Checking the meter
177
Did you perform the control test according to
the User’s Manual?
yes
Did you wait for the application prompt to be
displayed before applying control solution?
yes
Did you apply the control solution immedi-
ately after opening the applicator?
yes
Did you enter the correct colour?
yes
Did you perform the test within the correct
temperature range (10–40 °C)?
yes
Has the expiry date for the control solution
been reached (see label on the bottom of the
applicator next to the symbol )?
no
Did you observe the proper storage condi-
tions for the meter, test cassette and control
solutions (see Chapter 15 and the test cas-
sette and control solution package inserts)?
yes
If your meter has been dropped, that can also lead to
control test results which are not within the concentra-
tion range displayed and for which
Control test not OK
is displayed. In this case, you should contact the
Accu-Chek Enquiry Line on 1800 251 816.