Maintenance manual site master cable and antenna analyzer with spectrum analyzer s331e, 2 mhz to 4 ghz s332e, 2 mhz to 4 ghz, spectrum analyzer, 9 khz to 4 ghz s361e, 2 mhz to 6 ghz s362e, 2 mhz to 6 ghz, spectrum analyzer, 9 khz to 6 ghz anritsu company 490 jarvis drive morgan hill, ca 95037-2809 u...
Trademark acknowledgments windows, windows xp, and windows vista are all registered trademarks of microsoft corporation. Acrobat reader is a registered trademark of adobe corporation. Site master is a trademark of anritsu company. Notice anritsu company has prepared this manual for use by anritsu co...
S3xxe mm pn: 10580-00253 rev. J safety-1 safety symbols to prevent the risk of personal injury or loss related to equipment malfunction, anritsu company uses the following symbols to indicate safety-related information. For your own safety, please read the information carefully before operating the ...
Safety-2 pn: 10580-00253 rev. J s3xxe mm for safety warning always refer to the operation manual when working near locations at which the alert mark, shown on the left, is attached. If the operation, etc., is performed without heeding the advice in the operation manual, there is a risk of personal i...
S3xxe mm pn: 10580-00253 rev. J safety-3 for safety warning this equipment is supplied with a rechargeable battery that could potentially leak hazardous compounds into the environment. These hazardous compounds present a risk of injury or loss due to exposure. Anritsu company recommends removing the...
Safety-4 pn: 10580-00253 rev. J s3xxe mm.
S3xxe mm pn: 10580-00253 rev. J contents-1 table of contents chapter 1—general information 1-1 introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1 1-2 anritsu customer service centers . . . . . . . . . . . ....
Contents-2 pn: 10580-00253 rev. J s3xxe mm table of contents (continued) chapter 6—assembly replacement 6-1 opening the site master case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1 6-2 pcb assembly removal. . . . . . . . . . . . . . . . . . ....
S3xxe mm pn: 10580-00253 rev. J 1-1 chapter 1 — general information 1-1 introduction this manual provides maintenance instructions for anritsu site master models s331e, s332e, s361e and s362e. The manual includes: • general information in this chapter, including: • lists of necessary test equipment ...
1-3 recommended test equipment general information 1-2 pn: 10580-00253 rev. J s3xxe mm 1-3 recommended test equipment the following test equipment is recommended for use in testing and maintaining anritsu site master model s3xxe. Table 1-1 is a list of test equipment that is required for verifying t...
General information 1-3 recommended test equipment s3xxe mm pn: 10580-00253 rev. J 1-3 table 1-2. Test equipment required for cable and antenna analyzer verification instrument critical specification recommended manufacturer/model frequency counter frequency: 2 ghz anritsu model mf2412b open/short f...
1-4 replaceable parts general information 1-4 pn: 10580-00253 rev. J s3xxe mm 1-4 replaceable parts table 1-4. List of replaceable parts part number description nd70931 s331e mb/vna pcb assembly (units without option 21) a s/n 1129004 nd73215 s331e mb/vna pcb assembly (units without option 21) a s/n...
General information 1-4 replaceable parts s3xxe mm pn: 10580-00253 rev. J 1-5 3-67304-9 model s331e id label 3-67304-6 model s332e id label 3-67304-5 model s361e id label 3-67304-7 model s362e id label nd70320 gps module (opt 31) nd72101 ethernet pcb assy (opt 411) nd82421 cpri pcb assy (opt 751) 3-...
1-6 pn: 10580-00253 rev. J s3xxe mm 3-71625-1 cable, lcd to keypad, units with lcd display 3-15-165 3-70675-1 cable, lcd to keypad, units with lcd display 3-15-174 3-72621-4 cable, lcd to main pcb, units with lcd display 3-15-147 and 3-15-165 3-70674-4 cable, lcd to main pcb, units with lcd display ...
