Ramsey Medical statMAP Instructions Manual - Cuff Will Not Inflate.

Manual is about: Blood Pressure Measurement Device

Summary of statMAP

  • Page 1

    Www.Statmap.Net 800-231-6370 or 813-289-5555 patents pending ramsey medical, inc. Blood pressure measurement device instructions for ref # 7200-0001.

  • Page 2: Table of Contents

    Page 2 table of contents description 3 explanation of symbols 4 important information 5 precautions for use 6 guidance and manufacturer’s declaration 9 accessories 13 controls and indicators 14 operating instructions (semi-automatic mode) 17 operating instructions (manual mode) 22 battery replacemen...

  • Page 3: Description

    Page 3 description statmap™ is a portable battery-powered blood pressure measuring device. Statmap utilizes the oscillometric, step deflate method for non-invasive blood pressure determination. Statmap measures all of the blood pressure parameters: systolic, diastolic, mean arterial pressure and hea...

  • Page 4: Explanation of Symbols

    Page 4 explanation of symbols type bf applied part (defibrillation-proof) ipx1 drip –proof attention, refer to accompanying documents on / off momentary switch multi-function button.

  • Page 5: Important Information

    Page 5 important information indications for use: statmap is a prescription device and is indicated for use to measure systolic and diastolic pressure, mean arterial pressure (map) and heart rate of persons ages 12 and over using the oscillometric method of measurement. Statmap may be used in hospit...

  • Page 6: Precautions For Use

    Page 6 precautions for use cuff size – only use the cuff when the artery index marker falls within the printed range indicated on the cuff, otherwise erroneous readings may result. Use only cuffs approved by ramsey medical inc. (refer to accessories section for ordering information). Do not allow th...

  • Page 7

    Page 7 precautions for use (con’t) caution: care should be taken when using the statmap to observe the maximum allowed pump-up pressures as follows: adults 280mmhg pediatrics (12 to 16 years) 200mmhg do not immerse the statmap in water. If the device is splashed with water or becomes wet, wipe it im...

  • Page 8

    Page 8 precautions for use (con’t) the statmap is designed to conform to electromagnetic compatibility (emc) standard en 60601-1-2, 2001 and will operate accurately in conjunction with other medical equipment which also meets this requirement. To avoid interference problems affecting the statmap, do...

  • Page 9: Declaration

    Page 9 guidance and manufacturer’s declaration electromagnetic emissions the statmap is intended for use in the electromagnetic environment specified below. The customer or user of the device should assure that it is used in such an environment. Emissions test compliance electromagnetic environment ...

  • Page 10: Electromagnetic Immunity

    Page 10 electromagnetic immunity the statmap is intended for use in the electromagnetic environment specified below. The customer or user of the statmap should assure that it is used in such an environment. Immunity test iec60601 test level compliance level electromagnetic environment - guidance ele...

  • Page 11: Electromagnetic Immunity

    Page 11 electromagnetic immunity the statmap is intended for use in the electromagnetic environment specified below. The customer or user of the statmap should assure that it is used in such an environment . Immunity test iec60601 test level compliance level electromagnetic environment - guidance po...

  • Page 12: Statmap

    Page 12 recommended separation distances between portable and mobile rf communications equipment and the statmap the statmap is intended for use in an electromagnetic environment in which radiated rf disturbances are controlled. The customer or user of the device can help prevent electromagnetic int...

  • Page 13: Accessories

    Accessories the following blood pressure cuffs are approved for use with statmap : page 13 cuffs: reorder no. 8061 8062 8063 type small adult adult large adult arm circumference range 20.0 to 27.0 cm 25.3 to 34.3 cm 32.1 to 43.4 cm cuff size 10 size 11 size 12.

  • Page 14: Controls And Indicators

    Controls and indicators power button: power button turns the device on/off. (in order to conserve battery life, the statmap will automatically shut off after 10 minutes of “no use”). Multi-function button: the multi-function button is used to control two features of the statmap: 1) brightness of led...

  • Page 15

    Page 15 power/status indicator: this led indicator illuminates green to inform the user power is on. Status indicators: at the end of a determination, the leds in the status ovals will flash ambr and the automatic valve will open releasing pressure in the cuff. The user must purge the cuff by squeez...

  • Page 16: Inflation Bulb:

    Cuff connector: attach cuff connector securely to device connector. Inflation bulb: use to inflate cuff. Page 16 deflate trigger: when the red deflate trigger is in the up position (valve closed), cuff is ready for inflation. When the red deflate trigger is in the down position (valve open), air is ...

