Varian MidiVac 929-5000 Instruction Manual - page 27
INSTRUCTIONS FOR USE
22
87-900-080-01 (
D)
DESCRIPTION
REF. DESCRIPTION
with (12) set to OFF
with (12) set to ON
1
Downwards arrow When pressed, it decreases the value of the
function selected by means of the (16) pushbutton.
When pressed it changes the maximum output
voltage from 7 kV to 5 kV and from 5 kV to 3 kV.
2
Upwards arrow
When pressed, it increases the value of the
function selected by means of the (16) pushbutton.
When pressed it changes the maximum output
voltage from 3 kV to 5 kV and from 5 kV to 7 kV.
3
V
Control disabled.
When pressed the display (9) shows the value of the
output voltage to the pump.
4
I Control
disabled.
When
pressed the display (9) shows the pump
absorbed current.
5
Set 2
When it lights up the display (9) shows the Set
Point 2 value when selected by means of the (16)
pushbutton.
When it lights up the current value has reached the
Set Point 2 value.
6
Set 1
When it lights up the display (9) shows the Set
Point 1 value when selected by means of the (16)
pushbutton.
When it lights up the current value has reached the
Set Point 1 value.
7
Amp
When it lights up the display (9) shows the pump
absorbed current.
8
KV
When it lights up the display (9) shows controller
output voltage.
9
Display
It shows the voltage or current value selected by
means of the pushbutton (16).
It shows the controller output voltage, or the pump
absorbed current.
10
3Kv / 5Kv / 7Kv
The LED on corresponds to the maximum controller
output voltage.
The LED on corresponds to the maximum controller
output voltage.
11
PROTECT
When it lights up the “Protect” operating mode is
activated.
When it flashes the controller is in setting mode.
When it lights up the “Protect” operating mode is
activated.
12
HV ON/OFF
It switches on the high voltage.
It switched off the high voltage.
13
HV
When it lights up the high voltage is present.
14
REMOTE
When it lights up the MidiVac is controlled by an
host computer.
When it lights up the MidiVac is controlled by an host
computer.
15
START
PROT
When pressed it activates the “Protect” operating
mode.
LED (11) flashes during the setting operations.
When pressed it activates the “Protect” operating
mode.
16
MENU
If the MidiVac is in “Remote” mode, when pressed
it switches the operating mode to “Local”.
If the MidiVac is in “Local” mode, when pressed it
allows to access to the setting functions.
If the MidiVac is in “Remote” mode, when pressed it
switches the operating mode to “Local”.
If the MidiVac is in “Local” mode, when pressed it
allows to switch to “Remote”.