OBSERVATION & USE OF MICROSCOPE (continued)
10.3
IPD Adjustment
This sdjustment should be performedevery time the
observer is changed since theinterpupillary distance differs
between individuals. Adjust the interpupillary distance so
that the view field for each eye is merged into one. Move
while holding each sleeve with both hands.
10.4
Focusing of microscope
Fig. 11
I
J
1.
Switch on the desired illumination as per specimen
selected.
2.
Adjust both eye pieces at 0 diopter reading.
3.
Adjust IPD to user convinience.
4.
Place any specimen on Black & White plate or on DF/BF
base as per configuration provided.
5.
Setup the magnification factor at highest magnification of
Zoom pod using knob (I).
6.
Looking through the eyepieces, adjust the focusing of
microscope head using coarse knob (J) to focus the
specimen.
7.
Bring the image into critical focus.
8.
Change magnification factor from highest to lowest.
9.
At this stage image should remain in focus. This setting
is called as par-focal of microscope.
10. Microscope is ready to use.
10.5
Focusing torque adjustment
Luxeo 6z Microscope is pre-adjusted to a focusing torque of
20 kg with standard Binocular or Trinocular. Still some user
may feel uncomfortable while using additional accessories.
Torque of focusing mechanism can be adjusted as per
following procedure per user's convenience:
1. Hold firmly the right knob shown as (K). Refer Fig. 12.
2. Rotate left knob (L) to adjust torque of the focusing
mechanism of microscope.
3. By rotating knob (L) to clock wise will increase torque
and by rotating anticlockwise will decrease torque.
K
L
Fig. 10
Fig. 12