S3xxe mm pn: 10580-00253 rev. J 2-1 chapter 2 — spectrum analyzer verification 2-1 frequency accuracy verification and adjustment the following procedure is used to verify and adjust the cw frequency accuracy of the spectrum analyzer in the s332e and s362e site master. Adjustment to the frequency ac...
2-1 frequency accuracy verification and adjustment spectrum analyzer verification 2-2 pn: 10580-00253 rev. J s3xxe mm step 21 , and proceed to step 22 . If the marker frequency is outside the specification, proceed to the next step for adjustment. 19. Perform step 19 through step 21 only if the prev...
Spectrum analyzer verification 2-2 single side band (ssb) phase noise verification s3xxe mm pn: 10580-00253 rev. J 2-3 2-2 single side band (ssb) phase noise verification this test is used to verify the single side band (ssb) phase noise of the spectrum analyzer in the s332e and s362e site master. E...
2-3 spurious response (second harmonic distortion) verification spectrum analyzer verification 2-4 pn: 10580-00253 rev. J s3xxe mm 2-3 spurious response (second harmonic distortion) verification the following test is used to verify the input related spurious response of the spectrum analyzer in the ...
Spectrum analyzer verification 2-3 spurious response (second harmonic distortion) verification s3xxe mm pn: 10580-00253 rev. J 2-5 25. Press the freq soft key and press the center freq soft key. 26. Use the keypad to enter 100.2 and press the mhz soft key. 27. Press the shift key and then press the ...
2-4 resolution bandwidth accuracy verification spectrum analyzer verification 2-6 pn: 10580-00253 rev. J s3xxe mm 2-4 resolution bandwidth accuracy verification the following test is used to verify the resolution bandwidth accuracy of the spectrum analyzer in the s332e and s362e site master. Equipme...
Spectrum analyzer verification 2-5 spectrum analyzer absolute amplitude accuracy verification s3xxe mm pn: 10580-00253 rev. J 2-7 2-5 spectrum analyzer absolute amplitude accuracy verification the tests in the following two sections verify the absolute amplitude accuracy of the spectrum analyzer in ...
2-5 spectrum analyzer absolute amplitude accuracy verification spectrum analyzer verification 2-8 pn: 10580-00253 rev. J s3xxe mm test setup components characterization 1. Turn on the ml2438a power meter, the mg3692x signal source, and the s332e or s362e site master. 2. On the power meter, press the...
Spectrum analyzer verification 2-5 spectrum analyzer absolute amplitude accuracy verification s3xxe mm pn: 10580-00253 rev. J 2-9 measuring the unit for 50 mhz amplitude accuracy 1. Remove sensor a, add the adapter and connect it to the spectrum analyzer rf in connector of the s332e or s362e site ma...
2-5 spectrum analyzer absolute amplitude accuracy verification spectrum analyzer verification 2-10 pn: 10580-00253 rev. J s3xxe mm 11. Press the amplitude soft key and then press the reference level soft key. 12. Use the keypad to enter 10 and press the dbm soft key. 13. Press the atten lvl soft key...
Spectrum analyzer verification 2-5 spectrum analyzer absolute amplitude accuracy verification s3xxe mm pn: 10580-00253 rev. J 2-11 test setup component characterization 1. Connect both ma2442d power sensors to the power meter and calibrate the sensors. 2. Connect the equipment as shown in figure 2-3...
2-5 spectrum analyzer absolute amplitude accuracy verification spectrum analyzer verification 2-12 pn: 10580-00253 rev. J s3xxe mm setup measuring amplitude accuracy across frequency 1. Connect the equipment as shown in figure 2-4 . 2. Set the s332e or s362e to spectrum analyzer mode and then preset...
Spectrum analyzer verification 2-5 spectrum analyzer absolute amplitude accuracy verification s3xxe mm pn: 10580-00253 rev. J 2-13 9. Set the power meter to display channel b. Press the sensor key, the cal factor soft key, and then the freq soft key. Use the keypad to enter the value matching the fr...
