  • Page 1

    Maintenance manual lmr master s412e vector network analyzer 500 khz to 1.6 ghz (6 ghz with option 16) spectrum analyzer 9 khz to 1.6 ghz (6 ghz with option 6) signal generator 500 khz to 1.6 ghz anritsu company 490 jarvis drive morgan hill, ca 95037-2809 usa http://www.Anritsu.Com part number: 10580...

  • Page 3: Trademark Acknowledgments

    Trademark acknowledgments lmr master is a trademark of anritsu company. Notice anritsu company has prepared this manual for use by anritsu company personnel and customers as a guide for the proper installation, operation and maintenance of anritsu company equipment and computer programs. The drawing...

  • Page 5: Safety Symbols

    S412e mm pn: 10580-00324 rev. N safety-1 safety symbols to prevent the risk of personal injury or loss related to equipment malfunction, anritsu company uses the following symbols to indicate safety-related information. For your own safety, please read the information carefully before operating the ...

  • Page 6: For Safety

    Safety-2 pn: 10580-00324 rev. N s412e mm for safety warning always refer to the operation manual when working near locations at which the alert mark, shown on the left, is attached. If the operation, etc., is performed without heeding the advice in the operation manual, there is a risk of personal i...

  • Page 7: Table of Contents

    S412e mm pn: 10580-00324 rev. N contents-1 table of contents chapter 1—general information 1-1 introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1 1-2 product description . . . . . . . . . . . . . . . . . . ...

  • Page 8

    Contents-2 pn: 10580-00324 rev. N s412e mm table of contents (continued) 5-9 tetra evm verification, option 581 and 582 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-23 5-10 dmr2 modulation fidelity verification, option 591 and 592 . . . . . . . . . . . . . . . . . . . . ...

  • Page 9: 1-1

    S412e mm pn: 10580-00324 rev. N 1-1 chapter 1 — general information 1-1 introduction this manual provides maintenance instructions for the anritsu lmr master s412e. The information includes: • chapter 1, “general information” • chapter 2, “vector network analyzer verification” • chapter 3, “spectrum...

  • Page 10: 1-3

    1-3 related documents chapter 1 — general information 1-2 pn: 10580-00324 rev. N s412e mm 1-3 related documents other documents are available for the s412e at the anritsu web site at www.Anritsu.Com • lmr master s412e user guide, pn 10580-00318 • lmr master s412e programming manual, pn 10580-00319 •...

  • Page 11: 1-5

    Chapter 1 — general information 1-5 anritsu customer service centers s412e mm pn: 10580-00324 rev. N 1-3 1-5 anritsu customer service centers for the latest service and sales information in your area, please visit the following url: http://www.Anritsu.Com/contact.Asp choose a country for regional co...

  • Page 12

    1-6 recommended test equipment chapter 1 — general information 1-4 pn: 10580-00324 rev. N s412e mm table 1-2. Recommended test equipment for spectrum analyzer verification equipment critical specification recommended manufacturer/model frequency reference frequency: 10 mhz symmetricom model rubisour...

  • Page 13

    Chapter 1 — general information 1-6 recommended test equipment s412e mm pn: 10580-00324 rev. N 1-5 table 1-3. Recommended test equipment for signal generator verification equipment critical specification recommended manufacturer/model rf coaxial cable n(m) to n(m) connectors anritsu 15nn50-1.5c freq...

  • Page 14: 1-7

    1-7 replaceable parts and assemblies chapter 1 — general information 1-6 pn: 10580-00324 rev. N s412e mm 1-7 replaceable parts and assemblies to ensure that the correct options are provided on the replacement assembly when ordering either a main pcb assembly or main/spa pcb assembly, all installed i...

  • Page 15

    Chapter 1 — general information 1-7 replaceable parts and assemblies s412e mm pn: 10580-00324 rev. N 1-7 a. When ordering the main/vna pcb assembly, in order to ensure installation of correct options, all options that are installed on the instrument must be declared on the order. The options are lis...

  • Page 16: 1-8

    1-8 performance verification procedures chapter 1 — general information 1-8 pn: 10580-00324 rev. N s412e mm 1-8 performance verification procedures performance verification procedures are in the following chapters: • chapter 2, “vector network analyzer verification” • chapter 3, “spectrum analyzer v...

  • Page 17: Verification

    S412e mm pn: 10580-00324 rev. N 2-1 chapter 2 — vector network analyzer verification 2-1 introduction this chapter provides the performance verification procedure for the vector network analyzer section of s412e. The vector network analyzer performance verification tests consist of the following: • ...

