Velleman VTBAL6 User Manual - page 5
VTBAL6
V. 02 – 07/07/2015
5
©Velleman nv
7.
Other Functions
7.1
Setting the Auto Power-Off Timer
1. Switch on the scales.
2. Hold both
and
keys pressed to enter the setting mode.
3. Set the timer using the numeric keys (timer range from 0 ~ 30 min). The scales will not switch off
automatically when the timer is set to 0.
4. Press
to confirm your choice.
Note: This function is only valid when powering the scales with a battery.
7.2
Calibration
1. Switch on the scales.
2. Hold both
and
keys pressed to enter the calibration mode.
3. Calibrate the zero point first. The scales display unit.0 in the PIECE WEIGHT window. Make sure there
is no object on the platform. Press
.
4. The scales display unload in the PIECE WEIGHT window. Press
to calibrate the zero point.
5. The scales will now display load in the PIECE WEIGHT window. Place a reference weight onto the
platform. The reference weight must be 25% ~ 100% FS. Press
when the displayed value has
stabilised.
6. The scales display inp.ld in the PIECE WEIGHT window. Now, enter the reference weight’s value using
the numeric keys and press
to confirm.
7. Remove the reference weight from the platform and press
to confirm the zero point again.
8. The scales will now automatically reset and return to the normal weighing mode. Exit the calibration
mode by pressing
. A calibration error will be reported by cal.err.
8.
Display Description
- - - - - -
Overload or overrange
- - - - - -
Uncertain value
- - - - - -
Weight signal is too small
0- - - - -
Zero point is too high
0- - - - -
Zero point is too low
Eep.er0
EEPROM cannot be accessed
Eep.er1
The EEPROM parameters differ from the backup data
Eep.er2
The parameter has not been set
Cal.err
Calibration error
Ad- - - -
Input signal over maximum range
Ad- - - -
Input signal under minimum range
Cap
Capacity-related data display
Uol
Voltage-related data display
Add
Totalised data
Pc.t
Piece weighing
Spl.pcs
Sample quantity
Up.pcs
Upper limit
Lo.pcs
Lower limit