T&D TR-74Ui Introductory Manual

Other manuals for TR-74Ui: Introductory Manual
Manual is about: Illuminance UV Recorder

Summary of TR-74Ui

  • Page 1

    This "introductory manual" explains in a simple manner how to set up the tr-74ui and some of its basic operations. For details about the functions and features of the software "illuminance uv recorder for windows" and how to use them, please refer to the "operation guide (help)" included in the soft...

  • Page 2

    Notices about this introductory manual in order to properly use this product, please carefully read this manual before using. T&d corporation accepts no responsibility for any malfunction of and/ or trouble with this product or with your computer that is caused by the improper handling of this produ...

  • Page 3: Software User Agreement

    3 software user agreement disclaimers 1. Although t&d corporation has made operational tests on our software illuminance uv recorder for windows, we cannot guarantee that all operations will work properly under all conditions. 2. T&d corporation shall not accept any responsibility for any damage whe...

  • Page 4: Compliance Information

    4 compliance information radio, emc and safety regulations this device complies with part 15 of the federal communications commission (fcc) rules. Operation is subject to the following two conditions: (1)this device may not cause harmful interference, and (2)this device must accept any interference ...

  • Page 5: Contents

    5 contents safety precautions and instructions ---------------------------------- 6 basic procedures -------------------------------------------------------- 12 what is tr-74ui i lluminance uv recorder? ------------------- 13 package contents ------------------------------------------------------- 1...

  • Page 6: Explanation of Symbols

    6 safety precautions and instructions to ensure safety obey all of the following warnings. The following items should be strictly obeyed for the safe usage of this unit, and for protecting yourself and other people from bodily harm and/or damage to property. The tr-74ui illuminance uv recorder is he...

  • Page 7: Danger

    7 danger when installing and using the unit, make sure to follow all warnings and directions from your computer manufacturer. Do not take apart, repair or modify the unit. Doing so may cause fire or electrocution. If water or a foreign body enters into the unit, immediately turn off the power, remov...

  • Page 8: Caution

    8 caution this unit is not waterproof. If the unit gets dirty, wipe it with a clean cloth and a mild detergent. Harmful gases or chemicals may cause corrosion and/or other danger to the unit. Also, by coming in contact with hazardous substances, harm may occur to the people handling the unit. Theref...

  • Page 9

    9 we shall not guarantee the operation of the unit if you have connected it to your computer using a usb hub or a usb extension cable. Please do not insert your fingers or any foreign objects into any of the units' jacks. Do not use any other batteries than those that are specified in this introduct...

  • Page 10

    10 cautions about using the illuminance-uv sensors do not connect the isa-3151 sensor to any data logger other than the tr-74ui. Use the sensor within the operating environment specifications. When measuring ultraviolet light or other light which may cause damage or injury to your eyes or skin make ...

  • Page 11

    11 cautions about using the temperature-humidity sensors do not connect the sensor to any data loggers other than tr-74ui. Use the sensor within the operating environment specifications. If extremely severe temperature changes occur, the humidity measurements may appear abnormal. Once the sensor's t...

  • Page 12: Basic Procedures

    12 basic procedures the following outline shows the basic procedures for getting ready, making settings and using the product. 1. Getting ready (provided in this manual) 1. Install the software "illuminance uv recorder for windows" 2. Get the tr-74ui ready to use 3. Install the usb device driver and...

  • Page 13

    13 what is tr-74ui illuminance uv recorder? The tr-74ui illuminance uv recorder is a data logger designed to simultaneously measure and record illuminance, uv light, temperature and humidity. Storage capacity is up to 8,000 readings (one reading is a set of data which includes illuminance, uv light,...

  • Page 14: Package Contents

    14 package contents the following items are included in the package: tr-74ui illuminance uv recorder tr-74ui illuminance-uv sensor isa-3151 (cable length: 1.5 m) temperature-humidity sensor tha-3151 (cable length: 1.5 m) software cd-rom aa alkaline battery usb communication cable us-15c (cable lengt...

  • Page 15: Main Unit

    15 tr-74ui appearance diagram and part names main unit 78 18 55 unit : millimeters lcd display display button: pressing this button will change the lcd display pattern. Interval button: pressing this button will make it possible to make or change recording interval settings. Rec/stop button: pressin...

  • Page 16

    16 illuminance-uv sensor (isa-3151) unit : millimeters 23 12.6 65 1500 φ2.6 26.5 illuminance sensor area ultraviolet sensor area temperature-humidity sensor (tha-3151) unit : millimeters 18 10 51.5 1500 26.5 φ2.6 temperature-humidity sensor area temperature-humidity sensor (hha-3151) unit : millimet...

