Operating instructions submersible pressure transmitter with ceramic measuring cell vegawell 52 4 … 20 ma/hart document id: 36785.
2 contents vegawell 52 • 4 … 20 ma/hart 36785-en-160923 contents 1 about this document 1.1 function ........................................................................................................................... 4 1.2 target group ............................................................
3 contents vegawell 52 • 4 … 20 ma/hart 36785-en-160923 9.2 disposal ......................................................................................................................... 29 10 supplement 10.1 technical data ...........................................................................
4 1 about this document vegawell 52 • 4 … 20 ma/hart 36785-en-160923 1 about this document 1.1 function this operating instructions manual provides all the information you need for mounting, connection and setup as well as important instruc- tionsformaintenanceandfaultrectification.Pleasereadthisinf...
5 2 for your safety vegawell 52 • 4 … 20 ma/hart 36785-en-160923 2 for your safety 2.1 authorised personnel all operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator. During work on and with the device the...
6 2 for your safety vegawell 52 • 4 … 20 ma/hart 36785-en-160923 2.5 safety label on the instrument the safety approval markings and safety tips on the device must be observed. 2.6 ce conformity thedevicefulfilsthelegalrequirementsoftheapplicableecguide- lines.Byaffixingthecemarking,vegaconfirmsucce...
7 3 product description vegawell 52 • 4 … 20 ma/hart 36785-en-160923 3 product description 3.1 configuration the scope of delivery encompasses: • vegawell 52 pressure transmitter with suspension cable • optionally available with straining clamp, screwed connection or housing with cable locking • doc...
8 3 product description vegawell 52 • 4 … 20 ma/hart 36785-en-160923 1 2 3 4 5 7 8 9 10 11 12 13 14 6 fig. 2: layout of the type label (example) 1 instrument type 2 product code 3 approvals 4 process temperature 5 signal output/operating voltage 6 materials transmitter/measuring cell/measuring cell ...
9 3 product description vegawell 52 • 4 … 20 ma/hart 36785-en-160923 1 2 3 4 5 fig. 3: front-flush installation of the certec ® measuring cell with double seal 1 housing, sensor 2 measuring cell 3 lateral seal for measuring cell 4 additional, front seal for measuring cell 5 diaphragm 4 … 20 ma/hart ...
10 3 product description vegawell 52 • 4 … 20 ma/hart 36785-en-160923 up to the time of installation, the packages must be left closed and stored according to the orientation and storage markings on the outside. Unless otherwise indicated, the packages must be stored only under the following conditi...
11 4 mounting vegawell 52 • 4 … 20 ma/hart 36785-en-160923 4 mounting 4.1 general instructions make sure that all parts of the instrument coming in direct contact with the process, especially the sensor element, process seal and processfitting,aresuitablefortheexistingprocessconditions,such as proce...
12 4 mounting vegawell 52 • 4 … 20 ma/hart 36785-en-160923 fig. 5: mounting example: vegawell 52 in an open basin with breather hous- ing vegabox 03 4.2 mounting steps with straining clamp 2 3 1 fig. 6: straining clamp 1 suspension cable 2 suspension opening 3 clamping jaws mount vegawell 52 with st...
13 4 mounting vegawell 52 • 4 … 20 ma/hart 36785-en-160923 4.3 mounting steps with screwed connection 1 2 3 4 5 6 fig. 7: threaded fitting 1 suspension cable 2 sealing screw 3 cone bushing 4 seal cone 5 threaded fitting 6 seal mount vegawell 52 with screwed connection as follows: 1. Weld the welded ...
14 4 mounting vegawell 52 • 4 … 20 ma/hart 36785-en-160923 4.4 mounting steps with threaded connection or housing 2 3 1 fig. 8: plastic housing 1 housing 2 seal 3 thread mount vegawell 52 as follows: 1. Weld the welded socket g1½ a or 1½ npt to the vessel top 2. Shift transmitter through the mountin...
15 5 connecting to power supply vegawell 52 • 4 … 20 ma/hart 36785-en-160923 5 connecting to power supply 5.1 preparing the connection generally connect the instrument only in the complete absence of line voltage. The instrument is equipped with an integrated overvoltage protection. For additional p...