2-6 residual spurious response verification spectrum analyzer verification 2-14 pn: 10580-00253 rev. J s3xxe mm 2-6 residual spurious response verification the following two tests are used to verify the residual spurious response of the spectrum analyzer of the s332e and s362e site master and is per...
Spectrum analyzer verification 2-6 residual spurious response verification s3xxe mm pn: 10580-00253 rev. J 2-15 residual spurious response test with preamp on equipment required • anritsu 28n50-2 50 ohm termination procedure 1. Connect the 50 ohm termination to the s332e or s362e spectrum analyzer r...
2-7 displayed average noise level (danl) spectrum analyzer verification 2-16 pn: 10580-00253 rev. J s3xxe mm 2-7 displayed average noise level (danl) the following test is used to verify the displayed average noise level (danl) of the spectrum analyzer systems in the s332e and s362e site master. Thi...
Spectrum analyzer verification 2-7 displayed average noise level (danl) s3xxe mm pn: 10580-00253 rev. J 2-17 for example, if the marker shows a value of –100 dbm at 100 khz rbw, the calculated value at 10 hz rbw is –140 dbm. 23. Enter the calculated values in the test records. Use the calculated for...
2-8 third order intercept (toi) verification spectrum analyzer verification 2-18 pn: 10580-00253 rev. J s3xxe mm 2-8 third order intercept (toi) verification the following test verifies the third order intercept point (also known as toi or ip3) of the spectrum analyzer in the s332e and s362. Equipme...
Spectrum analyzer verification 2-8 third order intercept (toi) verification s3xxe mm pn: 10580-00253 rev. J 2-19 19. Choose the larger of the two values from step 16 and step 18 , and put this value into the following equation as the “max” variable. Toi = – 20 + [(– 20 – max) / 2] dbm 20. Record the...
2-8 third order intercept (toi) verification spectrum analyzer verification 2-20 pn: 10580-00253 rev. J s3xxe mm.
S3xxe mm pn: 10580-00253 rev. J 3-1 chapter 3 — cable and antenna analyzer verification 3-1 frequency accuracy verification the following test is used to verify the cw frequency accuracy of the rf source on the s3xxe in cable and antenna analyzer mode. Equipment required • frequency counter frequenc...
3-2 return loss accuracy verification cable and antenna analyzer verification 3-2 pn: 10580-00253 rev. J s3xxe mm 3-2 return loss accuracy verification the following test is used to verify the accuracy of return loss measurements on the s3xxe in cable and antenna analyzer mode. Equipment required • ...
S3xxe mm pn: 10580-00253 rev. J 4-1 chapter 4 — option verification 4-1 introduction this chapter describes the verification process for options available on the s3xxe site master. 4-2 option 10, bias tee verification this test verifies that the optional bias tee in the cable and antenna analyzer of...
4-2 option 10, bias tee verification option verification 4-2 pn: 10580-00253 rev. J s3xxe mm high current test verification the tests in this section verify the bias tee, option 10, high current operation of the s3xxe in cable and antenna analyzer mode. Equipment required • anritsu 40-187-r external...
Option verification 4-2 option 10, bias tee verification s3xxe mm pn: 10580-00253 rev. J 4-3 fault verification the tests in this section verify the bias tee, option 10, fault condition of the s3xxe in cable and antenna analyzer mode. Equipment required • anritsu 40-187-r external power supply • anr...
4-3 option 21, transmission measurement, system dynamic range option verification 4-4 pn: 10580-00253 rev. J s3xxe mm 4-3 option 21, transmission measurement, system dynamic range the following test is used to verify the system dynamic range for site masters with option 21, 2-port transmission measu...
Option verification 4-4 option 29, power meter level accuracy s3xxe mm pn: 10580-00253 rev. J 4-5 4-4 option 29, power meter level accuracy the following test verifies the level accuracy of the s332e or s362e power meter for site masters with option 29 installed. Equipment required • anritsu mg3692x...