  • Page 17: Operating Instructions:

    Page 17 operating instructions: semi-automatic mode 1. The ideal circumstance for accurate bp measurement is to have the subject relaxed and seated comfortably with the arm free of clothing. The cuffed arm should be supported to maintain the cuff at the patient’s heart level. 2. Turn the statmap on....

  • Page 18: Operating Instructions:

    Page 18 7. If there is no nsv stored in the unit, there will be a sequential illumination of each of the leds to verify that all are function- ing properly. (subsequently, several other leds will then flash indicating the version of the operating software. These can be ignored unless needed by custo...

  • Page 19: Operating Instructions:

    Page 19 11. When both ambr lights flash in the status indicator areas, the bp determination is complete. The remaining air must be released from the cuff by squeezing the deflate trigger. The statmap will indicate bp and heart rate at the end of the deflation on the leds to the nearest 5mmhg tic mar...

  • Page 20: Automatic

    14 . Resetting statmap to clear the nsv clearing the nsv from statmap occurs automatically or manually. Automatic 1) turn unit off and back on. (record the nsv on the patient’s chart.) 2) begin new bp session with same or next patient by inflation of the cuff. 3) unit is cleared of all previous sess...

  • Page 21: Bright Dim

    Page 21 blood pressure display and map (mean arterial pressure) to the nearest mmhg the statmap will indicate blood pressure at the end of each measurement on the led display to the nearest 5mmhg tick mark. The multi-function button can be used to display the bp measurement to the nearest 1 mmhg. 1)...

  • Page 22: Operating Instructions:

    Page 22 operating instructions: manual mode 1. The ideal circumstance for accurate bp measurement is to have the subject relaxed and seated comfortably with the arm free of clothing. The cuffed arm should be supported to maintain the cuff at the patient’s heart level. 2. When using statmap in the ma...

  • Page 23: Battery Replacement:

    8. As the pressure falls, note systolic pressure at the first appearance of repetitive sounds (phase i korotkoff sounds). 9. Note the diastolic pressure when the korotkoff sounds disappear. 10. The air must be released from the cuff by squeezing the deflate trigger. Page 23 battery replacement: loos...

  • Page 24: Cleaning:

    Page 24 cleaning: statmap : clean with damp cloth. Do not immerse in water or other liquid. Do not use alcohol. Do not use steam or heat to sterilize the statmap. Cuffs, tubing and port fittings : before washing the cuff, remove it from the statmap, close off the tube with a plug, and place the hook...

  • Page 25: Leak Test:

    Page 25 leak test: the statmap should be checked annually for leakage or when there is doubt about the validity of measurements. This check may be accomplished by plugging the outlet port, inflating the device to 200-260 mmhg, and checking for needle movement. Caution: calibration and leak testing s...

  • Page 26: Troubleshooting

    Page 26 troubleshooting note: at the first indication of any statmap malfunction, the first attempt at resolving the problem should be to: replace both batteries with known fresh batteries disconnect the cuff squeeze out all residual air verify using correct size cuff reattach cuff to the statmap wh...

  • Page 27: Cuff Will Not Inflate.

    Page 27 difficulty obtaining consistent bp readings (greater than +/- 15 mmhg from reading to reading) 1. Excessive movement. Keep patient from moving during the measurement cycle 2. Wrong size cuff used. 3. Cuff not properly snug, i.E., not tight enough on initial placement. Tighten cuff on the lim...

  • Page 28: Method Of Bp Measurement:

    Indications/contraindications for use: statmap is a prescription device and is indicated for use to measure systolic and diastolic pressure, mean arterial pressure (map) and heart rate of persons ages 12 and over using the oscillometric method of measurement. There are no contraindications for the u...

  • Page 29: Battery Life:

    Page 29 battery life: depends on cuff size and battery type. Alkaline: adult cuff average 150-200. Lithium batteries give approximately 2x the readings of alkaline . Operating environment: temperature: 10°c to 40°c (50°f to 104°f) humidity: 15-90%, non-condensing altitude: - 500 feet (152 meters) be...

  • Page 30: Warranty

    Warranty the ramsey medical, inc. Statmap™ and cuffs are covered by a one year warranty against defects in materials and workmanship. Damage resulting from inappropriate use or physical abuse is not covered by the warranty. Units returned for warranty service will be repaired or replaced at the disc...

  • Page 32

    Awr400057-d 130604 phone: 800-231-6370 813-289-5555 fax: 813-289-5454 website: www.Statmap.Net e-mail: support@ramseymedical.Com tampa, fl 33607 usa statmap is a registered trademark of ramsey medical, inc. Patents pending manual revision 06/24/13 4920 w. Cypress st. Suite 110 tampa, fl 33607.