  • Page 18: 2-3

    2-3 transmission dynamic range chapter 2 — vector network analyzer verification 2-2 pn: 10580-00324 rev. N s412e mm 2-3 transmission dynamic range the following test is used to verify the transmission dynamic range of the vector network analyzer. The procedures are automated by using the lmr master ...

  • Page 19: 2-4

    Chapter 2 — vector network analyzer verification 2-4 return loss measurement accuracy verification s412e mm pn: 10580-00324 rev. N 2-3 2-4 return loss measurement accuracy verification the following test can be used to verify the accuracy of return loss measurements. Measurement calibration of the s...

  • Page 20

    2-4 return loss measurement accuracy verification chapter 2 — vector network analyzer verification 2-4 pn: 10580-00324 rev. N s412e mm.

  • Page 21: 3-1

    S412e mm pn: 10580-00324 rev. N 3-1 chapter 3 — spectrum analyzer verification 3-1 introduction this chapter provides the performance verification procedure for the spectrum analyzer section of s412e. The spectrum analyzer performance verification tests consist of the following: • “frequency accurac...

  • Page 22: 3-2

    3-2 frequency accuracy verification and adjustment chapter 3 — spectrum analyzer verification 3-2 pn: 10580-00324 rev. N s412e mm 3-2 frequency accuracy verification and adjustment the following procedure is used to verify and adjust the frequency accuracy of the spectrum analyzer in the s412e. Adju...

  • Page 23

    Chapter 3 — spectrum analyzer verification 3-2 frequency accuracy verification and adjustment s412e mm pn: 10580-00324 rev. N 3-3 the measured value in table a-7 . Skip step 17 through step 19 and proceed to step 20 . If the marker frequency is outside the specification, proceed to the next step for...

  • Page 24: 3-3

    3-3 residual spurious response chapter 3 — spectrum analyzer verification 3-4 pn: 10580-00324 rev. N s412e mm 3-3 residual spurious response the following test is used to verify the residual spurious response of the spectrum analyzer of the s412e. Equipment required • 50 ohm termination, anritsu mod...

  • Page 25

    Chapter 3 — spectrum analyzer verification 3-3 residual spurious response s412e mm pn: 10580-00324 rev. N 3-5 residual spurious test with preamp on 19. Touch the pre amp on/off submenu button to turn it on. 20. Touch the freq main menu button and set the start freq to 10 mhz. 21. Set the stop freq t...

  • Page 26: 3-4

    3-4 displayed average noise level (danl) chapter 3 — spectrum analyzer verification 3-6 pn: 10580-00324 rev. N s412e mm 3-4 displayed average noise level (danl) the following test is used to verify the displayed average noise level (danl) of the spectrum analyzer in the s412e. This test is performed...

  • Page 27

    Chapter 3 — spectrum analyzer verification 3-4 displayed average noise level (danl) s412e mm pn: 10580-00324 rev. N 3-7 danl test with preamp on 21. Touch the amplitude button on the touchscreen main menu and set the reference level to –50 dbm. 22. Touch the pre amp on/off submenu button to turn the...

  • Page 28: 3-5

    3-5 single side band (ssb) phase noise chapter 3 — spectrum analyzer verification 3-8 pn: 10580-00324 rev. N s412e mm 3-5 single side band (ssb) phase noise this test is used to verify the single side band (ssb) phase noise of the spectrum analyzer in the s412e. Equipment required • synthesized sign...

  • Page 29: 3-6

    Chapter 3 — spectrum analyzer verification 3-6 spurious response (second harmonic distortion) s412e mm pn: 10580-00324 rev. N 3-9 3-6 spurious response (second harmonic distortion) the following test is used to verify the input related spurious response of the spectrum analyzer in the s412e. Equipme...

  • Page 30

    3-6 spurious response (second harmonic distortion) chapter 3 — spectrum analyzer verification 3-10 pn: 10580-00324 rev. N s412e mm 24. Set the # of average to 5. 25. Wait until the trace count displays “5/5”. 26. Touch the marker main menu button and then peak search submenu button. 27. Record the a...

  • Page 31: 3-7

    Chapter 3 — spectrum analyzer verification 3-7 resolution bandwidth accuracy s412e mm pn: 10580-00324 rev. N 3-11 3-7 resolution bandwidth accuracy the following test is used to verify the resolution bandwidth accuracy of the spectrum analyzer in the s412e. Equipment required • synthesized signal ge...

  • Page 32: 3-8

    3-8 amplitude accuracy chapter 3 — spectrum analyzer verification 3-12 pn: 10580-00324 rev. N s412e mm 3-8 amplitude accuracy the following tests are used to verify the amplitude accuracy of the spectrum analyzer in the s412e. The tests consist of: • “50 mhz amplitude accuracy verification” on page ...

