Zeiss ConfoCor 3 Quick Manual - Starting The System
2
04/2009
Starting the System
Switching on the LSM system
! Switch on the main switch (Fig. 1/1) and the
safety lock (Fig. 1/2).
! When set to ON the power remote switch
labeled System/PC provides power to the
computer. This allows use of the computer and
ZEN software offline
! To completely switch on the system, now press
the Components switch to ON. This starts the
other components and the complete system is
ready to be initialized by the ZEN software.
Switching on the X-Cite 120 or the HBO
100 mercury lamp
! Switch on the main switch of the X-Cite 120 /
HBO 100 lamp for reflected light illumination
via the power supply as described in the
respective operating manual.
Switching on the Ar-ML Laser
! If the Ar-ML laser is required, switch it on via
the toggle switch (Fig. 2/
2) on the power supply
and turn the key (Fig. 2/
1).
The laser is automatically kept in standby mode
for 5 minutes to warm up.
! Set the idle-run-switch (Fig. 2/3) to run. It takes
about 50s until the laser has reached the set
output power (green LED) provided the warm
up time of 5 minutes is already completed.
! Adjust the required power level with the control
knob (Fig. 2/
4) (default position should be
11 o’clock).
Fig. 1
Power remote switch
0
Las
er - F
ertig
ung
G
mbH
laser
run
Fig. 2
Power supply of Ar-ML laser