3B LASER PEN 200 Operating Instructions Manual - page 11
Treatment ends automatically after the selected therapy time has elapsed or when the selected energy dose has
been achieved. The end of the therapy is indicated by short triple sound signal.
Treatment can be terminated manually at any time by pressing the LASER key (1).
Once the therapy has ended, the screen changes back to the “Base setting” display.
Note: When no laser application is taking place, the unit turns “OFF” automatically, if it is not operated for a
period of >= 10 minutes (AUTO OFF).
1. The 3B LASER PEN must only be operated with inserted laser tip.
2. Risk of burns:
When the laser out put is >300 mW, especially in CW mode, the skin is heated. Please alert the patient to this fact.
Terminate the therapy session immediately when the patient finds the laser therapy too hot.
No direct irradiation of tattoos, moles, highly pigmented skin or from a distance to the treatment area less than 2 cm.
When applying a laser output of more than 300 mW, please note:
No irradiation with direct contact.
Distance to the skin approx. 2 cm.
Move the applicator slowly over the treatment area making circular or linear motions.
Max. dosage in treatments per patient: 60 Joule.
Please double-check on the monitor that the correct dosage has been set before starting treatment. The therapy time
is calculated automatically on the basis of the selected laser power and programme.
Maximum treatment time per patient is 10 min.
3. Risk of infection:
When treating injured skin and wounds, avoid direct contact with the skin to prevent contamination with germs.
In the presence of wounds and injuries, keep a distance of approx. 1 cm to the skin.
4. Please replace the tip after each treatment. A second tip has been provided.
Disinfect the laser tip after each treatment.
The tip may be disinfected in an autoclave or via spray-and-wipe disinfection with a 70% alcohol .
The 3B LASER PEN works with a maximum laser output of 500 mW. At maximum power (500 mW in CW mode) and
a fully charged battery, the battery time is approx. 1 h.
If the battery time becomes significantly shorter, replace the batteries with new ones.
As a general rule, the unit should be placed in the charger station after every application. But note that the laser pen
must be charged whenever the battery level indicator is <50%.
Charge the unit as long as possible, preferably until the charge symbol “disappears” (4 h).
Always keep a fully charged replacement battery at hand. Please note the user manual for the charger that has been
supplied.
Use only batteries of the same kind (AA – NiMH 1.2 V – min. 2,450 mAh). How to replace the batteries is explained
on p. 15 of the User Manual.