the rules in the ...

  • Page 154

    Series 2200 programmable dc power supplies reference manual appendix c: status model 2200s-901-01 rev. C / march 2016 c-11 system errors the following table lists the system errors that can occur during power supply operation. System error messages (dde bit 3) code message –310 system error –350 too...

  • Page 155

    Appendix c: status model series 2200 programmable dc power supplies reference manual c-12 2200s-901-01 rev. C / march 2016 device dependent errors the following table lists the device errors that can occur during power supply operation. These errors may indicate that the power supply needs repair. D...

  • Page 156: Display Character Codes

    In this appendix: ascii code chart ..................................................................... D-2 appendix d display character codes.

  • Page 157: Ascii Code Chart

    Appendix d: display character codes series 2200 programmable dc power supplies reference manual d-2 2200s-901-01 rev. C / march 2016 ascii code chart figure 34: ascii code chart.

  • Page 158: Index

    A abbreviating commands, queries, and parameters • 3-7 accessories • 1-2 optional • 1-3 standard • 1-2 arrow keys • 2-6 b bnf notation • 3-3 c calibration • b-1 cc message • 2-11 chaining commands and queries • 3-8 cleaning • a-1 command and query structure • 3-4 commands abbreviating • 3-7 alphabet...

  • Page 159

    Index series 2200 programmable dc power supplies reference manual 2 2200s-901-01 rev. C / march 2016 n navigation wheel • 2-6 lock the navigation wheel • 2-22 numeric keypad • 2-6 special operations • 2-8 unlock • 2-8 o on/off key • 2-6, 2-7 optional accessories • 1-3 p parameter types • 3-7 perform...

  • Page 160

    Specifications are subject to change without notice. All keithley trademarks and trade names are the property of keithley instruments. All other trademarks and trade names are the property of their respective companies. Keithley instruments corporate headquarters • 28775 aurora road • cleveland, ohi...