Varian Cary 5000 User Manual
1
UserguideforCary5000absorptionspectrometer
withexternalDRA1800attachment.
(lastupdated03/17/2016)
ThisguideisforuseoftheCary5000withtheDRAonly.ForusewithouttheDRAasa
standardtransmissionspectrometerseetheUserguideforCary5000inabsorption
mode.
Importantwarnings!
• DonotunplugorplugintheexternalDRAattachmentwhentheinstrumentis
on.Thiswillruinthedetectorsoftheintegratingsphere
• DonotputwhitelightintotheDRAwiththeDRAelectricallyconnectedtothe
Cary.
• Minimizetheamountoftimethattheintegratingsphereisopen.Long-termlight
exposurehurtthedetectorovertime.
• Wearcleangloveswhileusingtheinstrument
• Donotuseliquidorpowdersamplesintheintegratingsphere.
QuickStartforusingDRA
1. Turnoncomputer
2. Checklensandmirrorarecorrectforyourmeasurement.
a. InstallSmallSampleKitmirror(M3SSK)orregularM3mirrorifneeded
b. IfusingSSKinstallcorrectlensL2.
3. MakesurecoverfortheDRAisfullyclosedandturnCaryon.
4. ClickontheCaryicononthetaskbar
5. Wait20minutesforthelampstowarmup.
6. Select
Setup
a. setwavelengthrange,%Tor%R
b. UnderOptionstab:setslitwidthandheight,SBWto2nm,Detectorand
Gratingchangewavelengths
c. UnderBaselinetab:chosezero/baseline
d. UnderStoragetab:Setyourfilename
7. Makesurethatallexitsoftheintegratingsphereareblockedwithfullreflectors.
8.
Baseline
andrun100%Tscan,thenblockthesamplebeamattheentrancetothe
integratingsphereforthe0%Tscan.Ifusinganapertureholderforreflection
measurements,pleaseseethe“Aperturekitforsmallsamples”sectiononpage4.
9. Mountyoursample
10. Run
Spectra
11. Whendoneremovesampleandremountreflectors.
12. TurnoffCary