  • Page 33

    Chapter 3 — spectrum analyzer verification 3-8 amplitude accuracy s412e mm pn: 10580-00324 rev. N 3-13 50 mhz amplitude accuracy verification test setup characterization 1. Turn on the 10 mhz reference standard, the ml2438a power meter, the mg3692x signal generator, and the s412e. 2. Install the 34r...

  • Page 34

    3-8 amplitude accuracy chapter 3 — spectrum analyzer verification 3-14 pn: 10580-00324 rev. N s412e mm 50 mhz amplitude accuracy measurements 1. Disconnect sensor a from the 10 db fixed attenuator and then install the 34nn50a adapter to the end of the attenuator. 2. Connect the open end of the 34nn5...

  • Page 35

    Chapter 3 — spectrum analyzer verification 3-8 amplitude accuracy s412e mm pn: 10580-00324 rev. N 3-15 4. Preset the s412e by pressing the shift key, then the preset (1) key, then the preset button on the touchscreen. Using the sweep menu, change the sweep mode to performance. 5. Touch the freq main...

  • Page 36

    3-8 amplitude accuracy chapter 3 — spectrum analyzer verification 3-16 pn: 10580-00324 rev. N s412e mm do this for both sensor a and sensor b. This sets the power meter to use the proper power sensor cal factor. Press the system key to display the power reading. 9. Adjust the mg3692x output level so...

  • Page 37: 3-9

    Chapter 3 — spectrum analyzer verification 3-9 third order intercept (toi) s412e mm pn: 10580-00324 rev. N 3-17 3-9 third order intercept (toi) the following test verifies the third order intercept point (also known as toi or ip3) of the spectrum analyzer in the s412e. Equipment required • 10 mhz re...

  • Page 38

    3-9 third order intercept (toi) chapter 3 — spectrum analyzer verification 3-18 pn: 10580-00324 rev. N s412e mm setup figure 3-3. Third order intercept (toi) pre-test setup s412e lmr master power charge +/- . 0 3 sweep 2 calibrate 1 preset 6 limit 5 trace 4 measure 9 mode 8 system 7 file shift back ...

  • Page 39

    Chapter 3 — spectrum analyzer verification 3-9 third order intercept (toi) s412e mm pn: 10580-00324 rev. N 3-19 800 mhz toi test 1. Connect the equipment as shown in figure 3-3 . The power splitter will be used as an rf combiner (the normal rf outputs will become inputs, and the normal input will be...

  • Page 40

    3-9 third order intercept (toi) chapter 3 — spectrum analyzer verification 3-20 pn: 10580-00324 rev. N s412e mm 15. Turn on the rf of both mg3692x #1 and mg3692x #2. 16. Press the shift key and then the trace (5) key. Press trace a operations, and set # of averages to 2. 17. After two sweeps have oc...

  • Page 41

    Chapter 3 — spectrum analyzer verification 3-9 third order intercept (toi) s412e mm pn: 10580-00324 rev. N 3-21 20. Determine which signal is larger (the 799.851 mhz signal or the 800.151 mhz signal). Use the larger value for max in the following formula to calculate the toi for 800 mhz (refer to th...

  • Page 42

    3-9 third order intercept (toi) chapter 3 — spectrum analyzer verification 3-22 pn: 10580-00324 rev. N s412e mm.

  • Page 43: Verification

    S412e mm pn: 10580-00324 rev. N 4-1 chapter 4 — cw signal generator verification 4-1 introduction this chapter provides the procedures for verifying the performance and the functionality of the cw signal generator in the s412e. The test procedures consist of the following: • “frequency accuracy veri...

  • Page 44

    4-2 frequency accuracy verification chapter 4 — cw signal generator verification 4-2 pn: 10580-00324 rev. N s412e mm.

  • Page 45: 5-1

    S412e mm pn: 10580-00324 rev. N 5-1 chapter 5 — option verification 5-1 introduction this chapter provides the procedures for verifying the performance and the functionality of various options in the s412e. The test procedures consist of the following: • “vna mode bias tee operational check, option ...

  • Page 46

    5-2 vna mode bias tee operational check, option 10 chapter 5 — option verification 5-2 pn: 10580-00324 rev. N s412e mm voltage and current test 8. Connect the 105 ohm load to vna port 2 connector of the s412e. 9. Ensure that the int voltage p2 setting is 12.0 v. 10. Touch the bias tee submenu button...

