10
•
Problem:
Total Vu measurement reports “Too Many: (any number > 8)”
•
Possible Causes:
1. There is dirt on a window. Remove the air purge and inspect for spots, dots, or glowing flecks
on any of the windows. Follow the window cleaning procedure described earlier in this
document. Check each axis, via the test port, for correct waveform.
2. There is more than one part in the field. Check for lint attached to the air purge. Check for any
additional fibers in the scan field. Check each axis, via the test port, for correct waveform.
•
Problem:
Total Vu measurement reports “No Scan”
•
Possible Causes:
1. The scanner is not powered. Check that the power cord is completely plugged in at both ends,
and that it is not plugged into a dead circuit (e.g. a power strip that is turned off).
2. The scanner data cable is not connected. Check that the scanner data cable is plugged into the
scanner, and into the TLAser400 in the PC.
•
Problem:
There are many problems that are not related to the scanner.
•
Possible Causes:
1. Please contact LaserLinc for further help.