Oakton -02 Operating Instructions Manual - page 4
3. Preparation
Inserting the batteries
The battery compartment is found at the
back of the instrument. To open the
battery compartment, push the lid up.
See figure
Note the polarity of battery before insert-
ing the batteries into position. After
replacement, place the cover back into
its position and press down until it
locks tight.
A "LO" annunciator in the LCD alerts
you when battery power is running
low. Replace batteries with a fresh set
as soon as possible.
Connecting your
Temperature sensor
Insert the mini connector plug of the temperature sensor into the connector on the side
of the meter. Note: the negative pin is larger than the positive pin; be sure to properly
align the pins. Unplug the sensor when not in use.
7
6
3.1
A
A
temperature
probe connections
on side of Acorn
TC meter
3.2
Switching the meter on
Press the ON/OFF key to power up your meter. All
the LCD segments display for a few seconds as the
meter goes through a self-diagnostic test. The LCD
then switches into measurement mode.
If the LCD then displays "oPEn", the temperature
sensor is faulty, or there is an open circuit.
See figure
See page 11 for more troubleshooting information.
3.3
A
A