Width Model
Description
Downflow
Velocity
(fpm)
Out-
Flow
Exhaust
Volume
(CFM)
Noise
Pressure
db(A)
Average
Outflow
Face
Velocity
(fpm)
2'
2' Filtered PCR Enclosure
30
40
50
60
120
165
210
250
53-57
58-60
60-61
62-63
85
115
145
170
3'
3' Filtered PCR Enclosure
30
40
50
60
180
250
310
370
58-60
60-62
61-63
63-54
85
115
145
170
4'
4' Filtered PCR Enclosure
30
40
50
60
240
330
420
500
60-62
62-64
66-69
68-70
85
115
145
170
Electrical Requirements
Standard duplex electrical receptacles should be nearby for
connecting the Filtered PCR Enclosure, or other equipment. The
enclosure includes iris pass-throughs to allow electrical cords
through the back of the enclosure without leaving a large hole.
Space Requirements
The dimensions for all the different models are shown in Appendix