Plusmed pM-KN03T Instruction Manual - Trouble Shooting Guide
26
TROUBLE SHOOTING GUIDE
SYMPTOM
PROBABLE CAUSE
SOLUTION
Power on, the
equipment is not
working
1.No power supply from the
battery
2.Technical fault happens
1.Use AC/DC power to charge the battery
2. Contact your home care provider.
The battery is not
chargeable or AD/
DC power cannot
supply power to
equipment,
Fuse inside the power adapter
blowout
1.Change a new fuse
(Fuse model: T3.15AH250V,
size: 8.5*4*8.5)
2. Call service provider or dealer.
Battery
power
display on LCD
screen is empty and
with beep alarm
Lack of power
Use AC/DC power to charge the battery
LCD screen display
alarm code “E01”
Machine do not detect breath
for a period of 2 minutes (120
seconds) or more
Check the connection from the cannula
to the device. Make sure the nasal
cannula is properly positioned on your
face and that you are breathing through
your nose. Make sure the cannula
tubing
is not kinked or obstructed.
The LCD screen
display alarm code
“E02”
This indicates that the user’s
breath rate is exceeding the
capacity of the device.
The indicator resets itself when breath
rate is reduced. If this indicator occurs
on a routine basis, contact your home
care provider.
The LCD screen will
display alarm code
“H04”
This alarm occurs when the
device is delivering a lower
concentration of oxygen than
specified. This alarm can be
silenced by pressing the Alarm
symbol
Change to another source of oxygen
and contact your equipment provider.
The LCD screen will
display alarm code
“H05”
1.Unit overheating due to
blocked air intake
2.The surrounding weather is
hot
3.Other inside cooling system
has fault
1. Make sure no block to the machine’s
air inlet our exhaust window.
2.Stop the machine for a while to cool
the temperature and
The LCD screen will
display alarm code
“H06”
Technical Fault
Contact your home care provider.
.
All other problems
Call service provider
QUALITY WARRANTY
Model
Warranty Period
pM-KN03T
24 Months