  • Page 47: 5-3

    Chapter 5 — option verification 5-3 gps verification, option 31 s412e mm pn: 10580-00324 rev. N 5-3 5-3 gps verification, option 31 the following test is used to verify the operation of the gps option on the s412e. • “touch the gps submenu button.” on page 5-3 • “spectrum analyzer frequency accuracy...

  • Page 48

    5-3 gps verification, option 31 chapter 5 — option verification 5-4 pn: 10580-00324 rev. N s412e mm 16. Press the esc key to dismiss the gps info dialog box. 17. Touch the gps on/off submenu button to turn gps feature off. 18. Remove the adapter and terminator from the gps connector of the s412e. 19...

  • Page 49

    Chapter 5 — option verification 5-3 gps verification, option 31 s412e mm pn: 10580-00324 rev. N 5-5 9. Touch the amplitude button on the touchscreen main menu and set the reference level to –10 dbm. 10. Touch the span submenu button and set to 10 khz. 11. Touch the bw main menu button and set the rb...

  • Page 50: 5-4

    5-4 fixed wimax signal analyzer verification, option 46 and 47 chapter 5 — option verification 5-6 pn: 10580-00324 rev. N s412e mm 5-4 fixed wimax signal analyzer verification, option 46 and 47 the tests in this section verify the performance of the optional fixed wimax signal analyzer of the s412e ...

  • Page 51

    Chapter 5 — option verification 5-4 fixed wimax signal analyzer verification, option 46 and 47 s412e mm pn: 10580-00324 rev. N 5-7 procedure 1. Connect the ma2482d power sensor to the power meter and zero the sensor. 2. Set the cal factor frequency of the power sensor to 2600.5 mhz. 3. Connect the e...

  • Page 52

    5-4 fixed wimax signal analyzer verification, option 46 and 47 chapter 5 — option verification 5-8 pn: 10580-00324 rev. N s412e mm 19. Verify that the error is within specification. 20. Set the cal factor frequency of the power sensor to 5600.5 mhz. 21. Set the mg3700a frequency to 5600.5 mhz. 22. C...

  • Page 53

    Chapter 5 — option verification 5-4 fixed wimax signal analyzer verification, option 46 and 47 s412e mm pn: 10580-00324 rev. N 5-9 7. Set the s412e as follows: a. Press the freq main menu key and set the center frequency to 2600.5 mhz. B. Press the setup submenu key and set the bandwidth to 10 mhz. ...

  • Page 54: 5-5

    5-5 mobile wimax signal analyzer verification, option 66 and 67 chapter 5 — option verification 5-10 pn: 10580-00324 rev. N s412e mm 5-5 mobile wimax signal analyzer verification, option 66 and 67 the tests in this section verify the functionality of the mobile wimax signal analyzer of the s412e. Th...

  • Page 55

    Chapter 5 — option verification 5-5 mobile wimax signal analyzer verification, option 66 and 67 s412e mm pn: 10580-00324 rev. N 5-11 procedure 1. Connect the ma2482d power sensor to the power meter and zero the sensor. 2. Set the power meter measurement mode to true rms, set averaging mode to moving...

  • Page 56

    5-5 mobile wimax signal analyzer verification, option 66 and 67 chapter 5 — option verification 5-12 pn: 10580-00324 rev. N s412e mm 16. Set the calibration factor frequency of the power sensor to 3600.5 mhz. 17. Set the mg3700a frequency to 3600.5 mhz. 18. Change the s412e center frequency to 3600....

  • Page 57

    Chapter 5 — option verification 5-5 mobile wimax signal analyzer verification, option 66 and 67 s412e mm pn: 10580-00324 rev. N 5-13 mobile wimax residual evm and frequency error tests (option 67) the tests in this section verify the function of the optional mobile wimax signal analyzer in model s41...

  • Page 58

    5-5 mobile wimax signal analyzer verification, option 66 and 67 chapter 5 — option verification 5-14 pn: 10580-00324 rev. N s412e mm k. Press the mod on/off key and verify that the led is on. Confirm that the “playing” indicator is displaying the moving pattern. L. Press the frequency key, then ente...

  • Page 59

    Chapter 5 — option verification 5-5 mobile wimax signal analyzer verification, option 66 and 67 s412e mm pn: 10580-00324 rev. N 5-15 residual evm and frequency error (5 mhz bandwidth and 5 ms frame length) 19. Set the mg3700a as follows: a. Press the yellow preset button (answer yes to the question)...

  • Page 60: 5-6

    5-6 p25/p25p2 modulation fidelity verification, option 521 and 522 chapter 5 — option verification 5-16 pn: 10580-00324 rev. N s412e mm 5-6 p25/p25p2 modulation fidelity verification, option 521 and 522 the following test is used to verify the modulation fidelity of the signal generator on the s412e...

