Vacuubrand DCP 3000 Instructions For Use Manual - Safety Information
page 5 of 36
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999162 // 19/01/2011
General information
To operate the vacuum gauge DCP 3000, at least one external
pressure transducer VSK 3000 or VSP 3000 or MPT 100 is required!
Read and comply with this manual before installing or operating the equip-
ment.
Remove all packing material, remove the product from its packing-box, remove the
protective covers and keep, inspect the equipment.
If the equipment is damaged, notify the supplier and the carrier in writing within three
days; state the item number of the product together with the order number and the
supplier’s invoice number. Retain all packing material for inspection.
Do not use the equipment if it is damaged.
If the equipment is not used immediately, replace the protective covers. Store the
equipment in suitable conditions.
Intended use
☞
Operate the device only in combination with VACUUBRAND genuine accesso-
ries. Make sure that the individual components are only connected, combined
and operated according to their design and as indicated in the instructions for
use.
☞
Comply with notes on correct vacuum and electrical connections, see section
”Use and operation”.
• The device is designed for
ambient and gas temperatures
at the pressure trans-
ducer connection of +10°C to +40°C at continuous operation, up to +80°C for
short periods (
a cabinet or a housing check maximum temperature. Ensure that the maximum
permitted gas temperature at the pressure transducer (see ”Technical data”) is
not exceeded.
Use the equipment for the intended use only, i.e. for measurement of vacuum in ves-
sels designed for that purpose.
Connecting the device
☞
Avoid uncontrolled overpressure
(e.g., when connecting to a locked or blocked
tubing system).
Risk of bursting.
•
Comply with max. permitted pressure
at the pressure transducer, see section
”Technical data”.
•
Connect hoses at the pressure transducer gas tight.
•
The device is equipped with a
short circuit proof wide range power supply
with an integrated overload protection.
•
Check that mains voltage and current conform with the equipment (see rating
plate).
•
Prior to using the wall power supply, assemble and lock the suitable mains plug
(included in delivery) to the power supply.