CAMBRIDGE CSW-15AT Installation, Setup And Operation Instructions

Manual is about: Wireless Remote Control Digital Weight Indicator

Summary of CSW-15AT

  • Page 1

    1.

  • Page 2

    2 table of contents meter connections ------------------------------------------------------------------ 3 1.0 operation ------------------------------------------------------------------------ 4 1.1 key functions ---------------------------------------------------------- 4 1.2 error messages ------...

  • Page 3: Meter Connections

    3 meter connections to connect power via ac wall adapter cable to the csw-15at meter. First make sure the on/off switch on the rear of the meter is in the off position. Connect the wall adapter to your 110vac outlet. Next, connect the other end of the cable to the connector on the rear of the indica...

  • Page 4

    4 1.0 operation 1.1 key functions: zero brings the scale to a zero balance reading. If the zero key is pressed and held for 5 seconds the calibration zero value will be displayed. Grs/net toggles the display between gross weight and net weight. This key is also used to enter setup & calibration mode...

  • Page 5

    5 1.3 numeric entries: once you have entered setup and calibration mode, (as described on the previous page) some entries may require entering a numeric value. To enter a numeric value: 1. Press and release the lb/kg key to move into the desired menu field. 2. Press and release the zero key, the fir...

  • Page 6

    6 2.0 scale procedure 2.1 software navigation flowchart: scale test rs-232 end calibration s1 ntep ( ntep) n0 angle c1 zero & clear show raw counts then “0” r1 baud rate 1200,8,n,1- 19200,8,n,1? T1 version software version (gross,tare,net) weight only net only gross only s2 capacity 1 to 999,999 (50...

  • Page 7

    7 2.1 software navigation flowchart: during setup you will be required to enter make numeric entries.(ex: capacity, zero band, etc…) the following table outlines the keys used to perform these entries along with their function. Zero -------------- key is used to move up. Grs/net --------- key is use...

  • Page 8

    8 2.2 scale menu definitions: enter calibration / setup mode by pressing and holding the grs/net key until parameter ( pxxx ) event counter is displayed, then release. Immediately after code is displayed, enter in sequence (within 5 sec.) tare , lb/kg , grs/net , and print/enter . Scale will be disp...

  • Page 9

    9 2.2 scale menu definitions continued: s11 zero delay 0 to 99 updates. The weight display must be within thezero band for the selected number of updates to be considered zero. 4 (default) s12 zero 0 0.5 divisions. Tracking (1) 1 division. (default) 2 3 division press the right arrow key to enter. U...

  • Page 10

    10 3.0 calibration procedure 3.1 calibration menu definitions: c1 zero all raw counts, (pitch and roll if in angle mode) will be displayed. When zero is pressed an analog zero is done and all calibration span points will be cleared. If the indicator is in angle mode the pitch and roll offsets will a...

  • Page 11

    11 3.2 calibration: press and hold the grs/net key, as described previously in section 1.4. Scale will be displayed. Press the lb/kg key to move right until calib is displayed. Press the grs/net key to move down, c1 will be displayed. Press the lb/kg key to move right, the raw counts will be display...

  • Page 12

    12 4.0 communications setup 4.1 communications menu definitions: r1 baud rate 1200to 115200 baud (9600) default.8, n, 1 r2 output 0 gross, tare, net (default) format “cr, lf, cr, lf [32 bytes of 2eh], cr, lf, cr, lf, gross (lb or kg), :, six ascii characters [indicated weight], cr, lf, tare (lb or k...

  • Page 13

    13 5.0 testing procedures 5.1 testing menu definitions: t1 software displays the software version. Press print/enter to exit. Version t2 display flashes all display segments then all indicating enunciators. Press print/enter to exit. T3 keys press the zero key, b1 will be displayed. Press the grs/ne...

  • Page 14

    14 6.0 warranty cambridge warrants the csw-15at to be free of defects in workmanship and/or materials for 12 months from the date of shipment. This warranty of workmanship and/or materials is valid, if in the opinion of cambridge the equipment has not been mechanically, environmentally, or electrica...