4-4 option 29, power meter level accuracy option verification 4-6 pn: 10580-00253 rev. J s3xxe mm procedure component characterization: 1. Connect both ma2442d power sensors to the power meter and calibrate the sensors. 2. Connect the model 1870a power splitter to the mg3692a/b output and sensor b t...
Option verification 4-5 option 31, gps verification s3xxe mm pn: 10580-00253 rev. J 4-7 2. Verify that the s3xxe is in the power meter mode and preset the unit. 3. Set the s3xxe span to 3 mhz. 4. Set the s3xxe center frequency to 50 mhz. 5. Adjust the mg3692a/b power so that the power meter sensor b...
4-5 option 31, gps verification option verification 4-8 pn: 10580-00253 rev. J s3xxe mm 7. Connect the external 10 mhz reference to the anritsu mg3692x synthesized signal generator. 8. Connect the output of the synthesized signal generator to the spectrum analyzer rf in of the s332e or s362e. 9. Set...
Option verification 4-5 option 31, gps verification s3xxe mm pn: 10580-00253 rev. J 4-9 12. Press the gps info soft key. Record the gps antenna current reading in the measured value column of table a-21 . Verify that it is within specification..
4-5 option 31, gps verification option verification 4-10 pn: 10580-00253 rev. J s3xxe mm.
S3xxe mm pn: 10580-00253 rev. J 5-1 chapter 5 — battery information 5-1 general information the following information relates to the care and handling of the anritsu battery pack and lithium-ion batteries. • the battery supplied with the site master may need charging before use. Before using the sit...
5-2 battery pack removal and replacement battery information 5-2 pn: 10580-00253 rev. J s3xxe mm 5-2 battery pack removal and replacement this section provides instructions for the removal and replacement of the site master battery pack. 1. Locate the battery access door illustrated in figure 5-1 . ...
Battery information 5-2 battery pack removal and replacement s3xxe mm pn: 10580-00253 rev. J 5-3 4. With the battery access door completely removed, grasp the battery lanyard and pull the battery straight out of the unit, as illustrated in figure 5-3 . 5. Replacement is the opposite of removal. Note...
5-2 battery pack removal and replacement battery information 5-4 pn: 10580-00253 rev. J s3xxe mm.
S3xxe mm pn: 10580-00253 rev. J 6-1 chapter 6 — assembly replacement 6-1 opening the site master case this procedure provides instructions for opening the site master case. With the case opened, the internal assemblies can be removed and replaced, as detailed in the following sections. 1. Remove the...
6-1 opening the site master case assembly replacement 6-2 pn: 10580-00253 rev. J s3xxe mm 4. Use a phillips screwdriver to remove the six screws securing the two halves of the site master case together ( figure 6-2 ). 5. Carefully lift up on the side of the case indicated above and begin to separate...
Assembly replacement 6-2 pcb assembly removal s3xxe mm pn: 10580-00253 rev. J 6-3 6-2 pcb assembly removal this procedure provides instructions to remove the motherboard/vna and spa (if installed) from the site master case. 1. Open the case as described in section 6-1 “opening the site master case” ...
6-3 spa assembly replacement assembly replacement 6-4 pn: 10580-00253 rev. J s3xxe mm 6-3 spa assembly replacement this procedure provides instructions to remove the spa assembly. 1. Open the case as described in section 6-1 “opening the site master case” . 2. Remove the pcb assembly from the front ...
Assembly replacement 6-4 spa and mb/vna n connector replacement s3xxe mm pn: 10580-00253 rev. J 6-5 6-4 spa and mb/vna n connector replacement this procedure provides instructions to replace the n connector attached to the spa (if installed) or mb/vna assembly. 1. Open the case as described in secti...
6-5 gps (opt. 31) replacement assembly replacement 6-6 pn: 10580-00253 rev. J s3xxe mm 6-5 gps (opt. 31) replacement this procedure provides instructions to replace the gps assembly. 1. Open the case as described in section 6-1 “opening the site master case” . 2. Remove the pcb assembly from the fro...
