Maintenance
Even if everything has been conceived so that to be forgotten, your water purifier requires a minimal care.
Here is the regular maintenance that RO400C requires so to produce a high quality water:
Information below concern a water purifier that is fed with softened water.
Ø
To clean the grid :
At every filter replacing (1 or 2).
Ø
To replace filter 1 :
At least every 6 months or when the water purifier requires it (see “Trouble-
shooting table”).
Ø
To replace filter 2 :
At least every 6 months, or every 2000 functioning hours, or when the water
purifier requires it (see “Trouble-shooting table”).
Ø
Preventive descaling :
Every time you replace filter 2.
Ø
Curative descaling :
If the purified water’s capacity decreases till being half of the original
capacity, even if the preventive descaling has been regularly carried out.
Ø
To replace the membrane :
If the purified water’s capacity decreases till being half of the original
capacity, even if the curative descaling has been regularly carried out.
Qualified technician only.
Ø
Preparation to storage and protection against freezing: If the water purifier will not be used for a long time
(for ex. annual closing). To repeat it every 9 months in case of long non-use.
To clean the grid
The grid ( a coarse filter) is conceived to intercept large particles; it’s located inside the union where you connect the
RO400C’s feed water hose. If the LEDs on the control panel show that the intake pressure is too low, the reason may be
that the grid is clogged.
1. Disconect the appliance from the electrical supply.
2. Unscrew the flexible feed water hose, on the back of the
appliance.
3. Use suitable pliers to grip and remove the grid (picture A).
4. Clean the grid. Then put it back inside the connector.
5. Reconnect the feed water hose to the appliance.
6. Connect the RO400C to the electrical supply.
The water purifier will flush automatically.