Bacharach Halogen Monitor Instruction - page 25
Halogen Monitor
Operation
Instruction 3015-4584 Rev 9
19
•
If HyperTerminal has never been run in association with the monitor, then a new
HyperTerminal connection needs to be made as follows:
a) Click
Start
; click
Run
. In the Run box, type “hypertrm.exe
”
and click
OK.
The
New
Connection
dialog box should appear.
b) Type in a New Connection Name. Click
OK
to select the default icon (if desired, select a
different icon before clicking OK). The “
Connect To”
dialog box should appear.
c) Click on
Connect from the
drop-down menu and choose the
COM
port to which the
analyzer is connected. Then click
OK
to display the
COM Properties
dialog box for that
COM port.
d) Set Port Settings to:
- Bits per second: ........ 19200
- Data bits: .......................... 8
- Parity: ........................ None
- Stop bits: .......................... 1
- Flow control: .............. None
Then click
OK
to display HyperTerminal’s main window.
e) Click File; click Save. This creates a file with an .ht extension in the folder C:\Program
Files\Accessories\Hyper Terminal, using the name that was entered in Step b.
3. After HyperTerminal is running: click
Transfer
, click
Capture Text
, and then click
Start
to accept
the default drive\directory\filename to which all received data will be stored as an ASCII text file.
NOTE:
By default, the captured text file will be stored on the hard drive at
C:\Program Files\Accessories\HyperTerminal\CAPTURE.TXT. If desired, click
Browse
to select a different drive, directory and filename. The next time data is
captured, it will be stored under the same location and filename.
NOTE:
Changing the filename extension to “CSV” (comma separated variable) will
allow the file to be directly opened in the spreadsheet program. For example, double
clicking the filename CAPTURE.CSV in Windows Explorer will automatically start
Microsoft Excel and load the file.
4. From the
Data Display
screen, press both
the
Keypad
Left
and
Right
buttons at the same time
to display the first
Function Menu
screen. Next, use the
Keypad
buttons to move the arrow (>)
until it points to
PC-DUMP
, and then press the
ENTER
button to begin the download process.
5. Observe that the monitor’s screen goes blank and that HyperTerminal displays the data as it is
being received. After all data has been transmitted, the Refrigerant Monitor re-displays the
Function Menu
screen.
6. To stop capturing data and save it, click
Transfer
, click
Capture Text
, click
Stop
.
7. HyperTerminal and the connection to the monitor can now be closed.
2ndUNIT PRINT
PLOT
>
PC-DUMP