OHAUS Valor 1000 Instruction Manual - 4.2 Backlight
EN-6
OHAUS Valor 1000 Series
Press SELECT to change the alarm setting.
Press ENTER to confirm setting and display [End]
Press and hold ZERO key then press SELECT to return to weighing.
4.2 Backlight
Press and hold ZERO button until [b9---x] is displayed, where x=
0 = backlight is turned off
1 = backlight is on during weighing
2 = backlight is always on
Press TARE to change the setting. Press ZERO to confirm the setting and return to
weighing.
4.3 CALIBRATION
Span Calibration:
Turn on the scale. When display flashes all segments ("888888") press and hold
ZERO and SELECT at the same time. "SET--- " will be displayed at the end of power on
sequence. Press the SELECT key until "SET-F1" is shown on the display.
Press ZERO key to display "A____x"(x=divisions in grams:1,2,5,10.20.50).
Press ZERO key to accept the default division setting which is shown on the display.
Press ZERO key to display "b____x"(x=decimal point:0,1,2,3). Press ZERO key to
accept the default decimal point setting which is shown on the display.
Press ZERO key to display "Cxxxxx"(xxxxx=Span weight or capacity of scale). Press
ZERO key to accept the default span calibration weight setting which is shown on the
display, or press M+ or TARE to increment or decrement the span calibration weight
setting.
Press ZERO to display "dx"(x=Zero tracking setting:0-9 from 0.125d to 2d). Press
ZERO key to accept the default setting which is shown on the display, or press M+ or
TARE to increment or decrement setting. Default setting is 3 which = 0.5d.
Press ZERO and display will show "Load 0". Make sure pan is empty then press ZERO
key to accept. Display will show "CAL 0" briefly, then "Load F” will be displayed.
Press ZERO to display "Fxxxxx"(xxxxx=Span weight ). Put displayed weight on the
scale, and press ZERO to accept. After completion, scale will be in weighing mode.