LaserLinc TLAser2120 Operator's Manual

Manual is about: Laser Scan Micrometers

Summary of TLAser2120

  • Page 1

    Tlaserx30 / x60 / x120 laser scan micrometers (130 / 160 / 1120 / 230 / 260 / 2120) operator’s manual april 2010 laserlinc, inc. 777 zapata drive fairborn, ohio 45324 telephone: 937.318.2440 fax: 937.318.2445.

  • Page 2

    Copyright laserlinc, inc. Copyrights this manual with all rights reserved. Under the copyright laws, this manual may not be reproduced in any form, in whole or part, without prior written consent of laserlinc, inc. Disclaimer laserlinc, inc. Has reviewed this manual thoroughly in order that it will ...

  • Page 3

    2 laser micrometers – theory of operation .....................................................3 laser micrometers – safety considerations .................................................4 tlaser130™ / 160™ / 1120™ laser micrometer ....................................................5 specification...

  • Page 4

    3 laser micrometers – theory of operation laser scan micrometers use a laser light source to measure diameter, gap, position, width, and other dimensions. The light source is focused onto the facets of a rotating mirror that reflects the laser beam into a collimating lens. The lens refracts the beam...

  • Page 5

    4 laser micrometers – safety considerations be sure to avoid direct exposure of human eyes to laser beams emitted from laser diodes. Even though barely visible and/or invisible to the human eye, they can be quite harmful. In particular, avoid looking directly into a laser diode or collimated beam al...

  • Page 6

    5 tlaser130™ / 160™ / 1120™ laser micrometer getting started • be gentle with your scanner, it is a precision optical instrument. Rough treatment can reduce the precision, and the life expectancy of your scanner. Specifications 130 / 160 / 1120 mounting tlaser130 / 160 / 2120 • mounting hole drawing...

  • Page 7

    6 cabling 130 / 160 / 1120 typical connectors 130 / 160 / 1120 • power and data – both are keyed, and cannot be connected incorrectly. Connectors (plugs), if oriented correctly, will slide snugly into sockets. To remove, grasp metal plug housing and pull directly away from scanner. Plugs should slid...

  • Page 8

    7 tlaser230™ / 260™ / 2120™ laser micrometer getting started • be gentle with your scanner, it is a precision optical instrument. Rough treatment can reduce the precision, and the life expectancy of your scanner. Specifications 230 / 260 / 2120 power is received from ac cord plugged into 110-220vac ...

  • Page 9

    8 cabling 230 / 260 / 2120 typical connecters 230 / 260 / 2120 • power and data – both are keyed, and cannot be connected incorrectly. Connectors (plugs), if oriented correctly, will slide snugly into sockets. To remove, grasp metal plug housing and pull directly away from scanner. Plugs should slid...

  • Page 10

    9 preventive maintenance & troubleshooting tlaserx30 / x60 / x120 cleaning and routine maintenance • cleaning – the windows of the tlaserx30/x60/x120 must always be clean to ensure the highest measurement accuracy. To clean the windows: 1. Remove the air purge or air purge if applicable. 2. Inspect ...

  • Page 11

    10 • problem: total vu measurement reports “too many: (any number > 8)” • possible causes: 1. There is dirt on a window. Remove the air purge and inspect for spots, dots, or glowing flecks on any of the windows. Follow the window cleaning procedure described earlier in this document. Check each axis...