  • Page 61

    Chapter 5 — option verification 5-6 p25/p25p2 modulation fidelity verification, option 521 and 522 s412e mm pn: 10580-00324 rev. N 5-17 23. Repeat step 21 through step 22 for the other frequencies listed in table a-28 . 24. Press the setup main menu button and set rx pattern to silence and tx patter...

  • Page 62: 5-7

    5-7 nxdn modulation fidelity verification, option 531 and 532 chapter 5 — option verification 5-18 pn: 10580-00324 rev. N s412e mm 5-7 nxdn modulation fidelity verification, option 531 and 532 the following test is used to verify the modulation fidelity of the signal generator on the s412e. This tes...

  • Page 63: 5-8

    Chapter 5 — option verification 5-8 lte signal analyzer verification, option 541 and 542 s412e mm pn: 10580-00324 rev. N 5-19 5-8 lte signal analyzer verification, option 541 and 542 the tests in this section verify the functionality of the lte signal analyzer of the s412e lmr master. The following ...

  • Page 64

    5-8 lte signal analyzer verification, option 541 and 542 chapter 5 — option verification 5-20 pn: 10580-00324 rev. N s412e mm lte channel power accuracy verification (option 541) the tests in this section verify the function of the optional lte signal analyzer in model s412e lmr master. Procedure 1....

  • Page 65

    Chapter 5 — option verification 5-8 lte signal analyzer verification, option 541 and 542 s412e mm pn: 10580-00324 rev. N 5-21 13. Record the s412e channel power reading in the 797 mhz, –50 dbm row, measured channel power column of table a-30 . 14. Calculate the channel power error by subtracting the...

  • Page 66

    5-8 lte signal analyzer verification, option 541 and 542 chapter 5 — option verification 5-22 pn: 10580-00324 rev. N s412e mm 6. Adjust the mg3700a level setting with the knob so that the power meter reads –10.0 dbm ± 0.5 db. 7. Set the s412e to lte signal analyzer mode and preset the unit. 8. Set t...

  • Page 67: 5-9

    Chapter 5 — option verification 5-9 tetra evm verification, option 581 and 582 s412e mm pn: 10580-00324 rev. N 5-23 5-9 tetra evm verification, option 581 and 582 the following test is used to verify evm of the signal generator on the s412e. This test requires the tetra options to be installed: equi...

  • Page 68

    5-10 dmr2 modulation fidelity verification, option 591 and 592 chapter 5 — option verification 5-24 pn: 10580-00324 rev. N s412e mm 5-10 dmr2 modulation fidelity verification, option 591 and 592 the following test is used to verify the modulation fidelity of the signal generator on the s412e. This t...

  • Page 69

    Chapter 5 — option verification 5-11 positive train control (ptc) verification, option 721 and 722 s412e mm pn: 10580-00324 rev. N 5-25 5-11 positive train control (ptc) verification, option 721 and 722 the following verification procedure is split into two procedures depending on the s412e’s firmwa...

  • Page 70

    5-11 positive train control (ptc) verification, option 721 and 722 chapter 5 — option verification 5-26 pn: 10580-00324 rev. N s412e mm procedure for units with firmware version v1.10 and higher 1. Connect the ac power adapter (anritsu part number 40-187-r) to the s412e and then plug the power cord ...

  • Page 71: Replacement

    S412e mm pn: 10580-00324 rev. N 6-1 chapter 6 — battery information and replacement 6-1 introduction this chapter describes the removal and replacement procedures for the instrument battery. 6-2 replaceable parts and assemblies the following replaceable parts are related to the battery pack and powe...

  • Page 72

    6-3 battery information chapter 6 — battery information and replacement 6-2 pn: 10580-00324 rev. N s412e mm unit not in use • when the s412e or the charger is not in use, disconnect it from the ac power source. • if left unused, a fully charged battery will discharge itself over time. • temperature ...

  • Page 73: 6-4

    Chapter 6 — battery information and replacement 6-4 battery pack removal and replacement s412e mm pn: 10580-00324 rev. N 6-3 6-4 battery pack removal and replacement this section provides instructions for the removal and replacing the s412e battery pack. 1. Locate the battery access door, as shown i...

  • Page 74

    6-4 battery pack removal and replacement chapter 6 — battery information and replacement 6-4 pn: 10580-00324 rev. N s412e mm 4. With the battery access door completely removed, grasp the battery lanyard and pull the battery straight out of the unit, as illustrated in figure 6-2 . 5. Replacement is t...