Assembly replacement 6-6 option assembly replacement s3xxe mm pn: 10580-00253 rev. J 6-7 6-6 option assembly replacement this procedure provides instructions to replace the ethernet (option 411) or cpri (option 751) assembly. 1. Open the case as described in section 6-1 “opening the site master case...
6-8 fan assembly replacement assembly replacement 6-8 pn: 10580-00253 rev. J s3xxe mm 6-8 fan assembly replacement this procedure provides instructions to replace the fan assembly. 1. Open the case as described in section 6-1 “opening the site master case” . 2. Remove the pcb assembly from the front...
Assembly replacement 6-9 lcd assembly replacement s3xxe mm pn: 10580-00253 rev. J 6-9 6-9 lcd assembly replacement this procedure provides instructions to replace the liquid crystal display (lcd). 1. Open the case as described in section 6-1 “opening the site master case” . 2. Remove the pcb assembl...
6-9 lcd assembly replacement assembly replacement 6-10 pn: 10580-00253 rev. J s3xxe mm 4. Turn the lcd assembly over and disconnect the front half of the case from the lcd assembly ( figure 6-12 ). 5. Use a phillips screwdriver to remove the four screws securing the lcd to the housing, see figure 6-...
Assembly replacement 6-10 lcd backlight pcb removal and replacement s3xxe mm pn: 10580-00253 rev. J 6-11 6. Disconnect the lcd backlight cable from the lcd backlight pcb. 7. Disconnect the lcd cable from the side of the lcd. 8. Carefully remove the lcd. 9. Reverse the above steps to install the repl...
6-11 keypad and keypad pcb replacement assembly replacement 6-12 pn: 10580-00253 rev. J s3xxe mm 6-11 keypad and keypad pcb replacement this procedure provides instructions to replace the rubber keypad and the keypad pcb. 1. Open the case as described in section 6-1 “opening the site master case” . ...
Assembly replacement 6-12 touch screen replacement s3xxe mm pn: 10580-00253 rev. J 6-13 6-12 touch screen replacement this procedure provides instructions to replace the touch screen. 1. Open the case as described in section 6-1 “opening the site master case” . 2. Remove the pcb assembly from the fr...
6-12 touch screen replacement assembly replacement 6-14 pn: 10580-00253 rev. J s3xxe mm.
S3xxe mm pn: 10580-00253 rev. J 7-1 chapter 7 — troubleshooting 7-1 introduction this chapter describes the primary troubleshooting operations that can be performed by all anritsu service centers. Perform the troubleshooting suggestions in the order they are listed. Operators of the mp1026b should r...
7-3 other problems troubleshooting 7-2 pn: 10580-00253 rev. J s3xxe mm 7-3 other problems touch screen problems: unit boots correctly, but the touch screen is unresponsive. 1. The touch screen may have lost its calibration data. Press shift then 0 to enter the touch screen calibration procedure. Fol...
S3xxe mm pn: 10580-00253 rev. J a-1 appendix a — test records this appendix provides test records that can be used to record the performance of the s3xxe. Anritsu recommends that you make a copy of the following test record pages and document the measurements each time a performance verification is ...
A-1 test records for spectrum analyzer verification a-2 pn: 10580-00253 rev. J s3xxe mm a-1 test records for spectrum analyzer verification s3____e firmware rev: ______________ operator: ____________________ date: _____________ serial number: _______________ options: ________________________________...
A-1 test records for spectrum analyzer verification s3xxe mm pn: 10580-00253 rev. J a-3 s3____e firmware rev: ______________ operator: _________________ date: _____________ serial number: _______________ options: ___________________________________________________ table a-4. Spectrum analyzer resolu...
A-1 test records for spectrum analyzer verification a-4 pn: 10580-00253 rev. J s3xxe mm s3____e firmware rev: ______________ operator: _________________ date: _____________ serial number: _______________ options: ___________________________________________________ table a-6. Spectrum analyzer 50 mhz...
