NCI 6702 User Manual - Calibration Mode
Calibration Mode
The calibration (CAL) mode menu lets you calibrate your scale. The items in the
calibration menu are as follows:
Line Frequency (50 or 60H)
- Select the AC line frequency you are using.
Pounds/Kilograms (LB or KILO)
- Selects the unit of measure of your calibra-
tion test weights.
Capacity (9.995, etc..)
- Select the capacity of your scale.
Step-by-Step
Instructions for
CAL Mode
Follow these steps to calibrate your scale. refer to Figure 3.
1. From the dIAG display press the
ZERO key twice
OR
From normal weighing mode, move
switch 1 to the Menu Mode or
OPEN position, then press the
ZERO key twice. . .
CAL is displayed. This stands for calibration.
2. Press the TEST key. . .
50 h or 60 H is displayed. This is the AC line
frequency.
3. Press the ZERO key to toggle
between the choices. When the
choice you want is displayed, press
the TEST key. . .
That choice is accepted and LB or KILO is
displayed. This is the unit of measure of
your test weights.
4. Press the ZERO key to toggle
between the choices. When the
choice you want is displayed,
press the TEST key. . .
That choice is accepted and a scale capacity
is displayed. Example: 9.995.
5. Press the ZERO key to toggle
between the choices. When the
choice you want is displayed,
press the TEST key. . .
That choice is accepted and LOAD 0 is
displayed.
6. Clear all weight from the scale
platter and press the TEST key. . .
After a brief wait LOAD 10 is displayed .
10 is used only as an example here.
9