O2 Concepts, LLC Oxlife Independence User Manual - Troubleshooting Guide
23
TROUBLESHOOTING GUIDE
ALERT/ALARM CONDITION
DEVICE STATUS
SOLUTION
LOW BATTERY ALERT
When battery charge status reaches 10%,
the screen will turn yellow and “ALERT LOW
BATTERY” will be displayed in the System
Status window. The device will beep every
30 seconds and then go into Low Battery
Alarm mode.
Plug into an external power source.
OR
Replace depleted battery (batteries)
with a charged battery.
LOW PURITY ALERT
If Oxygen Purity drops below 85% the
screen will turn yellow and “ALERT LOW
PURITY” will be displayed in the System
Status window. The device will beep once
every 30 seconds until the condition is
corrected.
Cleanorreplaceairinletfilter.
Contact your authorized dealer.
NO BREATH DETECTED
If no breath is detected for 45 seconds
the screen will turn yellow and “ALERT NO
BREATH” will be displayed in the System
Status Window. The device will beep once.
If no breath is detected for an additional
15 seconds, the device will default to the
previous Continuous Flow Setting selected.
Ensure cannula tubing does not
exceed 7 feet (2.1m).
Note: No Breath Detected is used
when in Pulse Flow Mode ONLY.
LOW BATTERY ALARM
When the battery is depleted, the screen
will turn red and “LOW BATTERY” will be
displayed in the System Status window.
The device will beep for 15 seconds before
shutting down.
Plug into an external power source.
Replace depleted battery (batteries)
with a charged battery.
LOW PURITY ALARM
If Oxygen Purity drops below 72% the
screen will turn red and “LOW PURITY” will
be displayed in the System Status window.
The device will beep for 15 seconds before
shutting down.
Repairorreplacetubingorhumidifier.
Cleanorreplaceairinletfilter.
Contact your authorized dealer.
NO EXTERNAL POWER
ALARM
If no external power source is detected,
the screen will turn red and “NO EXTERNAL
POWER” will be displayed in the System
Status window. The device will beep every
second for 10 minutes before shutting
down.
Ensure all charging connections are
intact.
In the case of an extended power
outage, utilize an alternate, non-
powered, source of oxygen.
NO FLOW ALARM
Ifnooxygenflowisdetected,thescreen
will turn red and “NO FLOW” will be
displayed in the System Status window.
The device will beep for 15 seconds before
shutting down.
Check cannula connection.
Repairorreplacetubingorhumidifier.
Cleanorreplaceairinletfilter.
Move device to ensure adequate
airflow.
Contact your authorized dealer.
OVER TEMPERATURE ALARM
If the device is too hot to operate safely,
the screen will turn red and “OVER
TEMPERATURE” will be displayed in the
System Status window. The device will
beep for 15 seconds before shutting down.
Allow the device to reach operating
temperature.
Utilize an alternate source of oxygen if
necessary.
SERVICE REQUIRED ALARM
If service is required for the device to
operate safely, the screen will turn red and
“SERVICE REQUIRED” will be displayed in
the System Status window. The device
will beep for 15 seconds before shutting
down.
Contact your authorized dealer and
utilize an alternate supply of oxygen.
UNAUTHORIZED BATTERY
ALARM
If the device detects that an unauthorized
battery has been installed, the screen will
turn red and “UNAUTHORIZED BATTERY” will
be displayed in the System Status window.
The device will shut down.
Remove unauthorized battery and
replace with an authorized O2
Concepts battery.