Tanaka APM-7 Maintenance Manual - page 10
9
2
Adjustments
Adjustment of Indicated temperature correction, Initial heat value and some another parameter is
done by pressing panel keys. The operation procedure is as follows.
Press the FUNCTION key and RESET key, then consecutively release the RESET key and then
the FUNCTION key. The following MENU Screen will appear.
Fig. 3 MENU Screen
By pressing Right-Left key to select item, then press FUNCTION key to display each setting
screen. For the changing value, press Right key to blink the first character, then press Right-Left
key/ Up-Down key. Press RESET key after setting is completed.
Record changed value to “Appended Table 1 APM-7 Setting Table”. It will be required when
built-in battery goes dead (approx. 5 to 7 years).
2
.1 Filament Output Settings
Caution: Adjust filament output when a new igniter is installed. If
filament output value is high, the electric ignitor may be burnt
out in a short time.
Tanaka Scientific Limited will not be liable for burnt out
igniter, if appropriate adjustment was not carried out by the
user.
Each piece of igniter has a slightly different resistance and thus when a new piece is installed
supplied electric power needs to be adjusted. For the replacing procedure, see 4.2.5 on page 34.
Press Right-Left key to select Filament (blinking) from the MENU Screen, then press
FUNCTION key. The following Filament Output Setting Screen will appear.
Press Right-Left key to select Set output (output value is blinking), then change the output
value by pressing Up-Down key. The provisional values are as follows. If the AMP-7 is now in
APM Ver x.xF
Filament Para set Temp cal
Ext. out
Time set Baro cal
appears when selected Printer only
See page 15 2.4andpage 21 2.
5
.