IMPORTANT INFORMATION REGARDING ELECTRO MAGNETIC
COMPATIBILITY (EMC)
This product needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided,
and this unit can be affected by portable and mobile RF communications
equipment.
1) *Do not use a mobile phone or other devices that emit electromagnetic
fields, near the unit. This may result in incorrect operation of the unit.
2) Caution: This unit has been thoroughly tested and inspected to assure
proper performance and operation!
3) *Caution: this machine should not be used adjacent to or stacked
with other equipment and that if adjacent or stacked use is necessary,
this machine should be observed to verify normal operation in the
configuration in which it will be used.
4) Warning:
The use of ACCESSORIES, transducers and cables other than those
specified, with the exception of transducers and cables sold by the
MANUFACTURER of the ELECTRICAL STIMULATOR as replacement parts
for internal components, may result in increased EMISSIONS or decreased
IMMUNITY of the ME EQUIPMENT or ME SYSTEM.
Guidance and manufacture’s declaration – electromagnetic emission
The ELECTRICAL STIMULATOR is intended for use in the electromagnetic environ-
ment specified below. The customer or the user of the ELECTRICAL STIMULATOR
should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment –
guidance
RF emissions
CISPR 11
Group 1
The ELECTRICAL STIMULATOR uses RF
energy only for its internal function. There-
fore, its RF emissions are very low and are
not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The ELECTRICAL STIMULATOR is suit-
able for use in all establishments, including
domestic establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings used
for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies