MadgeTech RFOT Product User Manual - page 6
Product User Guide
6
Different wireless channels may be used to create multiple networks in one area, or to avoid wireless
interference from other devices. Any MadgeTech data logger or RFC1000 wireless transceiver that
is on the same network is required to use the same channel. If all of the devices are not on the
same channel, the devices will not communicate with one another. MadgeTech wireless data
loggers and RFC1000 wireless transceivers are programmed by default on channel 25.
Changing the channel settings of the RFOT
1.
Unscrew the end cap and gently separating the end from the
body of the data logger. Flip the black switch located inside
to the wireless position to ‘0’ (‘0’ indicates non-wireless
mode, ‘1’ indicates wireless mode).
2.
The RFOT has a set of DIP Switches located on the back of
the PCB, on the opposite side of the battery (diagram 1a).
3.
Change the DIP Switches to match the desired channel
(diagram 1b)
.
4.
Return the wireless switch to ‘1’.
5.
Screw the body of the logger back together making sure the
O-Rings are not visible.
Channel Programming
To configuring the channel settings of the RFC1000 wireless transceiver (sold separately),
please refer to the RFC1000 Product User Guide that shipped with the product or download it
from the MadgeTech website at
www.madgetech.com.
Channel DIP Switch Orientation
The images below show the orientations available of the switches for each channel.
CH 11
CH 19
CH 12
CH 20
CH 13
CH 21
CH 14
CH 22
CH 15
CH 23
CH 16
CH 24
CH 17
CH 25
CH 18
CH 26
NOT SUPPORTED
DEFAULT CHANNEL
1b.
1a.
CHANNEL NOTE: MadgeTech wireless data loggers and wireless transceivers purchased prior to
April 15, 2016 are programmed by default to channel 11. Please refer to the Product User Guide