Oakton -02 Operating Instructions Manual

Summary of -02

  • Page 1

    Distributed by: 68x243616 operating instructions oakton wd-35627-00, -02 temp tc acorn series meters r1 printed in the u.S.A. 9/00 temp jkt acorn series thermocouple thermometer on off cal hold enter ° c/ ° f j-k-t max min 9. Warranty oakton warrants this meter to be free from significant deviations...

  • Page 2

    3 2 table of contents 1. Introduction ............................................................................................4 2. Display and keypad functions .................................................................5 3. Preparation..........................................................

  • Page 3

    5 4 1. Introduction thank you for purchasing an oakton ® temp tc acorn meter. These economical, microprocessor-based meters work with type j, type k and type t probes for a wide temperature measurement range: • type j probes (iron-constantan) offer a wide -200 to 1000°c/ -328 to 1832°f range; use in...

  • Page 4

    3. Preparation inserting the batteries the battery compartment is found at the back of the instrument. To open the battery compartment, push the lid up. See figure note the polarity of battery before insert- ing the batteries into position. After replacement, place the cover back into its position a...

  • Page 5

    9 8 4. Measurement measuring temperature 1. Switch on meter. The acorn temp tc meter will automatically enter the temperature measurement mode. The "j", "k" or "t" annun- ciator will display, depending on what probe type was selected the last time the meter was used. The °c or °f annunicator will di...

  • Page 6

    11 10 5. Calibration temperature calibration using offset adjustment the temperature sensor included with your meter is factory calibrated. Over time, the temperature calibration may drift and require recalibration. If you replace the probe you should recalibrate. 1. Connect your temperature probe t...

  • Page 7

    Catalog number probe type temperature range time constant* general purpose probes. Wd-08517-55 j –90 to 760°c / -310 to 1400°f 3 seconds wd-08516-55 k -250 to 899°c / -418 to 1500°f 3 seconds wd-08500-55 t -250 to 400°c / -418 to 752°f 3 seconds penetration probes. Use for penetrating soft or semiso...

  • Page 8

    15 14.