  • Page 75: 7-1

    S412e mm pn: 10580-00324 rev. N 7-1 chapter 7 — assembly replacement 7-1 replaceable parts list refer to table 1-5, “list of replaceable parts and assemblies for the list of replaceable parts. Refer to the following sections for basic replacement instructions. 7-2 opening the s412e case this procedu...

  • Page 76

    7-2 opening the s412e case chapter 7 — assembly replacement 7-2 pn: 10580-00324 rev. N s412e mm 2. Remove the top and bottom bumpers ( figure 7-1 ) to expose the screw holes on the back of the unit remove the nut and washer on the gps connector. 3. Place the s412e face down on a stable work surface ...

  • Page 77

    Chapter 7 — assembly replacement 7-2 opening the s412e case s412e mm pn: 10580-00324 rev. N 7-3 6. Lay the s412e flat and remove the battery connector cable between the two halves ( figure 7-3 ). 7. Closing the case is the reverse of opening. Ensure all cables are properly seated and none are pinche...

  • Page 78: 7-3

    7-3 pcb assembly replacement chapter 7 — assembly replacement 7-4 pn: 10580-00324 rev. N s412e mm 7-3 pcb assembly replacement this section describes the removal and replacement of the spa and main/vna boards which are attached to each other and attached to the s412e case. 1. Open the case as descri...

  • Page 79: 7-4

    Chapter 7 — assembly replacement 7-4 spa assembly replacement s412e mm pn: 10580-00324 rev. N 7-5 7-4 spa assembly replacement this section describes the removal of the spa assembly board. 1. Open the case as described in section 7-2 “opening the s412e case” . 2. Remove the pcb assembly from the fro...

  • Page 80: 7-5

    7-5 spa and mb/vna n connector replacement chapter 7 — assembly replacement 7-6 pn: 10580-00324 rev. N s412e mm 7-5 spa and mb/vna n connector replacement this procedure provides instructions for replacing the n connector attached to the spa assembly or mb/vna assembly. 1. Open the case as described...

  • Page 81: 7-6

    Chapter 7 — assembly replacement 7-6 gps (option 31) replacement s412e mm pn: 10580-00324 rev. N 7-7 7-6 gps (option 31) replacement this procedure provides instructions for removing and replacing the gps module. 1. Open the case as described in section 7-2 “opening the s412e case” . 2. Remove the p...

  • Page 82: 7-7

    7-7 signal generator replacement chapter 7 — assembly replacement 7-8 pn: 10580-00324 rev. N s412e mm 7-7 signal generator replacement this procedure provides instructions for removing and replacing the signal generator pcb assembly. This procedure requires a cable removal tool, anritsu part number ...

  • Page 83

    Chapter 7 — assembly replacement 7-7 signal generator replacement s412e mm pn: 10580-00324 rev. N 7-9 5. Turn over the pcb and remove the 4 screws shown in figure 7-10 . 6. Carefully slide the top plate off of the connectors being careful not to damage the gps cable (if connected). 7. Gently pull up...

  • Page 84

    7-7 signal generator replacement chapter 7 — assembly replacement 7-10 pn: 10580-00324 rev. N s412e mm 8. Remove the cable shown in figure 7-11 using the cable removal tool, part number 783-1399. When inserting the cable onto the new signal generator make sure the cable connector is parallel to the ...

  • Page 85: 7-8

    Chapter 7 — assembly replacement 7-8 main/vna pcb assembly replacement s412e mm pn: 10580-00324 rev. N 7-11 7-8 main/vna pcb assembly replacement this procedure provides instructions for removing and replacing the main/vna pcb assembly. 1. Open the case as described in section 7-2 “opening the s412e...

  • Page 86

    7-10 lcd assembly replacement chapter 7 — assembly replacement 7-12 pn: 10580-00324 rev. N s412e mm 7-10 lcd assembly replacement this procedure provides instructions for removing and replacing the liquid crystal display (lcd) once the main/vna pcb assembly has been separated from the s412e. 1. Open...

  • Page 87

    Chapter 7 — assembly replacement 7-10 lcd assembly replacement s412e mm pn: 10580-00324 rev. N 7-13 4. Turn the lcd assembly over and disconnect the front half of the case from the lcd assembly ( figure 7-14 ). 5. Use a phillips screw driver to remove the four screws securing the lcd to the housing ...

  • Page 88

    7-11 lcd backlight pcb removal and replacement chapter 7 — assembly replacement 7-14 pn: 10580-00324 rev. N s412e mm 6. Disconnect the lcd backlight cable from the lcd backlight pcb. 7. Disconnect the lcd cable from the side of the lcd. 8. Carefully remove the lcd. 9. Reverse the above steps to inst...

