OHAUS CD-11 Instruction Manual

Manual is about: Deluxe bench scale and indicator

Summary of CD-11

  • Page 1

    Page 1 of 30 0908 ih-606 h-606 1 ohaus corporation ohaus corporation ohaus corporation ohaus corporation ohaus corporation 29 hanover road florham park, nj 07932-0900 instruction manual model cd-11 indicator 2 iec1010-1 & en60950:1992 safety regulations iec1010-1 & en60950:1992 sicherheitsbestimmung...

  • Page 2

    Page 2 of 30 0908 ih-606 h-606 3 note: this equipment has been tested and found to comply with the limits for a class a digital device, pursuant to part 15 of the fcc rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a co...

  • Page 3

    Page 3 of 30 0908 ih-606 h-606 5 table of contents (cont.) 4.6 clear tare operation ......................................................................................................................................... 22 4.7 unit switch operation ....................................................

  • Page 4

    Page 4 of 30 0908 ih-606 h-606 5 table of contents (cont.) 4.6 clear tare operation ......................................................................................................................................... 22 4.7 unit switch operation ....................................................

  • Page 5

    Page 5 of 30 0908 ih-606 h-606 7 2. Installation 2.1 unpacking and checking open the package and remove the instrument and the accessories. Check the completeness of the delivery. The following accessories are part of the standard equipment of your new indicator. • remove packing material from the i...

  • Page 6

    Page 6 of 30 0908 ih-606 h-606 7 2. Installation 2.1 unpacking and checking open the package and remove the instrument and the accessories. Check the completeness of the delivery. The following accessories are part of the standard equipment of your new indicator. • remove packing material from the i...

  • Page 7

    Page 7 of 30 0908 ih-606 h-606 9 2.5 connecting power the cd-11 indicator may be operated using the ac adapter supplied, or 6 alkaline c-type batteries (not supplied). 2.5.1 ac adapter • connect the ac adapter connector to the receptacle located at the right-hand side of the indicator and plug the a...

  • Page 8

    Page 8 of 30 0908 ih-606 h-606 9 2.5 connecting power the cd-11 indicator may be operated using the ac adapter supplied, or 6 alkaline c-type batteries (not supplied). 2.5.1 ac adapter • connect the ac adapter connector to the receptacle located at the right-hand side of the indicator and plug the a...

  • Page 9

    Page 9 of 30 0908 ih-606 h-606 11 2.7 initial setup (cont.) 2.7.3 load cell capacity information review the specifications of the scale base to be used with the indicator. Make sure the settings you select in the indicator are compatible with the scale base. Below is a load cell scale capacity (lb o...

  • Page 10

    Page 10 of 30 0908 ih-606 h-606 11 2.7 initial setup (cont.) 2.7.3 load cell capacity information review the specifications of the scale base to be used with the indicator. Make sure the settings you select in the indicator are compatible with the scale base. Below is a load cell scale capacity (lb ...

  • Page 11

    Page 11 of 30 0908 ih-606 h-606 13 2.7 initial setup (cont.) 2.7.5 readout menu the readout menu is used to adapt the indicator to environmental conditions, set measuring units on/off, parts counting, auto zero tracking, timer on/off and retain zero data. It contains 10 submenus. Review all of the s...

  • Page 12

    Page 12 of 30 0908 ih-606 h-606 13 2.7 initial setup (cont.) 2.7.5 readout menu the readout menu is used to adapt the indicator to environmental conditions, set measuring units on/off, parts counting, auto zero tracking, timer on/off and retain zero data. It contains 10 submenus. Review all of the s...

  • Page 13

    Page 13 of 30 0908 ih-606 h-606 15 auto power off • press zero/menu button, aotoff is displayed. This is the auto off timer. When set on, the indicator will shut off automatically after 5 minutes has elapsed on the condition that no button is pressed and the scale base is stable during that period. ...

  • Page 14

    Page 14 of 30 0908 ih-606 h-606 15 auto power off • press zero/menu button, aotoff is displayed. This is the auto off timer. When set on, the indicator will shut off automatically after 5 minutes has elapsed on the condition that no button is pressed and the scale base is stable during that period. ...

