Care and Maintenance
Error Codes
1. Disconnect the cuff tubing from the monitor prior to storing.
2. Only use a soft, dry cloth to clean your blood pressure monitor.
3. Avoid using any types of liquids on the monitor or cuff.
4. Do not store the unit where it will be exposed to direct sunlight, dust or
humidity.
5. Avoid extreme temperatures.
6. Never disassemble the monitor or cuff.
7. Dropping or subjecting your blood pressure monitor to strong shocks
should be avoided.
Err 0
No pulse is detected
Err 1
Inflation rate is too low
Err 2
Systolic reading unattainable
Err 3
Cuff pressure is too high
Err 4
Diastolic pressure is outside of operation range
Err 5
Cuff deflated too fast
Err 6
Cuff deflated too slow
Err 7
Systolic pressure is outside of operation range
Err 8
Overall pressure is outside of operation range
Err 9
Cuff is unable to inflate to the necessary pressure level
18
Troubleshooting
If any abnormality occurs during use, please check and correct the
following:
Condition
Correction
Display is blank when power is on.
Check and correct the polarity of
the installed batteries. Reinstall or
replace batteries.
Measurement incomplete or
Review and follow ‘Applying Your
abnormally low or high values
Blood Pressure Cuff’ and ‘Taking
displayed.
Your Blood Pressure Reading’
sections. Replace the batteries.
Measurements are different from
Remember, blood pressure
those typically measured by
readings are influenced by physical
physician or every measurement
and mental conditions and/or even
reading is different.
the time of day. Daily measurement
recordings should be interpreted
by your physician.