  • Page 17: Reading The Lcd Display

    17 reading the lcd display recording mark the recording status will appear. On: recording in progress. Blinking: waiting for programmed start. Off: recording has been stopped. Data capacity scale after every 2,000 readings the scale will be marked from left to right. Storage capacity is 8,000 readin...

  • Page 18: Main Display Messages

    18 main display messages the following are the main display messages that appear on the lcd display. Full (storage capacity full) upon reaching storage capacity of 8,000 readings, "full" will appear on the lcd display. When this message appears, recording has been stopped. (only when recording in on...

  • Page 19: Start Recording

    19 operations via buttons on tr-74ui note: - if no operations can be performed using buttons on the tr-74ui unit, it means the buttons have been deactivated via the software. Power button: turning on / off the power during recording, the power cannot be turned off by pressing the button on the unit....

  • Page 20

    20 interval button: setting / checking recording interval during recording or while waiting for a programmed recording to start via the software, recording interval settings cannot be made or changed. The factory default setting for the recording interval is 10 seconds. Setting the recording interva...

  • Page 21: Light.

    21 display button: changing the lcd display pattern there are two lcd display patterns for current readings: an alternating display and a fixed display. Alternating display: the lcd display shows all or selected multiple measurement items. Make settings for measurement items to be displayed via the ...

  • Page 22: Changing The Battery

    22 changing the battery besides the unit’s lcd display, the battery level can be displayed on the computer screen via the software if a unit is connected to the computer with a usb cable. If a unit is left without a battery for some time, all data may be lost, so please work quickly when changing th...

  • Page 23: Installing The Software

    23 installation setup and recording procedures do not connect a tr-74ui unit to the computer. Make sure to install the software before connecting the tr-74ui to the computer. Installing the software 11 start windows and login using a user account with administrator (computer administration) rights. ...

  • Page 24

    24 51 follow the directions as they appear to complete the installation. If a window appears such as the one below during installation, choose [install] button. 61 after installation, "illuminance uv recorder for windows" will appear in the windows start screen or start menu..

  • Page 25

    25 getting the tr-74ui ready to use 11 connect the sensor. The sensor jacks are common for both sensors. The temperature-humidity sensor or the illuminance-uv sensor can be connected to either of the sensor jacks. - to avoid poor connections, be sure to push the sensor connector securely into the ja...

  • Page 26: Starting Recording

    26 starting recording 1. Open illuminance uv recorder for windows and connect your device via usb to the computer. * the usb driver installation will start automatically. * when the usb driver installation is completed, a setting window will automatically open. * for details of the operations of ill...

  • Page 27: Faq

    27 faq q1 the power doesn't come on, what may be the problem? A check to make sure that the battery has been properly installed, and there is also sufficient battery power. Check to make sure that the battery is installed in the proper direction. If after checking and trying the above the power stil...

  • Page 28: Specifications

    28 specifications tr-74ui tr-74ui-h temperature- humidity sensor (external) tha-3151 hha-3151 (high-precision type) thermistor polymer resistance platinum resistance electrostatic capacitance measurement channels temperature 1ch humidity 1ch temperature 1ch humidity 1ch units of measurement °c, °f %...

  • Page 29

    29 responsiveness response time (90%): 3 sec. (at recording interval of 1 sec.) 6 sec. (at other intervals) logging capacity 8,000 data sets (one data set consists of readings for all channels in that type of unit.) recording interval select from 15 choices: 1, 2, 5, 10, 15, 20, 30 sec. Or 1, 2, 5, ...

  • Page 30

    30 relative spectral response characteristics graph (illuminance) broken line: the cie standard response function v ( λ ) solid line: isa-3151 relative spectral response characteristics graph (uv) isa-3151.

  • Page 31

    31 cosine correction characteristics (illuminance) broken line: cos θ solid line: measurement value cosine correction characteristics (uv) broken line: cos θ solid line: measurement values.

  • Page 32: Options

    32 options unit : millimeters sensor extension cable (tr-1c30) 37 tr-1c30: 3000 φ8 φ8.4 26 tr-1c30 cable length: 3 m temperature durability: -25 to 60°c usage figure isa-3151 tha-3151 tr-1c30 it is possible to use up to three extension cables for each sensor. Usb communication cable (us-15c) 1500 14...

  • Page 35: Product Support

    Product support for support, please contact the distributor from which you purchased the product. A list of distributors can be found at: http://www.Tandd.Com/purchasing/ product information product information can be found at: http://www.Tandd.Com/product/ illuminance uv recorder tr-74ui introducto...