W & H Lisa MB 17 User Manual - page 38
User manual
36
Annex 1 TECHNICAL CHARACTERISTICS
Electricity supply
Single phase 230 VAC ±10% – 50/60Hz – 10A
Sterilizer:
Working temperature / Humidity
Storage temperature / Humidity
Min. atmospheric pressure
Nominal voltage :
Max. absorbed power :
Max. current :
Dimensions overall :
Max. space required :
Clutter of the door movement :
Weight empty :
Max. mass in working condition fully loaded :
Max. heat output :
Max. noise level :
Name plate situated on the back side:
10°- 40°C / 0-90%.
-20°- 60°C / 0-90%
(empty)
0,5 bar
230 V
2100 W
9,2 A
W : 445mm / H : 410mm / D : 520mm (MB17) – 620mm (MB22)
W : 485mm / H : 460mm / D : 570mm (MB17) – 670mm (MB22)
W : 360mm / H : 400mm / D : 360mm
49 kg (MB17) or 56 kg (MB22)
129.4 N/m² (140 N/ foot) for MB17 or 177.3 N/m² (160 N/ foot) for MB22)
3000 KJ/hr
< 50dB
Steam generator
Power / Voltage :
Max. pressure / Max. temperature :
Safety overpressure valve :
Name plate on the steam generator:
1700 W / 230 VAC
4 bar / 150°C
5 bar
Sterilization chamber
Power / Voltage :
Max. pressure / Max. temperature :
Safety overpressure valve :
Total volume :
Usable space (identical for all cycles) :
Bacteriological filter :
Name plate on the chamber:
1000 W / 230 VAC
2,4 bar / 138°C
2,4 bar
17 l. Ø:250mm / Depth:350mm (MB17)
21 l. Ø:250mm / Depth:450mm (MB22)
12 l. W:195mm / H:205mm / D:300mm (MB17)
16 l. W:195mm / H:205mm / D:400mm (MB22)
0.3 µm
Distilled water (or demineralized )
:
Water quality :
Min. / Max. consumption :
Double tank / Autonomy :
Conform to EN 13060 annex C
0,2 l / 0,35 l (full porous load) on MB17
0,3 l / 0,50 l (full porous load) on MB22
Min 8 cycles (full porous load)
Connections
Parallel printer port
Miscellaneous
Fully micro-processor driven and controlled / touch screen
Mains filter / 2KV over tension filter.
Programmable stand-by mode.
STERILIZER CLASS B conform with following directives and norms
:
93/42/EEC
Medical Device Directive (MDD)
97/23/EEC
Pressure Equipment Directive (PED)
EN 13060
Small steam sterilizer.
EN 61010-1
Laboratory equipment – Safety requirements.
EN 61010-2-040
Laboratory equipment – Specific instructions for steam sterilizer
EN 61326
Electrical equipment for measurement, control and laboratory use - EMC requirements
Model
LISA
Jahr/Year
1999
SN 991240000
1,7 KW
4 Bar
0,05 L
Wasserinhalt / Water volume
Zul. Betriebsüberdruck / Max. Pressure
Zul. Warmeleistung / Max.Power
Steam generator
Prüfüberdruck und Ausführung entsprechen der
Dampfkesselverordnung. / Overpressure tests and
development are conform to vessel standards.
Model
LISA
Jahr/Year
1999
SN 991230000
1 KW
2,4 Bar
17 L
138°C
Sterilizer chamber
Zul. Warmeleistung / Max.Power
Zul. Betriebsüberdruck / Max. Pressure
Kammer inhalt / Chamber volume
Zul. Betriebstemperature / Max. T°