16 5 connecting to power supply vegawell 52 • 4 … 20 ma/hart 36785-en-160923 1 2 3 fig. 9: connect vegawell 52 to power supply 1 direct connection 2 connection via vegabox 03 3 connection via housing if screened cable is required, we recommend connecting the cable screen on both ends to ground poten...
17 5 connecting to power supply vegawell 52 • 4 … 20 ma/hart 36785-en-160923 5.2 connection procedure proceed as follows: 1. Wire the suspension cable up to the connection compartment 2) 2. Connect the wire ends to the screw terminals according to the wiring plan connect the vegawell 52 according to...
18 5 connecting to power supply vegawell 52 • 4 … 20 ma/hart 36785-en-160923 2 3 + 1 - fig. 11: wiring plan vegawell 52 for 4 … 20 ma, 4 … 20 ma/hart 1 to the sensor 2 to power supply or processing system 3 shielding 3) wire number wire colour/polarity terminal 1 brown (+) 1 2 blue (-) 2 shielding g...
19 5 connecting to power supply vegawell 52 • 4 … 20 ma/hart 36785-en-160923 sensor 1 2 + ( ) (-) power supply 3 4 + ( ) (-) r off on hart 1 4 2 3 fig. 13: wiring plan vegawell 52 4 … 20 ma/hart 1 to the sensor 2 switch for communication resistor (on = activated, off = deactivated) 3 terminal for co...
20 6 set up with vegadis 82 vegawell 52 • 4 … 20 ma/hart 36785-en-160923 6 set up with vegadis 82 6.1 principle of operation and connection the vegadis 82 is an external display and adjustment unit without additional external energy. The instrument is suitable for measured value indication and adjus...
21 6 set up with vegadis 82 vegawell 52 • 4 … 20 ma/hart 36785-en-160923 6.3 setup steps youcanfindadetaileddescriptionofthesetupstepsforvegawell 52 in the operating instructions manual "vegadis 82 - 4 … 20 ma/ hart"..
22 7 setup with pactware vegawell 52 • 4 … 20 ma/hart 36785-en-160923 7 setup with pactware 7.1 connect the pc 1 5 2 3 4 open twist usb lo ck fig. 15: connection of the pc to vegabox 03 or communication resistor 1 pc with pactware 2 usb interface 3 vegaconnect 4 communication resistor 250 Ω 5 power ...
23 7 setup with pactware vegawell 52 • 4 … 20 ma/hart 36785-en-160923 7.2 parameter adjustment with pactware for parameter adjustment of the instrument via a windows pc, the configurationsoftwarepactwareandasuitableinstrumentdriver (dtm) according to fdt standard are required. The latest pactware ve...
24 7 setup with pactware vegawell 52 • 4 … 20 ma/hart 36785-en-160923 the standard version is available as a download under www.Vega.Com/downloads. The full version is available on cd from the agency serving you. 7.3 saving the parameter adjustment data it is recommended to document or save the para...
25 8maintenanceandfaultrectification vegawell 52 • 4 … 20 ma/hart 36785-en-160923 8 maintenanceandfaultrectification 8.1 maintenance if the instrument is used properly, no special maintenance is required in normal operation. Insomeapplications,productbuilduponthediaphragmcaninfluence the measuring r...
26 8maintenanceandfaultrectification vegawell 52 • 4 … 20 ma/hart 36785-en-160923 error code cause rectification 4 … 20 ma signal miss- ing connection to volt- age supply wrong – check connection according to chapter "connection steps" and if necessary, correct according to chapter "wiring plan" no ...
27 8maintenanceandfaultrectification vegawell 52 • 4 … 20 ma/hart 36785-en-160923 1 2 fig. 17: step 4 1 sw 46 2 sw 22 4. Pull the suspension cable out of the screwed socket, remove the pressure screw, cone sleeve and seal cone from the cable 5. Removethefilteradapterfromthecapillaryline 4 3 2 5 6 7 ...