A-1 test records for spectrum analyzer verification s3xxe mm pn: 10580-00253 rev. J a-5 5000 mhz (s362e only) dbm dbm dbm 5990 mhz (s362e only) dbm dbm dbm table a-7. Spectrum analyzer absolute amplitude accuracy across frequency setup table frequency required sensor b reading for –2 dbm @ attenuato...
A-1 test records for spectrum analyzer verification a-6 pn: 10580-00253 rev. J s3xxe mm s3____e firmware rev: ______________ operator: _________________ date: _____________ serial number: _______________ options: ___________________________________________________ table a-8. Spectrum analyzer absolu...
A-1 test records for spectrum analyzer verification s3xxe mm pn: 10580-00253 rev. J a-7 100 –50 –40 15 on ±1.25 –30 –20 0 off ±1.25 –30 –20 5 off ±1.25 –30 –20 10 off ±1.25 –30 –20 20 off ±1.25 –2 0 30 off ±1.25 –2 0 40 off ±1.25 –2 0 50 off ±1.25 –2 0 55 off ±1.25 500 –50 –40 15 on ±1.25 –30 –20 0 ...
A-1 test records for spectrum analyzer verification a-8 pn: 10580-00253 rev. J s3xxe mm 2000 –50 –40 15 on ±1.25 –30 –20 0 off ±1.25 –30 –20 5 off ±1.25 –30 –20 10 off ±1.25 –30 –20 20 off ±1.25 –2 0 30 off ±1.25 –2 0 40 off ±1.25 –2 0 50 off ±1.25 –2 0 55 off ±1.25 3000 –50 –40 15 on ±1.25 –30 –20 ...
A-1 test records for spectrum analyzer verification s3xxe mm pn: 10580-00253 rev. J a-9 5000 (s362e only) –50 –40 15 on ±1.50 –30 –20 0 off ±1.50 –30 –20 5 off ±1.50 –30 –20 10 off ±1.50 –30 –20 20 off ±1.50 –2 0 30 off ±1.50 –2 0 40 off ±1.50 –2 0 50 off ±1.50 –2 0 55 off ±1.50 5990 (s362eo nly) –5...
A-1 test records for spectrum analyzer verification a-10 pn: 10580-00253 rev. J s3xxe mm s3____e firmware rev: ______________ operator: _________________ date: _____________ serial number: _______________ options: ___________________________________________________ table a-9. Spectrum analyzer resid...
A-1 test records for spectrum analyzer verification s3xxe mm pn: 10580-00253 rev. J a-11 s3____e firmware rev: ______________ operator: _________________ date: _____________ serial number: _______________ options: ___________________________________________________ table a-11. Spectrum analyzer danl...
A-2 test records for cable and antenna analyzer verification a-12 pn: 10580-00253 rev. J s3xxe mm a-2 test records for cable and antenna analyzer verification s3____e firmware rev: ______________ operator: _________________ date: _____________ serial number: _______________ options: ________________...
A-3 test records for options verification s3xxe mm pn: 10580-00253 rev. J a-13 a-3 test records for options verification s3____e firmware rev: ______________ operator: _________________ date: _____________ serial number: _______________ options: ___________________________________________________ ta...
A-3 test records for options verification a-14 pn: 10580-00253 rev. J s3xxe mm s3____e firmware rev: ______________ operator: _________________ date: _____________ serial number: _______________ options: ___________________________________________________ table a-18. Option 29, characterization char...
A-3 test records for options verification s3xxe mm pn: 10580-00253 rev. J a-15 table a-21. Option 31, gps receiver bias tee verification voltage setting measured value specification 3.3 v ma 32 ma ± 15 % (27.2 ma to 36.8 ma) 5.0 v ma 55.6 ma ± 15 % (47.3 ma to 63.9 ma).
A-3 test records for options verification a-16 pn: 10580-00253 rev. J s3xxe mm.
Anritsu utilizes recycled paper and environmentally conscious inks and toner. Anritsu company 490 jarvis drive morgan hill, ca 95037-2809 usa http://www.Anritsu.Com.