Labconco 3970202 User Manual - page 33
28
Chapter 6: Maintaining Your Filtered PCR Enclosure
when downflow velocity drops below the recommended 30-65 fpm
velocity and the speed control is adjusted to full speed. Replace
HEPA filter if it fails the HEPA Filter Leak Test in Chapter 6.
How to Install a New HEPA Filter
NOTE: Only a qualified certifier should service the HEPA filter.
After the HEPA filter is replaced, the enclosure MUST be
certified. See Figure 6-1.
1. Unplug the enclosure and remove the prefilter located on
top.
2. Remove the front panel by loosening the two screws that
secure it.
3. Using a 9/16" deep socket, loosen the filter clamp bolts
located on top. Refer to Figure 6-1.
4. With the clamp bolts loosened, the HEPA filter-clamping
frame should be clear of filter. Carefully pull the filter
straight out of the enclosure and discard properly.
5. Cover the surface of the new HEPA filter gasket with a
light coating of silicone grease, if desired.
6. Install the new HEPA filter by pushing it straight into the
cabinet, ensuring that it is correctly oriented with the gasket
facing down on the exhaust side of the enclosure.
7. Tighten the clamp bolts uniformly until the filter gasket is
properly compressed against the frame. Inspect the seal
thoroughly before proceeding.
8. Reinstall the front panel and prefilter.
9. Plug the enclosure in and have it check for leaks and
downflow velocity.
CAUTION: The filter clamp bolts should only be tightened
enough to ensure a proper seal at maximum tightness. The
filter gasket should be compressed 50% or less.