  • Page 89

    Chapter 7 — assembly replacement 7-12 keypad and keypad pcb replacement s412e mm pn: 10580-00324 rev. N 7-15 7-12 keypad and keypad pcb replacement this procedure provides instructions for removing and replacing the keypad and the keypad pcb. 1. Open the case as described in section 7-2 “opening the...

  • Page 90

    7-13 touchscreen replacement chapter 7 — assembly replacement 7-16 pn: 10580-00324 rev. N s412e mm 7-13 touchscreen replacement this procedure provides instructions for removing and replacing the touchscreen. 1. Open the case as described in section 7-2 “opening the s412e case” . 2. Remove the main/...

  • Page 91: 8-1

    S412e mm pn: 10580-00324 rev. N 8-1 chapter 8 — troubleshooting 8-1 introduction this chapter describes the primary troubleshooting operations that can be performed. Perform the troubleshooting suggestions in the order in which they are listed. Only qualified service personnel should replace interna...

  • Page 92: 8-3

    8-3 operating problems chapter 8 — troubleshooting 8-2 pn: 10580-00324 rev. N s412e mm boot-up self test fails the unit boot-up self test fails with an error message: 1. Perform a factory defaults reset. This done by holding the esc key down while pressing the on/off key to power on the unit. During...

  • Page 93

    Chapter 8 — troubleshooting 8-3 operating problems s412e mm pn: 10580-00324 rev. N 8-3 option 31, gps problems unable to acquire gps fix 1. Verify that the gps antenna is in direct line-of-sight relationship or is placed outside without any obstructions. 2. If the anritsu 2000-1528-r gps antenna is ...

  • Page 94

    8-3 operating problems chapter 8 — troubleshooting 8-4 pn: 10580-00324 rev. N s412e mm.

  • Page 95: Appendix A — Test Records

    S412e mm pn: 10580-00324 rev. N a-1 appendix a — test records a-1 introduction this appendix provides test records that can be used to record the performance of the s412e lmr master. Make a copy of the following test record pages and document the measured values each time performance verification is...

  • Page 96: A-2

    A-2 test records for vna verification appendix a — test records a-2 pn: 10580-00324 rev. N s412e mm s412e firmware rev: ______________ operator: ____________________ date: _____________ serial number: _______________ options: ___________________________________________________ a-2 test records for v...

  • Page 97: A-3

    Appendix a — test records a-3 test records for spectrum analyzer verification s412e mm pn: 10580-00324 rev. N a-3 s412e firmware rev: ______________ operator: ____________________ date: _____________ serial number: _______________ options: ___________________________________________________ a-3 test...

  • Page 98

    A-3 test records for spectrum analyzer verification appendix a — test records a-4 pn: 10580-00324 rev. N s412e mm s412e firmware rev: ______________ operator: ____________________ date: _____________ serial number: _______________ options: ___________________________________________________ table a-...

  • Page 99

    Appendix a — test records a-3 test records for spectrum analyzer verification s412e mm pn: 10580-00324 rev. N a-5 s412e firmware rev: ______________ operator: ____________________ date: _____________ serial number: _______________ options: ___________________________________________________ table a-...

  • Page 100

    A-3 test records for spectrum analyzer verification appendix a — test records a-6 pn: 10580-00324 rev. N s412e mm s412e firmware rev: ______________ operator: ____________________ date: _____________ serial number: _______________ options: ___________________________________________________ table a-...

  • Page 101

    Appendix a — test records a-3 test records for spectrum analyzer verification s412e mm pn: 10580-00324 rev. N a-7 s412e firmware rev: ______________ operator: ____________________ date: _____________ serial number: _______________ options: ___________________________________________________ table a-...

  • Page 102

    A-3 test records for spectrum analyzer verification appendix a — test records a-8 pn: 10580-00324 rev. N s412e mm s412e firmware rev: ______________ operator: ____________________ date: _____________ serial number: _______________ options: ___________________________________________________ table a-...

  • Page 103

    Appendix a — test records a-3 test records for spectrum analyzer verification s412e mm pn: 10580-00324 rev. N a-9 s412e firmware rev: ______________ operator: ____________________ date: _____________ serial number: _______________ options: ___________________________________________________ table a-...

  • Page 104

    A-3 test records for spectrum analyzer verification appendix a — test records a-10 pn: 10580-00324 rev. N s412e mm s412e firmware rev: ______________ operator: ____________________ date: _____________ serial number: _______________ options: ___________________________________________________ table a...