  • Page 15

    Page 15 of 30 0908 ih-606 h-606 17 2.7 initial scale setup (cont.) 2.7.7 lockout switch menu lockout switch menu (locsw) is a software controlled option which can lock the settings in the calibration, setup, readout, and print menus to prevent tampering. When used in conjunction with the lock switch...

  • Page 16

    Page 16 of 30 0908 ih-606 h-606 17 2.7 initial scale setup (cont.) 2.7.7 lockout switch menu lockout switch menu (locsw) is a software controlled option which can lock the settings in the calibration, setup, readout, and print menus to prevent tampering. When used in conjunction with the lock switch...

  • Page 17

    Page 17 of 30 0908 ih-606 h-606 19 3. Calibration and sealing model cd-11 indicator requires span calibration before using. Span calibration ensures that the indicator reads correctly within specifications using weight values of 20% to 100% of capacity. For best results, calibrate at or near full ca...

  • Page 18

    Page 18 of 30 0908 ih-606 h-606 19 3. Calibration and sealing model cd-11 indicator requires span calibration before using. Span calibration ensures that the indicator reads correctly within specifications using weight values of 20% to 100% of capacity. For best results, calibrate at or near full ca...

  • Page 19

    Page 19 of 30 0908 ih-606 h-606 21 4. Operation before using the indicator, make sure it has been properly set up and calibrated. Refer to sections 2 and 3 and review settings. Four buttons are used: on/off - power on or off, print/units - short press prints data, long press changes unit of measure,...

  • Page 20

    Page 20 of 30 0908 ih-606 h-606 21 4. Operation before using the indicator, make sure it has been properly set up and calibrated. Refer to sections 2 and 3 and review settings. Four buttons are used: on/off - power on or off, print/units - short press prints data, long press changes unit of measure,...

  • Page 21

    Page 21 of 30 0908 ih-606 h-606 23 4. Operation (cont.) 4.7 unit switch operation (cont.) if enabled in program mode, additional units of measure may be utilized beyond the primary unit of measure used for callibration or display. Switching between primary and secondary units is accomplished by pres...

  • Page 22

    Page 22 of 30 0908 ih-606 h-606 23 4. Operation (cont.) 4.7 unit switch operation (cont.) if enabled in program mode, additional units of measure may be utilized beyond the primary unit of measure used for callibration or display. Switching between primary and secondary units is accomplished by pres...

  • Page 23

    Page 23 of 30 0908 ih-606 h-606 25 4.13 printing data printing data to an external computer or printer requires that the communications parameters in the print menu, be set first. Refer to paragraph 2.7.6 print menu for proper set up. • to print data, press print/units button with a short press. The...

  • Page 24

    Page 24 of 30 0908 ih-606 h-606 25 4.13 printing data printing data to an external computer or printer requires that the communications parameters in the print menu, be set first. Refer to paragraph 2.7.6 print menu for proper set up. • to print data, press print/units button with a short press. The...

  • Page 25

    Page 25 of 30 0908 ih-606 h-606 27 5.2 error codes list the following list describes the various error codes and which can appear on the display and the suggested remedy. Lobat is indicated when batteries are weak. Approximately 20 minutes of operating time remain. Error 1 indicates an overload cond...

  • Page 26

    Page 26 of 30 0908 ih-606 h-606 27 5.2 error codes list the following list describes the various error codes and which can appear on the display and the suggested remedy. Lobat is indicated when batteries are weak. Approximately 20 minutes of operating time remain. Error 1 indicates an overload cond...

  • Page 27

    Page 27 of 30 0908 ih-606 h-606 29 limited warranty ohaus products are warranted against defects in materials and workmanship from the date of delivery through the duration of the warranty period. During the warranty period ohaus will repair, or, at its option, replace any component(s) that proves t...

  • Page 28

    Page 28 of 30 0908 ih-606 h-606 ...

  • Page 29

    Page 29 of 30 0908 ih-606 h-606 before calibration or use, remove the red plastic shipping tabs in the corners of the scale base. ...

  • Page 30