WaveSense jazz Oweners Manual - Important:
46
47
The WaveSense Jazz includes a unique positive feedback
feature where a happy face symbol is displayed with glucose
test results that fall within a specific target range. By entering
values for what your glucose levels should be before or after
a meal (target glucose ranges), you can customize when this
happy face symbol appears. Always consult your healthcare
professional for your recommended target ranges.
IMPORTANT:
It is advised to set the hypoglycemic and
hyperglycemic thresholds before making changes within
the glucose target range menu (refer to chapters 15-16 for
setting hypoglycemic and hyperglycemic warning alarms).
When setting pre and post meal ranges, the lower limits are
restricted by the hypoglycemic threshold and the meter will not
allow you to select a number below this threshold. (For example:
If you want the pre meal target range to begin at 90 mg/dL, then
the hypoglycemic alarm needs to be set at 90 mg/dL or less).
The upper limits for pre and post meal ranges are restricted by
the hyperglycemic threshold and the meter will not allow you to
select a number above this threshold. (For example: If you want
the post meal target range to end at 180 mg/dL, then the hyper-
glycemic alarm needs to be set at 180 mg/dL or higher).
Setting
The Glucose Target Range
17
CHAP
T
E
R
[1]
When your meter is deactivated, press the CHECK ( )
button once to enter menu mode. The word “MEM” is dis-
played.
[2]
Press the DOWN (
) or UP (
) arrow buttons
until the word “RANGE” is displayed.
[3]
Press the CHECK ( ) button to enter and edit
the glucose target range settings.
[4]
Set the lower limit of the pre meal glucose
target range (flashing) with the DOWN (
)
or UP (
) arrow buttons. Press the CHECK
( ) button to confirm your selection and con-
tinue on to the upper limit pre meal setting.
[5]
Set the upper limit of the pre meal
glucose target range (flashing) with the
DOWN (
) or UP (
) arrow buttons. Press
the CHECK ( ) button to confirm your se-
lection and continue on to the lower limit
post meal setting.
[6]
Set the lower limit of the post meal glucose
target range (flashing) with the DOWN (
) or
UP (
) arrow buttons. Press the CHECK ( )
button to confirm your selection and continue
on to the upper limit post meal setting.
[7]
Set the upper limit of the post meal
glucose target range (flashing) with the
DOWN (
) or UP (
) arrow buttons. Press
the CHECK ( ) button to confirm your
selection and return to the menu.
Hyp
ogl
y
cemic
T
hreshold
Hyp
ergl
y
cemic
T
hreshold
GLUCOSE TARGET RANGE
(
mg/d