  • Page 105

    Appendix a — test records a-3 test records for spectrum analyzer verification s412e mm pn: 10580-00324 rev. N a-11 s412e firmware rev: ______________ operator: ____________________ date: _____________ serial number: _______________ options: ___________________________________________________ 50 –50 ...

  • Page 106

    A-3 test records for spectrum analyzer verification appendix a — test records a-12 pn: 10580-00324 rev. N s412e mm s412e firmware rev: ______________ operator: ____________________ date: _____________ serial number: _______________ options: ___________________________________________________ 1000 –5...

  • Page 107

    Appendix a — test records a-3 test records for spectrum analyzer verification s412e mm pn: 10580-00324 rev. N a-13 s412e firmware rev: ______________ operator: ____________________ date: _____________ serial number: _______________ options: ___________________________________________________ 3000 (u...

  • Page 108

    A-3 test records for spectrum analyzer verification appendix a — test records a-14 pn: 10580-00324 rev. N s412e mm s412e firmware rev: ______________ operator: ____________________ date: _____________ serial number: _______________ options: ___________________________________________________ 5990 (u...

  • Page 109

    Appendix a — test records a-3 test records for spectrum analyzer verification s412e mm pn: 10580-00324 rev. N a-15 s412e firmware rev: ______________ operator: ____________________ date: _____________ serial number: _______________ options: ___________________________________________________ table a...

  • Page 110: A-4

    A-4 test records for signal generator verification appendix a — test records a-16 pn: 10580-00324 rev. N s412e mm s412e firmware rev: ______________ operator: ____________________ date: _____________ serial number: _______________ options: ___________________________________________________ a-4 test...

  • Page 111: A-5

    Appendix a — test records a-5 test records for options verification s412e mm pn: 10580-00324 rev. N a-17 s412e firmware rev: ______________ operator: ____________________ date: _____________ serial number: _______________ options: ___________________________________________________ a-5 test records ...

  • Page 112

    A-5 test records for options verification appendix a — test records a-18 pn: 10580-00324 rev. N s412e mm s412e firmware rev: ______________ operator: ____________________ date: _____________ serial number: _______________ options: ___________________________________________________ mobile wimax sign...

  • Page 113

    Appendix a — test records a-5 test records for options verification s412e mm pn: 10580-00324 rev. N a-19 s412e firmware rev: ______________ operator: ____________________ date: _____________ serial number: _______________ options: ___________________________________________________ p25/p25p2 verific...

  • Page 114

    A-5 test records for options verification appendix a — test records a-20 pn: 10580-00324 rev. N s412e mm s412e firmware rev: ______________ operator: ____________________ date: _____________ serial number: _______________ options: ___________________________________________________ table a-28. Optio...

  • Page 115

    Appendix a — test records a-5 test records for options verification s412e mm pn: 10580-00324 rev. N a-21 s412e firmware rev: ______________ operator: ____________________ date: _____________ serial number: _______________ options: ___________________________________________________ lte signal analyz...

  • Page 116

    A-5 test records for options verification appendix a — test records a-22 pn: 10580-00324 rev. N s412e mm s412e firmware rev: ______________ operator: ____________________ date: _____________ serial number: _______________ options: ___________________________________________________ tetra verificatio...

  • Page 117

    Appendix a — test records a-5 test records for options verification s412e mm pn: 10580-00324 rev. N a-23 s412e firmware rev: ______________ operator: ____________________ date: _____________ serial number: _______________ options: ___________________________________________________ table a-35. Units...

  • Page 118

    A-5 test records for options verification appendix a — test records a-24 pn: 10580-00324 rev. N s412e mm s412e firmware rev: ______________ operator: ____________________ date: _____________ serial number: _______________ options: ___________________________________________________.

  • Page 119: Index

    A to r s412e mm pn: 10580-00324 rev. N index-1 index a ac to dc power adapter . . . . . . . . . . . . . . . . . . . 1-6 adapter (for option 11) . . . . . . . . . . . . . . . . . . . . 1-4 anritsu pn 10580-00220 user guide . . . . . . . . . . . . . . . . 1-2 10580-00221 programming manual . . . . . ....

  • Page 120

    S to v index-2 pn: 10580-00324 rev. N s412e mm s safety symbols for safety . . . . . . . . . . . . . . . . . . . . . . . . Safety-2 in manuals . . . . . . . . . . . . . . . . . . . . . . . Safety-1 on equipment . . . . . . . . . . . . . . . . . . . . Safety-1 self test fails . . . . . . . . . . . . ...

  • Page 122

    Anritsu utilizes recycled paper and environmentally conscious inks and toner. Anritsu company 490 jarvis drive morgan hill, ca 95037-2809 usa http://www.Anritsu.Com.