6.1.2 Pulse Sound Setting
After turning the device on,the pulse sound is open.Long press the button can close the pulse sound
and the pulse sound indication icon disappears. Long press button again, the pulse sound is turned on
and pulse sound indication icon appears.
6.1.3 Alarm Setting
a.
Alarm includes the alarm of measure data's going beyond the limits (When the SpO
2
is below
90%, or the pulse rate is not between 50bpm and 120bpm, the alarm occurs), the alarm of
low-power.
b. In the open state of alarm, when the measure data is beyond the the normal measure range, the
device would give alarm sound and the corresponding value glitter. Alarm could be suspended by
short pressing button, and the alarm icon disappears, but the value still glitter.Alarm function will be
renewed in 30 seconds.
If low-power alarm occurs, please charge the battery;and if the battery has been out of normal
life, please replace it by a new one.
6.1.4 Display Mode Switch
In the normal measure or alarm pause state,short press the button can switch the display mode of the
screen.
6.1.5 Charge
Put the rechargeable buttony battery properly according to the polarity direction in the charger.
Connect one end of the adapter to the power supply socket, and connect another end of the adapter to
charger by data line.
The power adapter of the device accords with the requirements of IEC 60601-1.
The charger accords with the requirements of IEC 60601-1.
In charging state, the red indication light of the charger will shine, the green indication light of
the charger shining means the charge has been accomplished.
6.2 Attention for Operation
A. Please check the device before using, and confirm that it can work normally.
B. The finger should be in a proper position (see the attached illustration of figure 4 for reference),
or else it may result in inaccurate measure.
C. The ray between luminescent tube and photoelectric receiving tube must get across subject’s
arteriole.
D. The oximeter should not be used at a location or limb tied with arterial canal or blood pressure
cuff or receiving intravenous injection.
E. Ensure
nothing, such as a plaster, can impede the light passage, or else it may result in
inaccurate measure of SpO
2
and
pulse rate.
F. Excessive ambient light may affect measurement accuracy. It includes fluorescent lamp, dual
ruby light, infrared heater, direct sunlight and etc.
G. Intense activity of the subject or extreme electrosurgical interference may also affect the
accuracy.
H. Testee can not use enamel or other makeup.