28 8maintenanceandfaultrectification vegawell 52 • 4 … 20 ma/hart 36785-en-160923 8. Shift the seal screw, cone sleeve and seal cone to the suspension cable and insert the cable into the screwed socket, insert the wire ends through the cable entry into the mounting plate theworkstepsarefinished. 8.5...
29 9 dismount vegawell 52 • 4 … 20 ma/hart 36785-en-160923 9 dismount 9.1 dismounting steps warning: before dismounting, be aware of dangerous process conditions such as e.G. Pressure in the vessel or pipeline, high temperatures, cor- rosive or toxic products etc. Take note of chapters "mounting" an...
30 10 supplement vegawell 52 • 4 … 20 ma/hart 36785-en-160923 10 supplement 10.1 technical data note for approved instruments the technical data in the respective safety instructions are valid for approved instruments (e.G. With exapproval).Thesedatacandifferfromthedatalistedherein,forexampleregardi...
31 10 supplement vegawell 52 • 4 … 20 ma/hart 36785-en-160923 Ʋ straining clamp 0.2 kg (0.441 lbs) Ʋ threadedfitting 0.4 kg (0.882 lbs) Ʋ plastic housing 0.8 kg (1.764 lbs) Ʋ stainless steel housing 1.6 kg (3.528 lbs) input variable percentage value -10 … +110 % of the nominal measuring range pressu...
32 10 supplement vegawell 52 • 4 … 20 ma/hart 36785-en-160923 run-up time approx. 15 s step response time ≤200ms(ti:0s,0…63%) referenceconditionsandinfluencingvariables(accordingtodinen60770-1) reference conditions according to din en 61298-1 Ʋ temperature +15 … +25 °c (+59 … +77 °f) Ʋ relative humi...
33 10 supplement vegawell 52 • 4 … 20 ma/hart 36785-en-160923 long-termstability(accordingtodin16086,dinv19259-1andiec60770-1) applies to digital hart interface as well as to analoguecurrentoutput4…20ma.Specifications refer to the set span. Turn down (td) is the relation nominal measuring range/set ...
34 10 supplement vegawell 52 • 4 … 20 ma/hart 36785-en-160923 electromechanical data suspension cable Ʋ configuration four wires, one suspension wire, one breather capillary, screen braiding, foil, mantle Ʋ wire cross-section 0.5 mm² Ʋ wire resistance ≤0.036Ω/m Ʋ tensile strength ≥1200n(270poundforc...
35 10 supplement vegawell 52 • 4 … 20 ma/hart 36785-en-160923 electrical protective measures protection rating Ʋ transmitter ip 68 (60 bar) Ʋ housing ip 66/ip 67 overvoltage category zurversorgungüberspeisetrennerdurchnetzeder Überspannungskategorie iii protection class iii approvals instrumentswith...
36 10 supplement vegawell 52 • 4 … 20 ma/hart 36785-en-160923 10.2 dimensions vegawell 52, titanium with glass feedthrough sw46 (1.81") l 22 mm (0.87") g1 ½a/ npt1 ½ 316 l ~ 69 mm (2.72") ø 77 mm (3.03") 112 mm (4.41") m20x1,5/ ½ npt 3 2 1 sw 30 mm (1.18") g1½ a 175 mm (6.89") ø 33 mm (1.30") 250 mm...
37 10 supplement vegawell 52 • 4 … 20 ma/hart 36785-en-160923 10.3 industrial property rights vega product lines are global protected by industrial property rights. Further information see www.Vega.Com. Only in u.S.A.: further information see patent label at the sensor housing. Vega produktfamilien ...
38 index vegawell 52 • 4 … 20 ma/hart 36785-en-160923 index a accessories – measuring instrument holder 10 – mounting bracket 10 application area 8 c cable screening 16 connection – directly 17 – via housing 18 – via vegabox 03 18 – via vegadis 82 19 f faultrectification25 functional principle 8 i i...
39 notes vegawell 52 • 4 … 20 ma/hart 36785-en-160923.
Printing date: vega grieshaber kg am hohenstein 113 77761 schiltach germany 36785-en-160923 all statements concerning scope of delivery, application, practical use and operat- ing conditions of the sensors and processing systems correspond to the information available at the time of printing. Subjec...