TAMS 1800 Series Installation & Operation Manual - Appendix B: Error Messages
Appendix B: Error Messages
+0,”No error”
+1,”No scan list”
Some commands, like INIT, require that a
scan list be set.
+2,”Invalid during scan”
Some commands, like CLOSE and OPEN,
are invalid while the device is scanning.
-101,”Invalid char”
An invalid character was received. Valid
characters are A-Z, a-z, 0-9, : ; * @ ( ) ? +
space newline comma
-102,”Syntax error”
The command was not recognized. See
“Help/Manuals/Direct IO” for command
reference.
-120,”Numeric data error”
There was a numeric overflow while
parsing a number.
-200,”Execution error”
Internal error. Please call TAMS Inc.
-211,”Trigger ignored”
A *TRG or TRIG:IMM command was
sent, but the device is not scanning or the
TRIG:SOURCE is not BUS, MIX, or
HOLD.
-213,”Init ignored”
The device received an INIT command
while it was already scanning. (Use
ABORT to abort a scan.)
-220,”Parameter error”
A state recall command (*RCL) referenced
a state that has not yet been saved.
-222,”Data out of range”
A relay number was out of range.
-223,”Too much data”
A scan list or a relay list was too long.
-229,”Too many relays closed” In order the limit the current draw as
required by USB, some devices have a
fixed limit of the number of relays that can
be closed. A typical limit for most cards is
30 relays.
-400,”Query error”
A query was received by the device
BEFORE a previous query was finished.
- 21 -